Multi-wavelength spectrophotometric analysis for detection of xanthochromia in cerebrospinal fluid and accuracy for the diagnosis of subarachnoid hemorrhage

被引:10
作者
Smith, Andrew [1 ]
Wu, Alan H. B. [1 ]
Lynch, Kara L. [1 ]
Ko, Nerissa [2 ]
Grenache, David G. [3 ]
机构
[1] Univ Calif San Francisco, Dept Lab Med, San Francisco, CA 94143 USA
[2] Univ Calif San Francisco, Dept Neurovasc Serv, San Francisco, CA 94143 USA
[3] Univ Utah, Sch Med, Dept Pathol, Salt Lake City, UT USA
关键词
Xanthochromia; Cerebrospinal fluid; Bilirubin; BILIRUBIN; HEADACHE;
D O I
10.1016/j.cca.2013.06.017
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Cerebrospinal fluid (CSF) was examined for bilirubin, an important indicator for diagnosis of subarachnoid hemorrhage (SAH). Methods: A multi-wavelength (340, 415, and 460 nm) spectrophotometric assay was developed for the quantitative measurement of bilirubin in CSF, enabling the mathematical correction for absorbance of hemoglobin and proteins. Bilirubin and hemoglobin results were correlated to HPLC and a standard colorimetric assay, respectively. A subset of samples was sent for an absorbance reading at 450 nm following baseline correction. The multi-wavelength bilirubin assay was validated on 70 patients with confirmed SAH and 70 patients with neurologic symptoms who ruled out for SAH. Results: The multi-wavelength spectrophometric assay demonstrated no interferences due to proteins (albumin) up to 30 g/l or oxyhemoglobin up to 260 mg/l. The assay limit of detection was 0.2 mg/l, linear to 20 mg/l, and CVs ranged from 1 to 6% at bilirubin concentrations of 0.84 and 2.1 mg/l. The spectrophotometric assay correlated to HPLC and the colorimetric assay for bilirubin and hemoglobin, respectively. Results also correlated to the absorbance method (with removal of samples with high hemoglobin and proteins). The area under the ROC curve for diagnosis of SAH was 0.971 and 0.954 for the HPLC and spectrophotometric assay, respectively. At a cutoff of 0.2 mg/l, the clinical specificity was 100% for both assays, and the clinical sensitivity was 94.3% and 88.6% for SAH for the HPLC and spectrophotometric asays, respectively. Conclusions: The multi-wavelength spectrophotometric assay is an objective alternative to visual inspection, HPLC, and absorbance for CSF bilirubin. (c) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:231 / 236
页数:6
相关论文
共 25 条
[1]  
Ahmed DS, 2009, CLIN LAB, V55, P283
[2]   THE NATURE AND CLINICAL SIGNIFICANCE OF PIGMENTS IN THE CEREBROSPINAL FLUID [J].
BARROWS, LJ ;
HUNTER, FT ;
BANKER, BQ .
BRAIN, 1955, 78 (01) :59-80
[3]   Sentinel headache and the risk of rebleeding after aneurysmal subarachnoid hemorrhage [J].
Beck, Juergen ;
Raabe, Andreas ;
Szelenyi, Andrea ;
Berkefeld, Joachim ;
Gerlach, Ruediger ;
Setzer, Matthias ;
Seifert, Volker .
STROKE, 2006, 37 (11) :2733-2737
[4]  
Beetham R, 2009, SCAND J CLIN LAB INV, V69, P1, DOI [10.1080/00365510802651874, 10.1080/00365510802651866]
[5]   Spectrophotometric quantification of bilirubin in hemorrhagic spinal fluid using an innovative algorithm [J].
Bhadri, Prashant R. ;
Salgaonkar, Vasant A. ;
Pyne-Geithman, Gail J. ;
Caffery, James J., Jr. ;
Shukla, Rakesh ;
Beyette, Fred R., Jr. ;
Clark, Joseph F. .
MEDICINAL CHEMISTRY, 2007, 3 (01) :21-27
[6]  
Burnett RW, 1973, NATL BUR STAND, V378, P109
[7]  
Chalmers AH, 1998, CLIN CHEM, V44, P1740
[8]  
Chalmers AH, 2001, CLIN CHEM, V47, P147
[9]   XANTHOCHROMIA OF THE CEREBROSPINAL FLUID .2. PRELIMINARY DESCRIPTION OF SEVERAL NEW COLORED SUBSTANCES [J].
CROSBY, RMN ;
WEILAND, GL .
AMA ARCHIVES OF NEUROLOGY AND PSYCHIATRY, 1953, 69 (06) :732-736
[10]   Revised national guidelines for analysis of cerebrospinal fluid for bilirubin in suspected subarachnoid haemorrhage [J].
Cruickshank, Anne ;
Auld, Peter ;
Beetham, Robert ;
Burrows, Gillian ;
Egner, William ;
Holbrook, Ian ;
Keif, Geoff ;
Lewis, Emma ;
Patel, Dina ;
Watson, Ian ;
White, Peter .