Sensitivity of 3D Dose Verification to Multileaf Collimator Misalignments in Stereotactic Body Radiation Therapy of Spinal Tumor

被引:1
作者
Ni Xin-ye [1 ]
Ren, Lei [2 ]
Yan, Hui [2 ]
Yin, Fang-Fang [2 ]
机构
[1] Nanjing Med Univ, Peoples Hosp Changzhou 2, Dept Radiat Oncol, Changzhou 213003, Peoples R China
[2] Duke Univ, Med Ctr, Dept Radiat Oncol, Durham, NC USA
基金
中国博士后科学基金;
关键词
stereotactic body radiation therapy; spinal tumor; dose; multileaf collimator; misalignments; LEAF POSITION ACCURACY; QUALITY-ASSURANCE; IMRT QA; MLC; DELIVERY;
D O I
10.1177/1533034615610251
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: This study aimed to detect the sensitivity of Delt 4 on ordinary field multileaf collimator misalignments, system misalignments, random misalignments, and misalignments caused by gravity of the multileaf collimator in stereotactic body radiation therapy. Methods: (1) Two field sizes, including 2.00 cm (X) x 6.00 cm (Y) and 7.00 cm (X) x 6.00 cm (Y), were set. The leaves of X1 and X2 in the multileaf collimator were simultaneously opened. (2) Three cases of stereotactic body radiation therapy of spinal tumor were used. The dose of the planning target volume was 1800 cGy with 3 fractions. The 4 types to be simulated included (1) the leaves of X1 and X2 in the multileaf collimator were simultaneously opened, (2) only X1 of the multileaf collimator and the unilateral leaf were opened, (3) the leaves of X1 and X2 in the multileaf collimator were randomly opened, and (4) gravity effect was simulated. The leaves of X1 and X2 in the multileaf collimator shifted to the same direction. The difference between the corresponding 3-dimensional dose distribution measured by Delt 4 and the dose distribution in the original plan made in the treatment planning system was analyzed with gamma index criteria of 3.0 mm/3.0%, 2.5 mm/2.5%, 2.0 mm/2.0%, 2.5 mm/1.5%, and 1.0 mm/1.0%. Results: (1) In the field size of 2.00 cm(X) x 6.00 cm (Y), the gamma pass rate of the original was 100% with 2.5 mm/2.5% as the statistical standard. The pass rate decreased to 95.9% and 89.4% when the X1 and X2 directions of the multileaf collimator were opened within 0.3 and 0.5mm, respectively. In the field size of 7.00 (X) cm x 6.00 (Y) cm with 1.5mm/1.5% as the statistical standard, the pass rate of the original was 96.5%. After X1 and X2 of the multileaf collimator were opened within 0.3 mm, the pass rate decreased to lower than 95%. The pass rate was higher than 90% within the 3 mm opening. (2) For spinal tumor, the change in the planning target volume V-18 under various modes calculated using treatment planning system was within 1%. However, the maximum dose deviation of the spinal cord was high. In the spinal cord with a gravity of -0.25 mm, the maximum dose deviation minimally changed and increased by 6.8% than that of the original. In the largest opening of 1.00 mm, the deviation increased by 47.7% than that of the original. Moreover, the pass rate of the original determined through Delt 4 was 100% with 3 mm/3% as the statistical standard. The pass rate was 97.5% in the 0.25 mm opening and higher than 95% in the 0.5 mm opening A, 0.25 mm opening A, whole gravity series, and 0.20 mm random opening. Moreover, the pass rate was higher than 90% with 2.0 mm/2.0% as the statistical standard in the original and in the 0.25 mm gravity. The difference in the pass rates was not statistically significant among the -0.25 mm gravity, 0.25 mm opening A, 0.20 mm random opening, and original as calculated using SPSS 11.0 software with P >.05. Conclusions: Different analysis standards of Delt 4 were analyzed in different field sizes to improve the detection sensitivity of the multileaf collimator position on the basis of 90% throughout rate. In stereotactic body radiation therapy of spinal tumor, the 2.0 mm/2.0% standard can reveal the dosimetric differences caused by the minor multileaf collimator position compared with the 3.0 mm/3.0% statistical standard. However, some position derivations of the misalignments that caused high dose amount to the spinal cord cannot be detected. However, some misalignments were not detected when a large number of multileaf collimator were administered into the spinal cord.
引用
收藏
页码:NP25 / NP34
页数:10
相关论文
共 22 条
[1]  
[Anonymous], ESTRO BOOKLET 9 GUID
[2]   Requirements for leaf position accuracy for dynamic multileaf collimation [J].
Budgell, GJ ;
Mott, JHL ;
Williams, PC ;
Brown, KJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (05) :1211-1227
[3]   Automatic detection of single MLC leaf positions with corrections for penumbral effects and portal imager dose rate characteristics [J].
Eilertsen, K .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (02) :313-334
[4]   Development of a novel ArcCHECK™ insert for routine quality assurance of VMAT delivery including dose calculation with inhomogeneities [J].
Fakir, H. ;
Gaede, S. ;
Mulligan, M. ;
Chen, J. Z. .
MEDICAL PHYSICS, 2012, 39 (07) :4203-4208
[5]   On the sensitivity of common gamma-index evaluation methods to MLC misalignments in Rapidarc quality assurance [J].
Heilemann, G. ;
Poppe, B. ;
Laub, W. .
MEDICAL PHYSICS, 2013, 40 (03)
[6]   Quality assurance methodology for Varian RapidArc treatment plans [J].
Iftimia, Ileana ;
Cirino, Eileen T. ;
Xiong, Li ;
Mower, Herbert W. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2010, 11 (04) :130-143
[7]   Task Group 142 report: Quality assurance of medical accelerators [J].
Klein, Eric E. ;
Hanley, Joseph ;
Bayouth, John ;
Yin, Fang-Fang ;
Simon, William ;
Dresser, Sean ;
Serago, Christopher ;
Aguirre, Francisco ;
Ma, Lijun ;
Arjomandy, Bijan ;
Liu, Chihray ;
Sandin, Carlos ;
Holmes, Todd .
MEDICAL PHYSICS, 2009, 36 (09) :4197-4212
[8]   Dosimetric verification of RapidArc treatment delivery [J].
Korreman, Stine ;
Medin, Joakim ;
Kjaer-Kristoffersen, Flemming .
ACTA ONCOLOGICA, 2009, 48 (02) :185-191
[9]   Pretreatment quality assurance of flattening filter free beams on 224 patients for intensity modulated plans: A multicentric study [J].
Lang, Stephanie ;
Reggiori, Giacomo ;
Puxeu Vaque, Josep ;
Calle, Carlos ;
Hrbacek, Jan ;
Kloeck, Stephan ;
Scorsetti, Marta ;
Cozzi, Luca ;
Mancosu, Pietro .
MEDICAL PHYSICS, 2012, 39 (03) :1351-1356
[10]   Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy [J].
LoSasso, T ;
Chui, CS ;
Ling, CC .
MEDICAL PHYSICS, 1998, 25 (10) :1919-1927