Evaluation of stent visibility by flat panel detector CT in patients treated for intracranial aneurysms

被引:19
作者
Clarencon, Frederic [1 ,2 ]
Piotin, Michel [3 ]
Pistocchi, Silvia [3 ]
Babic, Drazenko [4 ]
Blanc, Raphael [3 ]
机构
[1] Hop La Pitie Salpetriere, Dept Neuroradiol, F-75013 Paris, France
[2] Grp Hosp Pitie Salpetriere, F-75013 Paris, France
[3] Fondat A de Rothschild, F-75019 Paris, France
[4] Philips Healthcare, Best, Netherlands
关键词
Flat panel detector; CT; Stent; Intracranial aneurysm; Artifacts; NEUROFORM STENT; CEREBRAL ANEURYSMS; ANGIOGRAPHIC CT; IN-VITRO;
D O I
10.1007/s00234-011-1002-8
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
This study aimed to evaluate the visibility of stents using high-resolution computed tomography (CT) acquisitions acquired with flat panel detector (XperCT, Allura series, Philips Healthcare, The Netherlands) for endovascular treatment of intracranial aneurysms. On a 24-month period, 48 patients endovascularly treated by coiling and stenting (59 stents) for intracranial aneurysms were explored by flat panel detector CT technique. A sequence of 620 2D images was acquired over an angle of 240A degrees using a 1,024 x 1,024 pixel matrix detector within a 48-cm field of view. The images were retrospectively analyzed independently by two neuroradiologists. Evaluation criteria were percentage of visualization of the stents and stent deployment (kinking or unsatisfactory deployment of the stent). Evaluation of the stent was feasible for all the patients. Stent visibility by XperCT was overall estimated at 76% of the stent length. Difficulties to analyze the stents were related to coil artifacts but not to packing density or aneurysm location. Stent length visualization was higher when the acquisition was performed before additional coiling (P < 0.0001). Mild kinking/misdeployment was noticed in 22% of the cases. XperCT technique provides multiplanar and 3D reconstructions that allows for a satisfying visualization of intracranial stents. This CT-like acquisition should be performed after the stent deployment and before coiling, in order to obtain better stent visualization.
引用
收藏
页码:1121 / 1125
页数:5
相关论文
共 18 条
[1]  
Benndorf G, 2005, AM J NEURORADIOL, V26, P1813
[2]   Wall Shear Stress in Intracranial Self-Expanding Stents Studied Using Ultra-High-Resolution 3D Reconstructions [J].
Benndorf, G. ;
Ionescu, M. ;
Alvarado, M. V. y ;
Hipp, J. ;
Metcalfe, R. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2009, 30 (03) :479-486
[3]  
Benndorf G, 2006, NEUROSURGERY, V58, DOI DOI 10.1227/01.NEU.0000205287.06739.E1
[4]   Neuroform stent-assisted coil embolization of wide-neck intracranial aneurysms: Strategies in stent deployment and midterm follow-up [J].
Biondi, Alessandra ;
Janardhan, Vallabh ;
Katz, Jeffrey M. ;
Salvaggio, Kimberly ;
Riina, Howard A. ;
Gobin, Y. Pierre .
NEUROSURGERY, 2007, 61 (03) :460-468
[5]   Angiographic CT with intravenous administration of contrast medium is a noninvasive option for follow-up after intracranial stenting [J].
Buhk, Jan-Hendrik ;
Lingor, Paul ;
Knauth, Michael .
NEURORADIOLOGY, 2008, 50 (04) :349-354
[6]   Rotational Flat-Panel Computed Tomography in Diagnostic and Interventional Neuroradiology [J].
Doerfler, A. ;
Struffert, T. ;
Engelhorn, T. ;
Richter, G. .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2008, 180 (10) :891-898
[7]  
Ebrahimi N, 2007, AM J NEURORADIOL, V28, P823
[8]   Balloon-in-stent technique for the constructive endovascular treatment of "ultra-wide necked" circumferential aneurysms [J].
Fiorella, D ;
Albuquerque, FC ;
Masaryk, TJ ;
Rasmussen, PA ;
McDougall, CG .
NEUROSURGERY, 2005, 57 (06) :1218-1226
[9]   Usefulness of the neuroform stent for the treatment of cerebral aneurysms: Results at initial (3-6-mo) follow-up [J].
Fiorella, D ;
Albuquerque, FC ;
Deshmukh, VR ;
McDougall, CG .
NEUROSURGERY, 2005, 56 (06) :1191-1201
[10]  
Fiorella D, 2006, NEUROSURGERY, V59, P291, DOI 10.1227/01.NEU.0000223650.11954.6C