The quality of reconstructed 3D images in multidetector-row helical CT: Experimental study involving scan parameters

被引:8
作者
Shin, JH
Lee, HK
Choi, CG
Suh, DC
Lim, TH
Kang, W
机构
[1] Univ Ulsan, Coll Med, Asan Med Ctr, Dept Radiol, Seoul 138736, South Korea
[2] Daejeon Univ, Dept Informat & Stat, Taejon, South Korea
关键词
computed tomography (CT); helical; three-dimensional; experimental study;
D O I
10.3348/kjr.2002.3.1.49
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To determine which multicletector-row helical CT scanning technique provides the best-quality reconstructed 3D images, and to assess differences in image quality according to the levels of the scanning parameters used. Materials and Methods: Four objects with different surfaces and contours were scanned using multicletector-row helical CT at three detector-row collimations (1.25, 2.50, 5.00 mm), two pitches (3.0, 6.0), and three different degrees of overlap between the reconstructed slices (0%, 25%, 50%). Reconstructed 3D images of the resulting 72 sets of data were produced using volumetric rendering. The 72 images were graded on a scale from 1 (worst) to 5 (best) for each of four rating criteria, giving a mean score for each criterion and an overall mean score. Statistical analysis was used to assess differences in image quality according to scanning parameter levels. Results: The mean score for each rating criterion, and the overall mean score, varied significantly according to the scanning parameter levels used. With regard to detector-row collimation and pitch, all levels of scanning parameters gave rise to significant differences, while in the degree of overlap of reconstructed slices, there were significant differences between overlap of 0% and of 50% in all levels of scanning parameters, and between overlap of 25% and of 50% in overall accuracy and overall mean score. Among the 18 scanning sequences, the highest score (4.94) was achieved with 1.25 mm detector-row collimation, 3.0 pitch, and 50% overlap between reconstructed slices. Conclusion: Comparison of the quality of reconstructed 3D images obtained using multidetector-row helical CT and various scanning techniques indicated that the 1.25 mm, 3.0, 50% scanning sequence was best. Quality improved as detector-row collimation decreased; as pitch was reduced from 6.0 to 3.0; and as overlap between reconstructed slices increased.
引用
收藏
页码:49 / 56
页数:8
相关论文
共 18 条
  • [11] Multi-detector row CT of thoracic disease with emphasis on 3D volume rendering and CT angiography
    Lawler, LP
    Fishman, EK
    [J]. RADIOGRAPHICS, 2001, 21 (05) : 1257 - 1273
  • [12] Dual-slice spiral versus single-slice spiral scanning: Comparison of the physical performance of two computed tomography scanners
    Liang, Y
    Kruger, RA
    [J]. MEDICAL PHYSICS, 1996, 23 (02) : 205 - 220
  • [13] COMPARISON OF SPIRAL COMPUTED-TOMOGRAPHY VERSUS CONVENTIONAL COMPUTED-TOMOGRAPHY MULTIPLANAR RECONSTRUCTIONS OF A FRACTURE DISPLACEMENT PHANTOM
    MCENERY, KW
    WILSON, AJ
    MURPHY, WA
    [J]. INVESTIGATIVE RADIOLOGY, 1994, 29 (07) : 665 - 670
  • [14] 3-DIMENSIONAL VOLUMETRIC DISPLAY OF CT DATA - EFFECT OF SCAN PARAMETERS UPON IMAGE QUALITY
    NEY, DR
    FISHMAN, EK
    MAGID, D
    ROBERTSON, DD
    KAWASHIMA, A
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1991, 15 (05) : 875 - 885
  • [15] COMPARISON OF HELICAL AND SERIAL CT WITH REGARD TO 3-DIMENSIONAL IMAGING OF MUSCULOSKELETAL ANATOMY
    NEY, DR
    FISHMAN, EK
    KAWASHIMA, A
    ROBERTSON, DD
    SCOTT, WW
    [J]. RADIOLOGY, 1992, 185 (03) : 865 - 869
  • [16] Multislice helical CT of focal and diffuse lung disease: Comprehensive diagnosis with reconstruction of contiguous and high-resolution CT sections from a single thin-collimation scan
    Schoepf, UJ
    Bruening, RD
    Hong, C
    Eibel, R
    Aydemir, S
    Crispin, A
    Becker, C
    Reiser, MF
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 177 (01) : 179 - 184
  • [17] Algorithm for image reconstruction in multi-slice helical CT
    Taguchi, K
    Aradate, H
    [J]. MEDICAL PHYSICS, 1998, 25 (04) : 550 - 561
  • [18] Liver lesions: Improved detection with dual-detector-array CT and routine 2.5-mm thin collimation
    Weg, N
    Scheer, MR
    Gabor, MP