A preliminary verification of the floating reference measurement method for non-invasive blood glucose sensing

被引:17
|
作者
Min, Xiaolin [1 ]
Liu, Rong [1 ]
Fu, Bo [1 ]
Xu, Kexin [1 ]
机构
[1] Tianjin Univ, State Key Lab Precis Measuring Technol & Instrume, Tianjin 300072, Peoples R China
基金
国家高技术研究发展计划(863计划); 中国国家自然科学基金;
关键词
Floating reference positions; Diffusion approximation equation; Monte Carlo simulation; Blood glucose; NEAR-INFRARED SPECTROSCOPY; TISSUE-SIMULATING PHANTOMS; LIGHT; REFLECTANCE; SCATTERING; TRANSPORT;
D O I
10.1016/j.optlastec.2016.12.011
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
In the non-invasive sensing of blood glucose by near-infrared diffuse reflectance spectroscopy, the spectrum is highly susceptible to the unstable and complicated background variations from the human body and the environment. In in vitro analyses, background variations are usually corrected by the spectrum of a standard reference sample that has similar optical properties to the analyte of interest. However, it is hard to find a standard sample for the in vivo measurement. Therefore, the floating reference measurement method is proposed to enable relative measurements in vivo, where the spectra under some special source-detector distance, defined as the floating reference position, are insensitive to the changes in glucose concentration due to the absorption effect and scattering effect. Because the diffuse reflectance signals at the floating reference positions only reflect the information on background variations during the measurement, they can be used as the internal reference. In this paper, the theoretical basis of the floating reference positions in a semi-infinite turbid medium was discussed based on the steady-state diffusion equation and its analytical solutions in a semi infinite turbid medium (under the extrapolated boundary conditions). Then, Monte-Carlo (MC) simulations and in vitro experiments based on a custom-built continuous-moving spatially resolving double-fiber MR measurement system, configured with two types of light source, a super luminescent diode (SLD) and a super-continuum laser, were carried out to verify the existence of the floating reference position in 5%, 10% and 20% Intralipid solutions. The results showed that the simulation values of the floating reference positions are close to the theoretical results, with a maximum deviation of approximately 0.3 mm in 1100-1320 nm. Great differences can be observed in 1340-1400 nm because the optical properties of Intralipid in this region don not satisfy the conditions of the steady-state diffusion equation. For the in vitro experiments, floating reference positions exist in 1220 nm and 1320 nm under two types of light source, and the results are quite close. However, the reference positions obtained from experiments are further from the light source compared with those obtained in the MC simulation. For the turbid media and the wavelengths investigated, the difference is up to 1 mm. This study is important for the design of optical fibers to be applied in the floating reference measurement.
引用
收藏
页码:7 / 12
页数:6
相关论文
共 50 条
  • [41] Internal floating reference method for noninvasive measurement of blood glucose with near-infrared spectroscopy
    Luo, Yun-Han
    Li, Shiping
    Huang, Furong
    OPTOELECTRONICS AND ADVANCED MATERIALS-RAPID COMMUNICATIONS, 2007, 1 (06): : 307 - 311
  • [42] Non-invasive blood glucose monitoring is an elusive goose
    Devi Dayal
    International Journal of Diabetes in Developing Countries, 2016, 36 : 399 - 400
  • [43] Non-invasive Assessment of Blood Glucose by Photo Plethysmography
    Jindal, G. D.
    Ananthakrishnan, T. S.
    Jain, Rajesh Kumar
    Sinha, Vineet
    Kini, A. R.
    Deshpande, Alakak
    IETE JOURNAL OF RESEARCH, 2008, 54 (03) : 217 - 222
  • [44] Non-invasive blood glucose monitoring is an elusive goose
    Dayal, Devi
    INTERNATIONAL JOURNAL OF DIABETES IN DEVELOPING COUNTRIES, 2016, 36 (04) : 399 - 400
  • [45] Review of Non-Invasive Glucose Sensing Techniques: Optical, Electrical and Breath Acetone
    Shokrekhodaei, Maryamsadat
    Quinones, Stella
    SENSORS, 2020, 20 (05)
  • [46] Recent Progress and Perspectives on Non-Invasive Glucose Sensors
    Davison, Nicholas B.
    Gaffney, Christopher J.
    Kerns, Jemma G.
    Zhuang, Qiandong D.
    DIABETOLOGY, 2022, 3 (01): : 56 - 71
  • [47] Extraction of Effective Signal in Non-Invasive Blood Glucose Sensing with Near-Infrared Spectroscopy
    Han Guang
    Liu Rong
    Xu Ke-xin
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38 (05) : 1599 - 1604
  • [48] Non-invasive blood glucose sensing on human body with near-infrared reflection spectroscopy
    Huang Zhen-hao
    Hao Chang-ning
    Zhang Lin-lin
    Huang Yan-chao
    Shi Yi-qin
    Jiang Geng-ru
    Duan Jun-li
    INTERNATIONAL SYMPOSIUM ON PHOTOELECTRONIC DETECTION AND IMAGING 2011: ADVANCES IN INFRARED IMAGING AND APPLICATIONS, 2011, 8193
  • [49] The feasibility investigation of non-invasive blood glucose sensing based on near-infrared spectroscopy
    Gu, Xiaoyu
    Deng, Bin
    An, Lin
    Chen, Wenliang
    Xu, Kexin
    OPTICAL DIAGNOSTICS AND SENSING VI, 2006, 6094
  • [50] Textile Organic Electrochemical Transistor for Non-Invasive Glucose Sensing
    Brendgen, Rike
    Grethe, Thomas
    Schwarz-Pfeiffer, Anne
    MICRO-SWITZERLAND, 2024, 4 (04): : 530 - 551