A preliminary verification of the floating reference measurement method for non-invasive blood glucose sensing

被引:17
|
作者
Min, Xiaolin [1 ]
Liu, Rong [1 ]
Fu, Bo [1 ]
Xu, Kexin [1 ]
机构
[1] Tianjin Univ, State Key Lab Precis Measuring Technol & Instrume, Tianjin 300072, Peoples R China
基金
国家高技术研究发展计划(863计划); 中国国家自然科学基金;
关键词
Floating reference positions; Diffusion approximation equation; Monte Carlo simulation; Blood glucose; NEAR-INFRARED SPECTROSCOPY; TISSUE-SIMULATING PHANTOMS; LIGHT; REFLECTANCE; SCATTERING; TRANSPORT;
D O I
10.1016/j.optlastec.2016.12.011
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
In the non-invasive sensing of blood glucose by near-infrared diffuse reflectance spectroscopy, the spectrum is highly susceptible to the unstable and complicated background variations from the human body and the environment. In in vitro analyses, background variations are usually corrected by the spectrum of a standard reference sample that has similar optical properties to the analyte of interest. However, it is hard to find a standard sample for the in vivo measurement. Therefore, the floating reference measurement method is proposed to enable relative measurements in vivo, where the spectra under some special source-detector distance, defined as the floating reference position, are insensitive to the changes in glucose concentration due to the absorption effect and scattering effect. Because the diffuse reflectance signals at the floating reference positions only reflect the information on background variations during the measurement, they can be used as the internal reference. In this paper, the theoretical basis of the floating reference positions in a semi-infinite turbid medium was discussed based on the steady-state diffusion equation and its analytical solutions in a semi infinite turbid medium (under the extrapolated boundary conditions). Then, Monte-Carlo (MC) simulations and in vitro experiments based on a custom-built continuous-moving spatially resolving double-fiber MR measurement system, configured with two types of light source, a super luminescent diode (SLD) and a super-continuum laser, were carried out to verify the existence of the floating reference position in 5%, 10% and 20% Intralipid solutions. The results showed that the simulation values of the floating reference positions are close to the theoretical results, with a maximum deviation of approximately 0.3 mm in 1100-1320 nm. Great differences can be observed in 1340-1400 nm because the optical properties of Intralipid in this region don not satisfy the conditions of the steady-state diffusion equation. For the in vitro experiments, floating reference positions exist in 1220 nm and 1320 nm under two types of light source, and the results are quite close. However, the reference positions obtained from experiments are further from the light source compared with those obtained in the MC simulation. For the turbid media and the wavelengths investigated, the difference is up to 1 mm. This study is important for the design of optical fibers to be applied in the floating reference measurement.
引用
收藏
页码:7 / 12
页数:6
相关论文
共 50 条
  • [1] Phantom Experimental Validation of Near Floating Reference Measuring Method in Non-Invasive Blood Glucose Sensing
    Xia Dan-dan
    Han Guang
    Yu Xu-yao
    Li Chen-xi
    Liu Rong
    Jiang Jing-ying
    Xu Ke-xin
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37 (06) : 1941 - 1945
  • [2] Study on optical measurement conditions for non-invasive blood glucose sensing
    Xu, KX
    Chen, WL
    Jiang, JY
    Qiu, QJ
    COMPLEX DYNAMICS, FLUCTUATIONS, CHAOS, AND FRACTALS IN BIOMEDICAL PHOTONICS, 2004, 5 (19): : 83 - 91
  • [3] Fundamental study on non-invasive blood glucose sensing
    Xu, Kexin
    Li, Qingbo
    Lu, Zhang
    Jiang, Jingying
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2002, 10 (3-4) : 187 - 197
  • [4] Theoretical Studies of Floating-Reference Method for NIR Blood Glucose Sensing
    Shi, Zhenzhi
    Yang, Yue
    Zhao, Huijuan
    Chen, Wenliang
    Liu, Rong
    Xu, Kexin
    OPTICAL DIAGNOSTICS AND SENSING XI: TOWARD POINT-OF-CARE DIAGNOSTICS AND DESIGN AND PERFORMANCE VALIDATION OF PHANTOMS USED IN CONJUNCTION WITH OPTICAL MEASUREMENT OF TISSUE III, 2011, 7906
  • [5] Simulation on how to customize glucose adjustment method for non-invasive blood glucose sensing by NIRS
    Min, Xiaolin
    Jiang, Jingying
    Zou, Da
    Liu, Rong
    Xu, Kexin
    DYNAMICS AND FLUCTUATIONS IN BIOMEDICAL PHOTONICS IX, 2013, 8580
  • [6] Multiparameter techniques for non-invasive measurement of blood glucose
    Amaral, Carlos F.
    Brischwein, Martin
    Wolf, Bernhard
    SENSORS AND ACTUATORS B-CHEMICAL, 2009, 140 (01): : 12 - 16
  • [7] Non-invasive blood glucose measurement Measurement of bioeletrical signal
    Anas, M. N.
    Nurun, N. K.
    Noral, A. N.
    Normahira, M.
    2012 IEEE EMBS CONFERENCE ON BIOMEDICAL ENGINEERING AND SCIENCES (IECBES), 2012,
  • [8] Effects of Skin Blood Flow Fluctuations on Non-Invasive Glucose Measurement and a Feasible Blood Flow Control Method
    Ge, Qing
    Han, Tongshuai
    Liu, Xueying
    Chen, Jiayu
    Liu, Wenbo
    Liu, Jin
    Xu, Kexin
    SENSORS, 2025, 25 (04)
  • [9] Evaluation and Validation on Sensitivity of Near-Infrared Diffuse Reflectance in Non-Invasive Human Blood Glucose Measurement
    Ge, Qing
    Han, Tongshuai
    Liu, Rong
    Zhang, Zengfu
    Sun, Di
    Liu, Jin
    Xu, Kexin
    SENSORS, 2024, 24 (18)
  • [10] Determination of the reference position in the near-infrared non-invasive blood glucose measurement in vivo
    Han, Guang
    Liu, Jin
    Liu, Rong
    Xu, Kexin
    OPTICS IN HEALTH CARE AND BIOMEDICAL OPTICS VII, 2017, 0024