Real time frequency domain fibreoptic temperature sensor using ruby crystals

被引:9
作者
Alcala, JR
Liao, SC
Zheng, J
机构
[1] Department of Biomedical Engineering, Case Western Reserve University, Cleveland
关键词
biomedical sensor; optical temperature sensor;
D O I
10.1016/1350-4533(95)00014-3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The excited state phosphorescence lifetime of ruby crystals is used to monitor temperature in the physiological range from 15 degrees to 45 degrees C with precision and accuracy less than 1 degrees C, in real time. Precision of 0.1 degrees C is attained with 3 min integration times. A 500 mu m cubic ruby crystal bounded to the distal end of an optical fibre of similar core dimensions is excited with pulsed Ne-He laser light of about 9 mu W average power The instrument uses a sampler for data acquisition, and frequency domain methods for data fitting. The instrument amplifies the a.c. components of the detector output and band limits the signal to 800 Hz. The fundamental frequency of the excitation is set to 24.41 Hz to obtain 32 or less harmonics. This band-limited signal is sampled and averaged between 20 and 100 cycles to obtain temperature measurements in real time. The frequency domain representation of the data is obtained by employing fast Fourier transform algorithms. The phase delay and the modulation ratio, of each sampled harmonic, is then computed. Five to 32 values of the phase and modulation are averaged before computing the sensor lifetime. The technique is capable of measuring precise and accurate excited state lifetimes from subpicowatt luminescent signals in plastic optical fibres. A least squares fit yields the lifetimes of single exponentials. A component of zero lifetime is introduced to account for the backscatter excitation sea by the photodetector leaking through optical interference filters. The phosphorescence lifetimes measured reproducibly to about six pa?ts in 1000, with a 2 s integration time, are used to monitor physiological temperature. Temperatures are computed employing empirical polynomials. The system drift is 3% over 5 h of continuous operation. The instrumentation and methods allow 2.7 s update times and 50 s full response times.
引用
收藏
页码:51 / 56
页数:6
相关论文
共 37 条
[1]   INTERPRETATION OF FLUORESCENCE DECAYS IN PROTEINS USING CONTINUOUS LIFETIME DISTRIBUTIONS [J].
ALCALA, JR ;
GRATTON, E ;
PRENDERGAST, FG .
BIOPHYSICAL JOURNAL, 1987, 51 (06) :925-936
[2]   FLUORESCENCE LIFETIME DISTRIBUTIONS IN PROTEINS [J].
ALCALA, JR ;
GRATTON, E ;
PRENDERGAST, FG .
BIOPHYSICAL JOURNAL, 1987, 51 (04) :597-604
[3]   RESOLVABILITY OF FLUORESCENCE LIFETIME DISTRIBUTIONS USING PHASE FLUOROMETRY [J].
ALCALA, JR ;
GRATTON, E ;
PRENDERGAST, FG .
BIOPHYSICAL JOURNAL, 1987, 51 (04) :587-596
[4]  
ALCALA JR, 1985, ANAL INSTRUM, V14, P225
[5]   DIGITAL PARALLEL ACQUISITION IN FREQUENCY-DOMAIN FLUOROMETRY - COMMENT [J].
ALCALA, JR .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (06) :1672-1673
[6]   DIGITAL PHOSPHORIMETER WITH FREQUENCY-DOMAIN SIGNAL-PROCESSING - APPLICATION TO REAL-TIME FIBEROPTIC OXYGEN SENSING [J].
ALCALA, JR ;
YU, C ;
YEH, GJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (06) :1554-1560
[7]   REAL-TIME FREQUENCY-DOMAIN FIBEROPTIC TEMPERATURE SENSOR [J].
ALCALA, JR ;
LIAO, SC ;
ZHENG, JL .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1995, 42 (05) :471-476
[8]   THE EFFECT OF HARMONIC CONFORMATIONAL TRAJECTORIES ON PROTEIN FLUORESCENCE AND LIFETIME DISTRIBUTIONS [J].
ALCALA, JR .
JOURNAL OF CHEMICAL PHYSICS, 1994, 101 (06) :4578-4584
[9]  
ALCALA JR, 1987, THESIS U ILLINOIS UR
[10]  
ALCALA JR, 1993, SPIE, V1885, P306