Timing of delayed perforation with the St. Jude Riata lead: A single-center experience and a review of the literature

被引:21
作者
Danik, Stephan B. [1 ]
Mansour, Moussa [1 ]
Heist, E. Kevin [1 ]
Ellinor, Patrick [1 ]
Milan, David [1 ]
Singh, Jagmeet [1 ]
Das, Saumya [1 ]
Reddy, Vivek [1 ]
D'Avila, Andre [1 ]
Ruskin, Jeremy N. [1 ]
Mela, Theofanie [1 ]
机构
[1] Harvard Univ, Sch Med, Massachusetts Gen Hosp, Cardiac Arrhythemia Serv, Boston, MA 02114 USA
关键词
Defibrillator; Right ventricular lead; Lead dislodgement; Lead revision; Cardiac perforation;
D O I
10.1016/j.hrthm.2008.09.017
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Recent studies have reported a possible increased risk of delayed perforation with the St. Jude Riata defibrillator lead. OBJECTIVE The purpose of this study was to determine the incidence and time of occurrence of this complication in a large number of patients who underwent implantationm at a single center. METHODS We reviewed the data and clinical course on all patients who underwent implantation of the St. Jude Riata (1580/1581, 1590/1591, and the 7000/7001 ST series) right ventricular defibrillator lead at the Massachusetts General Hospital between June 2001 and October 2007. Data were compared with all Medtronic Sprint Fidelis leads implanted during the same time period. RESULTS Of a total of 416 implanted Riata leads, follow-up data at 6 weeks and 3 months was available for 87% and 75% of patients, respectively. There were 6 cases of lead perforation (6 of 233, 2.6%) with the 1580/1581 series, no cases of lead perforation (0 of 92) with the 1590/1591 series, and 2 cases of perforation (2 of 92, 2.2%) with the 7000/7001 ST series. Interrogation the day after implantation did not reveal any abnormalities. Patients with perforation developed symptoms of chest pain or shortness of breath within 3 weeks post-implantation; interrogation revealed pacing thresholds of > 5 V at 0.5 ms within the first follow-up period of 6 weeks. CONCLUSION All cases of perforation with the St. Jude Riata lead occurred within 6 weeks post-implantation and showed pacing thresholds > 5 V at 0.5 ms. There were no cases of delayed perforation after 6 weeks in 307 patients with at least 3 months of follow-up.
引用
收藏
页码:1667 / 1672
页数:6
相关论文
共 16 条
[1]  
Carlson MD, 2008, PACE, V31, P13, DOI 10.1111/j.1540-8159.2007.00943.x
[2]   Does size really matter? A comparison of the Riata lead family based on size and its relation to performance [J].
Corbisiero, Raffaele ;
Armbruster, Rebecca .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2008, 31 (06) :722-726
[3]   Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator [J].
Danik, Stephan B. ;
Mansour, Moussa ;
Singh, Jagmeet ;
Reddy, Vivek Y. ;
Ellinor, Patrick T. ;
Milan, David ;
Heist, E. Kevin ;
d'Avila, Andre ;
Ruskin, Jeremy N. ;
Mela, Theofanie .
HEART RHYTHM, 2007, 4 (04) :439-442
[4]  
Fisher JD, 2008, PACE, V31, P7, DOI 10.1111/j.1540-8159.2007.00918.x
[5]   Early failure of a small-diameter high-voltage implantable cardioverter-defibrillator lead [J].
Hauser, Robert G. ;
Kallinen, Linda M. ;
Almquist, Adrian K. ;
Gornick, Charles C. ;
Katsiyiannis, William T. .
HEART RHYTHM, 2007, 4 (07) :892-896
[6]  
Hirschl DA, 2007, PACE, V30, P28
[7]  
Krivan L, 2008, PACE, V31, P3, DOI 10.1111/j.1540-8159.2007.00917.x
[8]  
LABORDERI J, 2008, HEART RHYTHM S, V5, pS139
[9]   More late perforations with the Riata™ defibrillator lead from a high-volume center:: An update on the numbers [J].
Lloyd, Michael S. ;
Shaik, M. Naushad ;
Riley, Michael ;
Langberg, Jonathan J. .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2008, 31 (06) :784-785
[10]   Late right ventricular perforation and hemothorax after transvenous defibrillator lead implantation [J].
Merla, Ramanna ;
Reddy, Nischita K. ;
Kunapuli, Sanjay ;
Schwarz, Ernst ;
Vitarelli, Antonio ;
Rosanio, Salvatore .
AMERICAN JOURNAL OF THE MEDICAL SCIENCES, 2007, 334 (03) :209-211