What Is the Critical Tibial Resection Depth During Unicompartmental Knee Arthroplasty? A Biomechanical Study of Fracture Risk

被引:13
作者
Houskamp, Daniel J. [1 ]
Tompane, Trevor [1 ]
Barlow, Brian T. [1 ]
机构
[1] Naval Med Ctr, 34800 Bob Wilson Dr, San Diego, CA 92134 USA
关键词
unicompartmental knee arthroplasty; fracture; tibial plateau fracture; biomechanical study; periprosthetic fracture; PLATEAU; COMPLICATION; FIXATION;
D O I
10.1016/j.arth.2020.04.005
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Fracture after medial unicompartmental knee arthroplasty (UKA) is a rare complication. Biomechanical studies evaluating association between depth of resection and maximum load to failure are lacking. The purpose of this study is to establish the relationship between depth of resection of the medial tibial plateau and mean maximum load to failure. Methods: Medial tibial resections were performed from 2 to 10 mm in 25 standardized fourth-generation Sawbones composite tibias (Sawbones, Vashon Island, Washington). A metal-backed tibial component with a 9-mm polyethylene bearing was used (Stryker PKR). Tibias were mounted on a biomechanical testing apparatus (MTESTQuattro) and axially loaded cyclically 10 times per cycle and incrementally increased until failure occurred. Results: Load to failure was recorded in 25 proximal tibia model samples after medial UKA using sequential resections from 2 to 10 mm. Analysis of variance testing identified significant differences in mean maximum load to failure between groups (P = .0003). Analysis of regression models revealed a statistically significant fit of a quadratic model (R-2 = 0.59, P = .0001). The inflection point of this quadratic curve was identified at 5.82 mm, indicating that the maximum load to failure across experimental models in this study began to decline beyond a resection depth of 5.82 mm. Conclusion: In this biomechanical model, medial tibial resections beyond 5.82 mm produced a significantly lower mean load to failure using a quadratic curve model. Resections from 2 to 6 mm showed no significant differences in mean load to failure. Identification of the tibial resection depth at which the mean load to failure significantly decreases is clinically relevant as this depth may increase the risk of periprosthetic fracture after a medial UKA. Published by Elsevier Inc.
引用
收藏
页码:2244 / 2248
页数:5
相关论文
共 24 条
[1]   Experimental model of tibial plateau fracture for biomechanical testing [J].
Ali, Ahmad M. ;
Saleh, Michael ;
Bolongaro, Stefano ;
Yang, Lang .
JOURNAL OF BIOMECHANICS, 2006, 39 (07) :1355-1360
[2]   Periprosthetic Fracture following Partial Knee Arthroplasty [J].
Brown, Nicholas M. ;
Engh, Gerard ;
Fricka, Kevin .
JOURNAL OF KNEE SURGERY, 2019, 32 (10) :947-952
[3]   Tibial plateau stress fracture - A complication of unicompartmental knee arthroplasty using 4 guide pinholes [J].
Brumby, SA ;
Carrington, R ;
Zayontz, S ;
Reish, T ;
Scott, RD .
JOURNAL OF ARTHROPLASTY, 2003, 18 (06) :809-812
[4]   Biomechanical impact testing of synthetic versus human cadaveric tibias for predicting injury risk during pedestrian-vehicle collisions [J].
Cameron, Muir W. ;
Schemitsch, Emil H. ;
Zdero, Radovan ;
Quenneville, Cheryl E. .
TRAFFIC INJURY PREVENTION, 2020, 21 (02) :163-168
[5]   Unicompartmental Knee Arthroplasty Provides Significantly Greater Improvement in Function than Total Knee Arthroplasty Despite Equivalent Satisfaction for Isolated Medial Compartment Osteoarthritis [J].
Casper, David S. ;
Fleischman, Andrew N. ;
Papas, Paraskevi Vivian ;
Grossman, Jamie ;
Scuderi, Giles R. ;
Lonner, Jess H. .
JOURNAL OF ARTHROPLASTY, 2019, 34 (08) :1611-1616
[6]   Periprosthetic tibial fractures in unicompartmental knee arthroplasty as a function of extended sagittal saw cuts: An experimental study [J].
Clarius, M. ;
Haas, D. ;
Aldinger, P. R. ;
Jaeger, S. ;
Jakubowitz, E. ;
Seeger, J. B. .
KNEE, 2010, 17 (01) :57-60
[7]   Management of the Failed Medial Unicompartmental Knee Arthroplasty [J].
Crawford, David A. ;
Berend, Keith R. ;
Lombardi, Adolph, V .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2018, 26 (20) :E426-E433
[8]   Mechanical validation of whole bone composite tibia models [J].
Cristofolini, L ;
Viceconti, M .
JOURNAL OF BIOMECHANICS, 2000, 33 (03) :279-288
[9]   Composite Bone Models in Orthopaedic Surgery Research and Education [J].
Elfar, John J. ;
Menorca, Ron Martin Garcia ;
Reed, Jeffrey Douglas ;
Stanbury, Spencer .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2014, 22 (02) :111-120
[10]   Mechanical Evaluation of Large-Size Fourth-Generation Composite Femur and Tibia Models [J].
Gardner, Matthew P. ;
Chong, Alexander C. M. ;
Pollock, Anthony G. ;
Wooley, Paul H. .
ANNALS OF BIOMEDICAL ENGINEERING, 2010, 38 (03) :613-620