Accuracy and predictability of the compensatory function of Orbscan II in intraocular pressure measurements after laser in situ keratomileusis

被引:13
作者
Lee, DH
Seo, S
Shin, SC
Chung, EH
Turner, TT
机构
[1] Inje Univ, Coll Med, Ilsan Paik Hosp, Dept Ophthalmol, Koyang 411410, Kyunggyi Do, South Korea
[2] Bausch & Lomb Inc, Claremont, CA USA
关键词
D O I
10.1016/S0886-3350(01)01286-X
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To evaluate the usefulness of the compensatory function of Orbscan II (Orbtak, Bausch & Lomb) for measuring postoperative intraoperative pressure (IOP) in laser in situ keratomileusis (LASIK) patients. Setting: Department of Ophthalmology, IIsan Pak Hospital, Inje University, IIsan, Korea. Methods; Two hundred ten patients had LASIK using the Hansatome(R) microkeratome (Bausch & Lomb) and the MEL 70 G-scan laser (Aesclepion-Meditec) and were examined retrospectively. The patients included in the study were divided into 2 groups: Group 1, 123 eyes with less than -6.0 diopters (D) of myopia; Group 2, 60 eyes with more than -6.0 D of myopia. The IOP was measured preoperatively and 2 and 4 weeks after LASIK using a noncontact tonometer (NCT CT-60, Topcon). The results were corrected with the Orbscan 11 program. Results: The mean IOPs in Group 1 were 16.35 mm Hg +/- 2.90 (SD) preoperatively, 10.80 +/- 2.21 mm Hg at 2 weeks, and 10.63 +/- 2.28 mm Hg at 4 weeks. After compensation with Orbscan 11, the 2-week and 4-week values were 16.81 +/- 3.14 mm Hg and 16.68 +/- 3.22 mm Hg, respectively, The mean IOPs in Group 2 were 17.63 +/- 2.93 mm Hg, 10.07 +/- 1.55 mm Hg, and 10.43 +/- 1.84 mm Hg, respectively; after compensation, they were 18.40 +/- 4.09 mm Hg and 18.05 +/- 4.09 mm Hg, respectively. After compensation with Orbscan 11, there were no statistically significant differences between the preoperative and postoperative IOPs. Conclusions: Orbscan II may help predict actual IOP values after LASIK and avoid the misinterpretation of high IOPs as normal IOPs, J Cataract Refract Surg 2002; 28: 259-264 (C) 2002 ASCRS and ESCRS.
引用
收藏
页码:259 / 264
页数:6
相关论文
共 18 条
[1]   Ocular hypertension and central corneal thickness [J].
Argus, WA .
OPHTHALMOLOGY, 1995, 102 (12) :1810-1812
[2]  
DRANCE SM, 1960, ARCH OPHTHALMOL-CHIC, V64, P494
[3]   Correlation of intraocular pressure and central corneal thickness in normal myopic eyes and after laser in situ keratomileusis [J].
Emara, B ;
Probst, LE ;
Tingey, DP ;
Kennedy, DW ;
Willms, LJ ;
Machat, J .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (10) :1320-1325
[4]   Accuracy of Goldmann tonometry after refractive surgery [J].
Faucher, A ;
Gregoire, J ;
Blondeau, P .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1997, 23 (06) :832-838
[5]   Intraocular pressure change measured by Goldmann tonometry after laser in situ keratomileusis [J].
Fournier, AV ;
Podtetenev, M ;
Lemire, J ;
Thompson, P ;
Duchesne, R ;
Perreault, C ;
Chehade, N ;
Blondeau, P .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (07) :905-910
[6]   Central corneal thickness in normal, glaucomatous, and ocular hypertensive eyes [J].
Herndon, LW ;
Choudhri, SA ;
Cox, T ;
Damji, KF ;
Shields, MB ;
Allingham, RR .
ARCHIVES OF OPHTHALMOLOGY, 1997, 115 (09) :1137-1141
[7]   Accuracy of Orbscan pachymetry measurements and DHG ultrasound pachymetry in primary laser in situ keratomileusis and LASIK enhancement procedures [J].
Iskander, NG ;
Penno, EA ;
Peters, NT ;
Gimbel, HV ;
Ferensowicz, M .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (05) :681-685
[8]   The effect of excimer laser photorefractive keratectomy on intraocular pressure measurements using the Goldmann applanation tonometer [J].
Mardelli, PG ;
Piebenga, LW ;
Whitacre, MM ;
Siegmund, KD .
OPHTHALMOLOGY, 1997, 104 (06) :945-948
[9]   CORNEAL CURVATURE IN APPLANATION TONOMETRY [J].
MARK, HH .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1973, 76 (02) :223-224
[10]  
Munger R, 1998, CAN J OPHTHALMOL, V33, P159