Continuous measurement of segmental and whole body bio-impedance

被引:0
|
作者
Zhu, FS [1 ]
Schneditz, D [1 ]
Wang, EJ [1 ]
Levin, NW [1 ]
机构
[1] Beth Israel Deaconess Med Ctr, New York, NY 10128 USA
来源
PROCEEDINGS OF THE 19TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 19, PTS 1-6: MAGNIFICENT MILESTONES AND EMERGING OPPORTUNITIES IN MEDICAL ENGINEERING | 1997年 / 19卷
关键词
bioimpedance; extracellular resistance; segmental measurement; regional fluid distribution; MULTIFREQUENCY IMPEDANCE; WATER; BIOIMPEDANCE; DIALYSIS;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Measurement of whole body bioimpedance provides a non-invasive and practical approach to estimate extracellular volume (ECV) in man. However, recent data indicate that ECV cannot be treated as a homogeneous compartment without obtaining paradoxical results. Therefore, bioimpedance measurements should include measurements of body segments such as the arm, the trunk, and the leg. A digital switch was developed and used to consecutively select different body segments for the continuous measurement of whole body and segmental bioimpedance at a minimum sampling period of 3 seconds. Bioimpedance data were modeled to determine the resistance of the ECV in the segments and in the whole body for a period of 30 minutes where subjects were in supine body position followed by a 30 minute standing phase. During supine body position modeled extracellular resistance increased in the leg, the arm, and in the whole body but slightly increased in the trunk. Changes in extracellular resistance reversed with subjects assuming a standing position. These measurements reflect the shifts in ECV between segments to be expected with changes in body position. Simultaneous measurement of bioimpedance in different body segments utilizing this approach might be useful to analyze the nature of fluid shifts in clinical applications such as hemodialysis and ultrafiltration.
引用
收藏
页码:2086 / 2088
页数:3
相关论文
共 50 条
  • [21] Monitoring Tissue Evolution on Electrodes with Bio-Impedance Test
    Maldonado, A.
    Perez, P.
    Huertas, G.
    Yufera, A.
    Rueda, A.
    Huertas, J. L.
    2016 CONFERENCE ON DESIGN OF CIRCUITS AND INTEGRATED SYSTEMS (DCIS 2016), 2016, : 213 - 217
  • [22] Dry Textile Electrodes for Wearable Bio-Impedance Analyzers
    Corchia, Laura
    Monti, Giuseppina
    Raheli, Federica
    Candelieri, Giulia
    Tarricone, Luciano
    IEEE SENSORS JOURNAL, 2020, 20 (11) : 6139 - 6147
  • [23] MEASUREMENT OF BIO-IMPEDANCE ON AN ISOLATED RAT SCIATIC NERVE OBTAINED WITH SPECIFIC CURRENT STIMULATING PULSES
    Rozman, Janez
    Zuzek, Monika C.
    Frangez, Robert
    Ribaric, Samo
    MATERIALI IN TEHNOLOGIJE, 2016, 50 (05): : 797 - 804
  • [24] A 1.55 μW Bio-Impedance Measurement System for Implantable Cardiac Pacemakers in 0.18 μm CMOS
    Zamani, Milad
    Rezaeiyan, Yasser
    Shoaei, Omid
    Serdijn, Wouter A.
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2018, 12 (01) : 211 - 221
  • [25] Design of Low Cost Bio-impedance Measuring Instrument
    Birok, Rajesh
    Kapoor, Rajiv
    INTERNATIONAL JOURNAL OF ADVANCED COMPUTER SCIENCE AND APPLICATIONS, 2022, 13 (01) : 738 - 749
  • [26] Do the Bio-impedance Models Exhibit Pinched Hysteresis?
    Abdelrahman, Doaa K.
    Mohammed, Rawan
    Fouda, Mohammed E.
    Said, Lobna A.
    Radwan, Ahmed G.
    2020 IEEE 63RD INTERNATIONAL MIDWEST SYMPOSIUM ON CIRCUITS AND SYSTEMS (MWSCAS), 2020, : 1016 - 1019
  • [27] 0.13pW/Hz Ring VCO-Based Continuous-Time Read-Out ADC for Bio-Impedance Measurement
    Danesh, Mohammadhadi
    Sanyal, Arindam
    IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS II-EXPRESS BRIEFS, 2020, 67 (12) : 2823 - 2827
  • [28] Output Impedance Comparison of Four Constant Current Sources for Bio-Impedance Applications
    Wang Wei
    Hu Rui
    Zhang Xiongxing
    Wang Kening
    JOURNAL OF NANOELECTRONICS AND OPTOELECTRONICS, 2017, 12 (05) : 437 - 441
  • [29] Soft electrodes for simultaneous bio-potential and bio-impedance study of the face
    Levit, Bara
    Funk, Paul F.
    Hanein, Yael
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2024, 10 (02)
  • [30] Cole-Cole Bio-Impedance Parameters Extraction From a Single Time-Domain Measurement
    Kapoulea, S.
    AbdelAty, A. M.
    Elwakil, A. S.
    Psychalinos, C.
    Radwan, A. G.
    2019 8TH INTERNATIONAL CONFERENCE ON MODERN CIRCUITS AND SYSTEMS TECHNOLOGIES (MOCAST), 2019,