Differentiation of thrombus from pannus as the cause of acquired mechanical prosthetic heart valve obstruction by non-invasive imaging: a review of the literature

被引:43
作者
Tanis, Wilco [1 ]
Habets, Jesse [2 ]
van den Brink, Renee B. A. [3 ]
Symersky, Petr [4 ]
Budde, Ricardo P. J. [2 ,5 ]
Chamuleau, Steven A. J. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Cardiol, NL-3508 GA Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, Dept Radiol, NL-3508 GA Utrecht, Netherlands
[3] Univ Amsterdam, Acad Med Ctr, Dept Cardiol, NL-1105 AZ Amsterdam, Netherlands
[4] Univ Amsterdam, Acad Med Ctr, Dept Thorac Surg, NL-1105 AZ Amsterdam, Netherlands
[5] Gelre Ziekenhuis, Dept Radiol, Apeldoorn, Netherlands
关键词
prosthetic heart valve obstruction; thrombus; pannus; echocardiography; fluoroscopy; computed tomography angiography; ROW COMPUTED-TOMOGRAPHY; JUDE MEDICAL VALVES; TRANSESOPHAGEAL ECHOCARDIOGRAPHY; AORTIC POSITION; SINGLE-CENTER; CINERADIOGRAPHY; THROMBOLYSIS; FIBRINOLYSIS; MANAGEMENT; TEE;
D O I
10.1093/ehjci/jet127
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
For acquired mechanical prosthetic heart valve (PHV) obstruction and suspicion on thrombosis, recently updated European Society of Cardiology guidelines advocate the confirmation of thrombus by transthoracic echocardiography, transesophageal echocardiography (TEE), and fluoroscopy. However, no evidence-based diagnostic algorithm is available for correct thrombus detection, although this is clinically important as fibrinolysis is contraindicated in non-thrombotic obstruction (isolated pannus). Here, we performed a review of the literature in order to propose a diagnostic algorithm. We performed a systematic search in Pubmed and Embase. Included publications were assessed on methodological quality based on the validated Quality Assessment of Diagnostic Accuracy Studies (QUADAS) II checklist. Studies were scarce (n 15) and the majority were of moderate methodological quality. In total, 238 mechanical PHVs with acquired obstruction and a reliable reference standard were included for the evaluation of the role of fluoroscopy, echocardiography, or multidetector-row computed tomography (MDCT). In acquired PHV obstruction caused by thrombosis, mass detection by TEE and leaflet restriction detected by fluoroscopy were observed in the majority of cases (96 and 100, respectively). In contrast, in acquired PHV obstruction free of thrombosis (pannus), leaflet restriction detected by fluoroscopy was absent in some cases (17) and mass detection by TEE was absent in the majority of cases (66). In case of mass detection by TEE, predictors for obstructive thrombus masses (compared with pannus masses) were leaflet restriction, soft echo density, and increased mass length. In situations of inconclusive echocardiography, MDCT may correctly detect pannus/thrombus based on the morphological aspects and localization. In acquired mechanical PHV obstruction without leaflet restriction and absent mass on TEE, obstructive PHV thrombosis cannot be confirmed and consequently, fibrinolysis is not advised. Based on the literature search and our opinion, a diagnostic algorithm is provided to correctly identify non-thrombotic PHV obstruction, which is highly relevant in daily clinical practice.
引用
收藏
页码:119 / 129
页数:11
相关论文
共 33 条
[1]  
[Anonymous], EUR RADIOL
[2]   Obstruction of mechanical valve prostheses: Clinical diagnosis and surgical or nonsurgical treatment [J].
Aoyagi, S ;
Fukunaga, S ;
Suzuki, S ;
Nishi, Y ;
Oryoji, A ;
Kosuga, K .
SURGERY TODAY-THE JAPANESE JOURNAL OF SURGERY, 1996, 26 (06) :400-406
[3]   Obstruction of St Jude Medical valves in the aortic position: Significance of a combination of cineradiography and echocardiography [J].
Aoyagi, S ;
Nishimi, M ;
Kawano, H ;
Tayama, E ;
Fukunaga, S ;
Hayashida, N ;
Akashi, H ;
Kawara, T .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2000, 120 (01) :142-147
[4]   Differentiating thrombus from pannus formation in obstructed mechanical prosthetic valves: An evaluation of clinical, transthoracic and transesophageal echocardiographic parameters [J].
Barbetseas, J ;
Nagueh, SF ;
Pitsavos, C ;
Toutouzas, PK ;
Quinones, MA ;
Zoghbi, WA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 32 (05) :1410-1417
[5]   Low radiation dose non-contrast cardiac CT: is it of value in the evaluation of mechanical aortic valve [J].
Bazeed, Mohamed Fayez ;
Moselhy, Mohamed Saleh ;
Rezk, Ahmad Ibrahim ;
Al-Murayeh, Mushabab Ayedh .
ACTA RADIOLOGICA, 2012, 53 (04) :389-393
[6]   OBSTRUCTION OF MECHANICAL HEART-VALVE PROSTHESES - CLINICAL ASPECTS AND SURGICAL-MANAGEMENT [J].
DEVIRI, E ;
SARELI, P ;
WISENBAUGH, T ;
CRONJE, SL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 17 (03) :646-650
[7]   Reoperation for prosthetic aortic valve obstruction in the era of echocardiography: Trends in diagnostic testing and comparison with surgical findings [J].
Girard, SE ;
Miller, FA ;
Orszulak, TA ;
Mullany, CJ ;
Montgomery, S ;
Edwards, WD ;
Tazelaar, HD ;
Malouf, JF ;
Tajik, AJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (02) :579-584
[8]   Imaging of Prosthetic Heart Valve Dysfunction Complementary Diagnostic Value of TEE and MDCT? [J].
Habets, Jesse ;
Tanis, Wilco ;
Mali, Willem P. Th M. ;
Chamuleau, Steven A. J. ;
Budde, Ricardo P. J. .
JACC-CARDIOVASCULAR IMAGING, 2012, 5 (09) :956-961
[9]   Diagnostic evaluation of left-sided prosthetic heart valve dysfunction [J].
Habets, Jesse ;
Budde, Ricardo P. ;
Symersky, Petr ;
van den Brink, Renee B. ;
de Mol, Bas A. ;
Mali, Willem P. ;
van Herwerden, Lex A. ;
Chamuleau, Steven A. .
NATURE REVIEWS CARDIOLOGY, 2011, 8 (08) :466-478
[10]   DIAGNOSIS OF PROSTHETIC HEART-VALVE THROMBOSIS - THE RESPECTIVE VALUES OF TRANSTHORACIC AND TRANSESOPHAGEAL DOPPLER ECHOCARDIOGRAPHY [J].
HABIB, G ;
CORNEN, A ;
MESANA, T ;
MONTIES, JR ;
DJIANE, P ;
LUCCIONI, R .
EUROPEAN HEART JOURNAL, 1993, 14 (04) :447-455