Meta-analyses comparing spine simulators with cadavers and finite element models by analysing range-of-motion data before and after lumbar total disc replacement

被引:5
作者
Bohn, Tobias [1 ,2 ]
Lang, Susanne A. J. [1 ]
Roll, Stephanie [3 ]
Schrader, Helene [1 ]
Pumberger, Matthias [4 ]
Buettner-Janz, Karin [1 ,5 ]
机构
[1] Charite Univ Med Berlin, Charitepl 1, D-10117 Berlin, Germany
[2] Ortenau Klinikum Offenburg Kehl, Inst Radiol, Ebertpl 12, D-77654 Offenburg, Germany
[3] Charite Univ Med Berlin, Inst Social Med Epidemiol & Hlth Econ, Charitepl 1, D-10117 Berlin, Germany
[4] Charite Univ Med Berlin, Ctr Musculoskeletal Surg, Charitepl 1, D-10117 Berlin, Germany
[5] Buttner Janz Spinefdn, Meinekestr 6, D-10719 Berlin, Germany
关键词
Meta-analysis; In vitro test methods; Range of motion; Total disc replacement; COMPRESSIVE FOLLOWER PRELOAD; ARTIFICIAL DISC; PRODISC-L; BIOMECHANICAL EVALUATION; ARTHROPLASTY; KINEMATICS; PROSTHESIS; LOAD; ANTERIOR; IMPLANT;
D O I
10.1016/j.jare.2020.06.017
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background: Range-of-motion (ROM) data generated by the in vitro test methods of spine simulators with cadavers (SSCs) and finite element models (FEMs) are used alternatively and complementarily for in vitro evaluations. Aim of Review: Our purpose is to compare exemplary segmental ROM data from SSCs and FEMs before and after ball-and-socket total disc replacement (bsTDR) to determine whether the two test methods provide the same data for the same evaluation subjects. Key Scientific Concepts of Review: We performed 70 meta-analyses (MAs) and 20 additional comparative analyses based on data from 21 SSC studies used for 39 MAs and 16 FEM studies used for 31 MAs. Only fifty-nine percent (n = 23/39) of SSC MAs show a restored ROM after bsTDR, whereas in FEM MAs, the ROM is restored in ninety percent (n = 28/31). Among the analyses comparing data from the same spinal segments, motion directions and bsTDR, SSC and FEM data are significantly different in ten percent (n = 2/20). According to our results, data generated by SSCs and FEMs cannot be used as alternative and complementary data without restriction. The quality of the evaluation methods itself as well as potential technical reasons for the discrepant results were not our evaluation target. Further SSC and FEM data should be compared using the same approach. (C) 2020 The Authors. Published by Elsevier B.V. on behalf of Cairo University.
引用
收藏
页码:29 / 41
页数:13
相关论文
共 43 条
[1]   Biomechanical comparison between lumbar disc arthroplasty and fusion [J].
Chen, Shih-Hao ;
Zhong, Zheng-Cheng ;
Chen, Chen-Sheng ;
Chen, Wen-Jer ;
Hung, Chinghua .
MEDICAL ENGINEERING & PHYSICS, 2009, 31 (02) :244-253
[2]   In Situ Contact Analysis of the Prosthesis Components of Prodisc-L in Lumbar Spine Following Total Disc Replacement [J].
Chen, Wen-Ming ;
Park, ChunKun ;
Lee, KwonYong ;
Lee, SungJae .
SPINE, 2009, 34 (20) :E716-E723
[3]   Finite element analysis of a ball-and-socket artificial disc design to suppress excessive loading on facet joints: A comparative study with ProDisc [J].
Choi, Jisoo ;
Shin, Dong-Ah ;
Kim, Sohee .
INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING, 2019, 35 (09)
[4]   Biomechanical Effects of the Geometry of Ball-and-Socket Artificial Disc on Lumbar Spine [J].
Choi, Jisoo ;
Shin, Dong-Ah ;
Kim, Sohee .
SPINE, 2017, 42 (06) :E332-E339
[5]   Biomechanical Effect of Constraint in Lumbar Total Disc Replacement A Study With Finite Element Analysis [J].
Chung, Sang Ki ;
Kim, Young Eun ;
Wang, Kyu-Chang .
SPINE, 2009, 34 (12) :1281-1286
[6]   Biomechanical evaluation of total disc replacement arthroplasty:: An in vitro human cadaveric model [J].
Cunningham, BW ;
Gordon, JD ;
Dmitriev, AE ;
Hu, NB ;
McAfee, PC .
SPINE, 2003, 28 (20) :S110-S117
[7]   Biomechanical Evaluation of the Kinematics of the Cadaver Lumbar Spine Following Disc Replacement With the Prodisc-L Prosthesis [J].
Demetropoulos, Constantine K. ;
Sengupta, Dilip K. ;
Knaub, Mark A. ;
Wiater, Brett P. ;
Abjornson, Celeste ;
Truumees, Eeric ;
Herkowitz, Harry N. .
SPINE, 2010, 35 (01) :26-31
[8]  
Di Mascio V, 2010, J APPL BIOMATER BIOM, V8, P97
[9]  
DiAngelo D., 2014, The Open Spine Journal, V6, P9
[10]   Load-Sharing Between Anterior and Posterior Elements in a Lumbar Motion Segment Implanted With an Artificial Disc [J].
Dooris, Andrew P. ;
Goel, Vijay K. ;
Grosland, Nicole M. ;
Gilbertson, Lars G. ;
Wilder, David G. .
SPINE, 2001, 26 (06) :E122-E129