Investigating the causes of low detectability of pesticides in fruits and vegetables analysed by high-performance liquid chromatography - Time-of-flight

被引:27
作者
Muehlwald, S. [1 ]
Buchner, N. [1 ]
Kroh, L. W. [2 ]
机构
[1] Fed Off Consumer Protect & Food Safety, Mauerstr 39-42, D-10117 Berlin, Germany
[2] Tech Univ Berlin, Inst Food Technol & Food Chem, Dept Food Anal, TIB 4-3-1,Gustav Meyer Allee 25, D-13355 Berlin, Germany
关键词
High-performance liquid chromatography; Time-of-flight mass spectrometry; Pesticides; Matrix effects; In-source collision-induced dissociation; TANDEM MASS-SPECTROMETRY; POSTCOLUMN INFUSION; RESIDUES; FOOD; EXTRACTION; SAMPLES; MS/MS; MS;
D O I
10.1016/j.chroma.2018.02.011
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Because of the high number of possible pesticide residues and their chemical complexity, it is necessary to develop methods which cover a broad range of pesticides. In this work, a qualitative multi-screening method for pesticides was developed by use of HPLC-ESI-Q-TOF. 110 pesticides were chosen for the creation of a personal compound database and library (PCDL). The MassHunter Qualitative Analysis software from Agilent Technologies was used to identify the analytes. The software parameter settings were optimised to produce a low number of false positive as well as false negative results. The method was validated for 78 selected pesticides. However, the validation criteria were not fulfilled for 45 analytes. Due to this result, investigations were started to elucidate reasons for the low detectability. It could be demonstrated that the three main causes of the signal suppression were the co-eluting matrix (matrix effect), the low sensitivity of the analyte in standard solution and the fragmentation of the analyte in the ion source (in-source collision-induced dissociation). In this paper different examples are discussed showing that the impact of these three causes is different for each analyte. For example, it is possible that an analyte with low signal intensity and an intense fragmentation in the ion source is detectable in a difficult matrix, whereas an analyte with a high sensitivity and a low fragmentation is not detectable in a simple matrix. Additionally, it could be shown that in-source fragments are a helpful tool for an unambiguous identification. (C) 2018 Elsevier B.V. All rights reserved.
引用
收藏
页码:37 / 49
页数:13
相关论文
共 23 条
[1]  
Anastassiades M, 2003, J AOAC INT, V86, P412
[2]  
Anderson Jared., 2016, Analytical Separation Science
[3]  
[Anonymous], 2013, AGILENT 6500 SERIES, V2
[4]   Comparison of dissociation of ions in an electrospray source, or a collision cell in tandem mass spectrometry [J].
Buré, C ;
Lange, C .
CURRENT ORGANIC CHEMISTRY, 2003, 7 (15) :1613-1624
[5]  
Federal Ministry of Food and Agriculture, RUCKST PFLANZ GES GE
[6]  
Ferrer C, 2012, COMP ANAL C, V58, P1, DOI 10.1016/B978-0-444-53810-9.00007-9
[7]   Overcoming matrix effects using the dilution approach in multiresidue methods for fruits and vegetables [J].
Ferrer, Carmen ;
Lozano, Ana ;
Agueera, Ana ;
Giron, A. Jimenez ;
Fernandez-Alba, A. R. .
JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (42) :7634-7639
[8]   Identification and quantitation of pesticides in vegetables by liquid chromatography time-of-flight mass spectrometry [J].
Ferrer, I ;
García-Reyes, JF ;
Fernandez-Alba, A .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2005, 24 (07) :671-682
[9]   Multi-residue method for the analysis of 101 pesticides and their degradates in food and water samples by liquid chromatography/time-of-flight mass spectrometry [J].
Ferrer, Imma ;
Thurman, E. Michael .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1175 (01) :24-37
[10]  
Haider K., 2000, UMWANDLUNG ABBAU PFL