Determination of Reference Values for NIST SRM 610-617 Glasses Following ISO Guidelines

被引:1665
作者
Jochum, Klaus Peter [1 ]
Weis, Ulrike [1 ]
Stoll, Brigitte [1 ]
Kuzmin, Dmitry [1 ,2 ]
Yang, Qichao [1 ]
Raczek, Ingrid [1 ]
Jacob, Dorrit E. [3 ,4 ]
Stracke, Andreas [5 ,6 ]
Birbaum, Karin [7 ]
Frick, Daniel A. [7 ]
Guenther, Detlef [7 ]
Enzweiler, Jacinta [8 ]
机构
[1] Max Planck Inst Chem, Abt Biogeochem, D-55020 Mainz, Germany
[2] Russian Acad Sci, Siberian Branch, VS Sobolev Inst Geol & Mineral, Novosibirsk 630090, Russia
[3] Johannes Gutenberg Univ Mainz, Inst Geowissensch, D-55099 Mainz, Germany
[4] Johannes Gutenberg Univ Mainz, Geocycles Forschungszentrum, D-55099 Mainz, Germany
[5] ETH, Inst Geochem & Petrol, CH-8092 Zurich, Switzerland
[6] Univ Munster, Inst Mineral, D-49149 Munster, Germany
[7] ETH, Inorgan Chem Lab, CH-8093 Zurich, Switzerland
[8] Univ Campinas UNICAMP, Inst Geosci, BR-13083970 Campinas, SP, Brazil
关键词
microanalysis; glass reference materials; NIST; characterisation; sample inhomogeneity; microanalyse; verre de reference; caracterisation; inhomogeneite de l'echantillon; INDUCTIVELY-COUPLED PLASMA; LA-ICP-MS; ENVIRONMENTAL REFERENCE MATERIALS; CERTIFIED REFERENCE MATERIALS; ASSOCIATION-OF-GEOANALYSTS; TRACE-ELEMENT ANALYSIS; INTERNATIONAL-ASSOCIATION; MASS SPECTROMETRY; SILICATE GLASS; ID-TIMS;
D O I
10.1111/j.1751-908X.2011.00120.x
中图分类号
P3 [地球物理学]; P59 [地球化学];
学科分类号
0708 ; 070902 ;
摘要
We present new reference values for the NIST SRM 610617 glasses following ISO guidelines and the International Association of Geoanalysts protocol. Uncertainties at the 95% confidence level (CL) have been determined for bulk- and micro-analytical purposes. In contrast to former compilation procedures, this approach delivers data that consider present-day requirements of data quality. New analytical data and the nearly complete data set of the GeoReM database were used for this study. Data quality was checked by the application of the Horwitz function and by a careful investigation of analytical procedures. We have determined quantitatively possible element inhomogeneities using different test portion masses of 1, 0.1 and 0.02 mu g. Although avoiding the rim region of the glass wafers, we found moderate inhomogeneities of several chalcophile/siderophile elements and gross inhomogeneities of Ni, Se, Pd and Pt at small test portion masses. The extent of inhomogeneity was included in the determination of uncertainties. While the new reference values agree with the NIST certified values with the one exception of Mn in SRM 610, they typically differ by as much as 10% from the Pearce (1997) values in current use. In a few cases (P, S, Cl, Ta, Re) the discrepancies are even higher.
引用
收藏
页码:397 / 429
页数:33
相关论文
共 47 条
[1]  
Albarede F., 1995, INTRO GEOCHEMICAL MO
[2]  
[Anonymous], 2009, GEN REQ COMP REF MAT
[3]  
[Anonymous], 2006, REF MAT GEN STAT PRI
[4]  
BONHAM RW, 1988, SPECTROSCOPY INT, V1, P42
[6]   Certificate of analysis of the reference material BRP-1 (Basalt Ribeirao Preto) [J].
Cotta, Aloisio J. B. ;
Enzweiler, Jacinta .
GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2008, 32 (02) :231-235
[7]   Compositional heterogeneity in NIST SRM 610-617 glasses [J].
Eggins, SM ;
Shelley, JMG .
GEOSTANDARDS NEWSLETTER-THE JOURNAL OF GEOSTANDARDS AND GEOANALYSIS, 2002, 26 (03) :269-286
[8]  
FRYER BJ, 1995, CAN MINERAL, V33, P303
[9]   Elemental fractionation during LA-ICP-MS analysis of silicate glasses: implications for matrix-independent standardization [J].
Gaboardi, Mabry ;
Humayun, Munir .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2009, 24 (09) :1188-1197
[10]   Determination of sulfur in fluid inclusions by laser ablation ICP-MS [J].
Guillong, Marcel ;
Latkoczy, Christopher ;
Seo, Jung Hun ;
Guenther, Detlef ;
Heinrich, Christoph A. .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2008, 23 (12) :1581-1589