Fully automatic on-line separation preconcentration system for electrothermal atomic absorption spectrometry: Determination of cadmium and lead in sea-water.

被引:47
作者
Fernandez, FM [1 ]
Stripeikis, JD [1 ]
Tudino, MB [1 ]
Troccoli, OE [1 ]
机构
[1] UNIV BUENOS AIRES,FAC CIENCIAS EXACTAS & NAT,DEPT QUIM INORGAN ANALIT & QUIM FIS,LAB ANAL TRAZAS,RA-1428 BUENOS AIRES,DF,ARGENTINA
关键词
flow injection; electrothermal atomic absorption spectrometry; cadmium; lead; sea-water;
D O I
10.1039/a607598c
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automatic separation preconcentration system coupled to an electrothermal (graphite furnace) atomic absorption spectrometer is described. The preconcentration step is performed on a chelating resin microcolumn (Chelex-100) placed in the injection tip of the autosampler, A time based manifold with two- and three-may solenoid valves commanded by an eight channel microcomputer programmable controller is used for column conditioning, preconcentration and mashing steps; no manual operations are involved, Elution is performed by the programmable graphite furnace autosampler and achieved in only one step, Operations involving complete and partial injection of the eluate into the graphite furnace are also discussed. The system was applied to the determination of Cd and Ph in near shore sea-mater from Patagonia, Argentina. Detection limits of I and 8 ng l(-1) mere obtained for Cd and Pb respectively, Analysis of a certified reference material (CASS-3) showed good agreement with the certified values.
引用
收藏
页码:679 / 684
页数:6
相关论文
共 47 条
[1]   DETERMINATION OF TRACE-METALS IN SEAWATER BY GRAPHITE-FURNACE ATOMIC-ABSORPTION FOLLOWING ONLINE SEPARATION AND PRECONCENTRATION [J].
AZEREDO, LC ;
STURGEON, RE ;
CURTIUS, AJ .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1993, 48 (01) :91-98
[2]  
BACKSTROM K, 1988, ANAL CHEM, V60, P1354
[3]   DESIGN AND EVALUATION OF AN ONLINE MICROCOLUMN PRE-CONCENTRATION TECHNIQUE FOR GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY [J].
BEINROHR, E ;
CAKRT, M ;
RAPTA, M ;
TARAPCI, P .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1989, 335 (08) :1005-1007
[4]   INTERCOMPARISON OF 5 METHODS FOR THE DETERMINATION OF TRACE-METALS IN SEA-WATER [J].
BONIFORTI, R ;
FERRAROLI, R ;
FRIGIERI, P ;
HELTAI, D ;
QUEIRAZZA, G .
ANALYTICA CHIMICA ACTA, 1984, 162 (AUG) :33-46
[5]   TOXICITIES OF TOTAL AND CHELEX-LABILE CADMIUM TO SALMON IN SOLUTIONS OF NATURAL-WATER AND DILUTED SEWAGE WITH POTENTIALLY DIFFERENT CADMIUM COMPLEXING CAPACITIES [J].
BUCKLEY, JA ;
YOSHIDA, GA ;
WELLS, NR ;
AQUINO, RT .
WATER RESEARCH, 1985, 19 (12) :1549-1554
[6]   DIRECT DETERMINATION OF CD, CU AND NI IN THE ELBE RIVER ESTUARY WITH ZEEMAN-AAS AND MATRIX MODIFICATION [J].
CALMANO, W ;
AHLF, W ;
SCHILLING, T .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1986, 323 (08) :865-868
[7]   KINETIC-STUDIES OF METAL SPECIATION USING CHELEX CATION-EXCHANGE RESIN - APPLICATION TO CADMIUM, COPPER, AND LEAD SPECIATION IN RIVER WATER AND SNOW [J].
CHAKRABARTI, CL ;
LU, YJ ;
GREGOIRE, DC ;
BACK, MH ;
SCHROEDER, WH .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1994, 28 (11) :1957-1967
[8]   ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRIC DETERMINATION OF MOLYBDENUM IN WATER, HUMAN HAIR AND HIGH-PURITY REAGENTS WITH FLOW-INJECTION ONLINE COPRECIPITATION PRECONCENTRATION [J].
CHEN, HW ;
XU, SK ;
FANG, ZL .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1995, 10 (08) :533-537
[9]   CHEMICAL SPECIATION OF CU, ZN, PB AND CD IN RAIN WATER [J].
CHENG, JG ;
CHAKRABARTI, CL ;
BACK, MH ;
SCHROEDER, WH .
ANALYTICA CHIMICA ACTA, 1994, 288 (03) :141-156
[10]   CRITICAL STUDY OF APPLICATION OF GRAPHITE-FURNACE NON-FLAME ATOMIC-ABSORPTION SPECTROMETRY TO DETERMINATION OF TRACE BASE METALS IN COMPLEX HEAVY-MATRIX SAMPLE SOLUTIONS [J].
CRUZ, RB ;
VANLOON, JC .
ANALYTICA CHIMICA ACTA, 1974, 72 (02) :231-243