Initial Stability Measurements of Implants Using a New Magnetic Resonance Frequency Analyzer With Titanium Transducers: An Ex Vivo Study

被引:16
作者
Bural, Canan [1 ]
Dayan, Cagatay [2 ]
Geckili, Onur [1 ]
机构
[1] Istanbul Cerrahpasa Univ, Fac Dent, Dept Prosthodont, Istanbul, Turkey
[2] Istanbul Univ, Program Dent Technicians, Istanbul, Turkey
关键词
osseointegration; implantology; experimental design; implant stability; resonance frequency; dental implant; IN-VITRO; DENTAL IMPLANTS; DEVICES; DESIGNS;
D O I
10.1563/aaid-joi-D-19-00126
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The establishment of dental implant stability is mandatory for successful osseointegration. Resonance frequency analysis (RFA) is the most frequently used method for the clinical measurement of implant stability. The purpose of the present study was to evaluate the reliability of the recently developed RF analyzer Penguin RFA and to compare it with the traditional RF analyzer Osstell ISQ. Sixty implants were inserted into fresh steer vertebrae and pelvis. Implant stability was measured using Penguin RFA by its transducers (multipegs) and Osstell ISQ by its transducers (smartpegs). Additionally, stability was measured by multipegs with Osstell ISQ and by smartpegs with Penguin RFA. The intraobserver and interobserver reliability of Penguin RFA were estimated by the intraclass coefficient (ICC). Mean implant stability quotients (ISQs) measured with Osstell ISQ were higher than the ISQs measured with Penguin RFA (P < .05). The intra- and interobserver reliability of Penguin RFA were considered as excellent (ICC > 0.7). For Osstell ISQ, no significance in ISQs was detected between the readings by smartpegs and multipegs (P > .05), while for Penguin RFA ISQs by smartpegs were significantly higher than the ISQs by multipegs (P <.05). The recently developed Penguin RFA is reliable and can be used in clinical practice for the measurement of dental implant stability regardless of the bone type. The multipegs originally manufactured for the Penguin RFA is also compatible with Osstell ISQ.
引用
收藏
页码:35 / 40
页数:6
相关论文
共 29 条
  • [11] Geckili O, 2009, INT J ORAL MAX IMPL, V24, P684
  • [12] Glauser R., 2001, Appl Osseointegration Res, V2, P27
  • [13] Associations of Resonance Frequency Analysis with Dynamic Mechanical Analysis of Dental Implant Systems
    Kim, Do-Gyoon
    Kwon, Hyun-Jung
    Jeong, Yong-Hoon
    Chien, Hua-Hong
    Crance, Samantha
    Agnew, Amanda M.
    Battula, Suneel
    Lee, Jin Whan
    Wen, Hai Bo
    [J]. CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2016, 18 (02) : 332 - 341
  • [14] Lachmann S, 2011, INT J ORAL MAX IMPL, V26, P347
  • [15] Lekholm U., 1985, TISSUE INTEGRATED PR
  • [16] Markovic A, 2011, INT J ORAL MAX IMPL, V26, P718
  • [17] Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis
    Meredith, N
    Alleyne, D
    Cawley, P
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 1996, 7 (03) : 261 - 267
  • [18] Meredith N, 1998, INT J PROSTHODONT, V11, P491
  • [19] Predicting osseointegration by means of implant primary stability - A resonance-frequency analysis study with delayed and immediately loaded ITI SLA implants
    Nedir, R
    Bischof, M
    Szmukler-Moncler, S
    Bernard, JP
    Samson, J
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2004, 15 (05) : 520 - 528
  • [20] Park JC, 2011, IMPLANT DENTISTRY - A RAPIDLY EVOLVING PRACTICE, P111