Initial Stability Measurements of Implants Using a New Magnetic Resonance Frequency Analyzer With Titanium Transducers: An Ex Vivo Study

被引:16
作者
Bural, Canan [1 ]
Dayan, Cagatay [2 ]
Geckili, Onur [1 ]
机构
[1] Istanbul Cerrahpasa Univ, Fac Dent, Dept Prosthodont, Istanbul, Turkey
[2] Istanbul Univ, Program Dent Technicians, Istanbul, Turkey
关键词
osseointegration; implantology; experimental design; implant stability; resonance frequency; dental implant; IN-VITRO; DENTAL IMPLANTS; DEVICES; DESIGNS;
D O I
10.1563/aaid-joi-D-19-00126
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The establishment of dental implant stability is mandatory for successful osseointegration. Resonance frequency analysis (RFA) is the most frequently used method for the clinical measurement of implant stability. The purpose of the present study was to evaluate the reliability of the recently developed RF analyzer Penguin RFA and to compare it with the traditional RF analyzer Osstell ISQ. Sixty implants were inserted into fresh steer vertebrae and pelvis. Implant stability was measured using Penguin RFA by its transducers (multipegs) and Osstell ISQ by its transducers (smartpegs). Additionally, stability was measured by multipegs with Osstell ISQ and by smartpegs with Penguin RFA. The intraobserver and interobserver reliability of Penguin RFA were estimated by the intraclass coefficient (ICC). Mean implant stability quotients (ISQs) measured with Osstell ISQ were higher than the ISQs measured with Penguin RFA (P < .05). The intra- and interobserver reliability of Penguin RFA were considered as excellent (ICC > 0.7). For Osstell ISQ, no significance in ISQs was detected between the readings by smartpegs and multipegs (P > .05), while for Penguin RFA ISQs by smartpegs were significantly higher than the ISQs by multipegs (P <.05). The recently developed Penguin RFA is reliable and can be used in clinical practice for the measurement of dental implant stability regardless of the bone type. The multipegs originally manufactured for the Penguin RFA is also compatible with Osstell ISQ.
引用
收藏
页码:35 / 40
页数:6
相关论文
共 29 条
  • [1] [Anonymous], 1986, ANGELES HERALD 1021, P1
  • [2] Implant stability measurements for implants placed at the time of extraction: A cohort, prospective clinical trial
    Becker, W
    Sennerby, L
    Bedrossian, E
    Becker, BE
    Lucchini, JP
    [J]. JOURNAL OF PERIODONTOLOGY, 2005, 76 (03) : 391 - 397
  • [3] Resonance frequency analysis: Comparing two clinical instruments
    Becker, William
    Hujoel, Philippe
    Becker, Burton E.
    [J]. CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2018, 20 (03) : 308 - 312
  • [4] Comparison of the Primary Stability of Two Implant Designs in Two Different Bone Types: An In Vitro Study
    Bilhan, Hakan
    Bilmenoglu, Caglar
    Urgun, Aliye Ceren
    Ates, Gokcen
    Bural, Canan
    Cilingir, Altug
    Geckili, Onur
    [J]. INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2015, 30 (05) : 1036 - 1040
  • [5] Reliability and validity of the instrumental assessment of implant stability in dry human mandibles
    Brouwers, J. E. I. G.
    Lobbezoo, F.
    Visscher, C. M.
    Wismeijer, D.
    Naeije, M.
    [J]. JOURNAL OF ORAL REHABILITATION, 2009, 36 (04) : 279 - 283
  • [6] Reliability of the Osstell Implant Stability Quotient and Penguin Resonance Frequency Analysis to Evaluate Implant Stability
    Buyukguclu, Gul
    Ozkurt-Kayahan, Zeynep
    Kazazoglu, Ender
    [J]. IMPLANT DENTISTRY, 2018, 27 (04) : 429 - 433
  • [7] Cehreli MC, 2009, INT J ORAL MAX IMPL, V24, P1015
  • [8] Influence of Stepped Osteotomy on Primary Stability of Implants Inserted in Low-Density Bone Sites: An In Vitro Study
    Degidi, Marco
    Daprile, Giuseppe
    Piattelli, Adriano
    [J]. INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2017, 32 (01) : 37 - 41
  • [9] A comparative in vitro evaluation of two different magnetic devices detecting the stability of osseo-integrated implants
    Geckili, O.
    Bilhan, H.
    Cilingir, A.
    Mumcu, E.
    Bural, C.
    [J]. JOURNAL OF PERIODONTAL RESEARCH, 2012, 47 (04) : 508 - 513
  • [10] Comparative Ex Vivo Evaluation of Two Electronic Percussive Testing Devices Measuring the Stability of Dental Implants
    Geckili, Onur
    Bilhan, Hakan
    Cilingir, Altug
    Bilmenoglu, Caglar
    Ates, Gokcen
    Urgun, Aliye Ceren
    Bural, Canan
    [J]. JOURNAL OF PERIODONTOLOGY, 2014, 85 (12) : 1786 - 1791