Characterization of recombinant human monoclonal tissue necrosis factor-α antibody using cation-exchange HPLC and capillary isoelectric focusing

被引:84
作者
Santora, LC
Krull, IS
Grant, K
机构
[1] BASF Biores Corp, Worcester, MA 01605 USA
[2] Northeastern Univ, Dept Chem, Boston, MA 02115 USA
关键词
ion-exchange HPLC; antibody analysis; cIEF; quantitation; TNF;
D O I
10.1006/abio.1999.4275
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Cation-exchange liquid chromatography (CIEX) and capillary isoelectric focusing (cIEF) methods have been developed for the routine analysis of a recombinant, human anti-tumor necrosis factor monoclonal antibody D2E7. Both of these methods can separate heavy-chain C-terminal variants of this antibody. Various enzymatic digestion methods have also been developed for the identification of the antibody C-termini lysine (Lys) variants. A comparison of conventional CIEX-HPLC and cIEF methods has been made for the analysis of antibodies, cIEF can also be used to determine the isoelectric points (pI) of antibody variants based on the use of internal pi standards. Different C-termini Lys variants have been separated and collected from the CIEX column and subsequently analyzed by cLEF and mass spectrometry. (C) 1999 Academic Press.
引用
收藏
页码:98 / 108
页数:11
相关论文
共 29 条
[1]  
BHOWN AS, 1988, PROTEIN PEPTIDE SEQU
[2]  
Bollag D.E., 1996, PROTEIN METHODS
[3]   Thermal stability studies of immunoglobulins using capillary isoelectric focusing and capillary zone electrophoretic methods [J].
Dai, HJ ;
Krull, IS .
JOURNAL OF CHROMATOGRAPHY A, 1998, 807 (01) :121-128
[4]  
Doonan S., 1996, METHODS MOL BIOL, P59
[5]  
GORDON AH, 1969, ELECTROPHORESIS PROT
[6]  
Graf H., 1994, Bioseparation, V4, P7
[7]   PROCESSING OF C-TERMINAL LYSINE AND ARGININE RESIDUES OF PROTEINS ISOLATED FROM MAMMALIAN-CELL CULTURE [J].
HARRIS, RJ .
JOURNAL OF CHROMATOGRAPHY A, 1995, 705 (01) :129-134
[8]   HIGH-PERFORMANCE ELECTROPHORESIS - ELIMINATION OF ELECTROENDOSMOSIS AND SOLUTE ADSORPTION [J].
HJERTEN, S .
JOURNAL OF CHROMATOGRAPHY, 1985, 347 (02) :191-198
[9]   CARRIER-FREE ZONE ELECTROPHORESIS, DISPLACEMENT ELECTROPHORESIS AND ISOELECTRIC-FOCUSING IN A HIGH-PERFORMANCE ELECTROPHORESIS APPARATUS [J].
HJERTEN, S ;
ELENBRING, K ;
KILAR, F ;
LIAO, JL ;
CHEN, AJC ;
SIEBERT, CJ ;
ZHU, MD .
JOURNAL OF CHROMATOGRAPHY, 1987, 403 :47-61
[10]   ADAPTATION OF THE EQUIPMENT FOR HIGH-PERFORMANCE ELECTROPHORESIS TO ISOELECTRIC-FOCUSING [J].
HJERTEN, S ;
ZHU, MD .
JOURNAL OF CHROMATOGRAPHY, 1985, 346 (OCT) :265-270