Liver Stiffness Measurements with MR Elastography: Agreement and Repeatability across Imaging Systems, Field Strengths, and Pulse Sequences

被引:108
作者
Trout, Andrew T. [1 ]
Serai, Suraj [1 ]
Mahley, Alana D. [1 ]
Wang, Hui [3 ]
Zhang, Yue [4 ]
Zhang, Bin [2 ]
Dillman, Jonathan R. [1 ]
机构
[1] Cincinnati Childrens Hosp Med Ctr, Dept Radiol, 3333 Burnet Ave,MLC 5031, Cincinnati, OH 45229 USA
[2] Cincinnati Childrens Hosp Med Ctr, Dept Biostat & Epidemiol, 3333 Burnet Ave,MLC 5031, Cincinnati, OH 45229 USA
[3] Philips Healthcare, Cincinnati, OH USA
[4] Univ Cincinnati, Dept Math Sci, Cincinnati, OH USA
关键词
MAGNETIC-RESONANCE ELASTOGRAPHY; NONINVASIVE ASSESSMENT; ADVANCED FIBROSIS; HEPATIC-FIBROSIS; HEALTHY-VOLUNTEERS; 3.0; T; DISEASE; PERFORMANCE; BIOPSY; REPRODUCIBILITY;
D O I
10.1148/radiol.2016160209
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To prospectively assess agreement and repeatability of magnetic resonance (MR) elastography liver stiffness measurements across imager manufacturers, field strengths, and pulse sequences. Materials and Methods: This prospective cross-sectional study was approved by the institutional review board; informed consent was obtained from all subjects. On the basis of an a priori power calculation, 24 volunteer adult subjects underwent MR elastography with four MR imaging systems (two vendors) and multiple pulse sequences (two-dimensional [2D] gradient-echo [GRE] imaging, 2D spin-echo [SE] echo-planar imaging, and three-dimensional [3D] SE echo-planar imaging). Each sequence was performed twice in each patient with each imaging system. Intraclass correlation coefficients (ICCs) were used to assess agreement and repeatability. P < .05 was considered indicative of a statistically significant difference. Results: Pairwise ICCs were 0.67-0.82 and 0.62-0.83 for agreement between pulse sequences across manufacturers (n = 4) and field strengths (n = 5), respectively. ICCs were 0.45-0.90 for pairwise agreement between sequences while fixing manufacturer and field strength (n = 8). Test-retest repeatability across the various manufacturer, field strength, and pulse sequence combinations (n = 10) was excellent (ICCs, 0.77-0.94). The overall ICC for all manufacturer, field strength, and sequence combinations (n = 10) was 0.68 (95% confidence interval [CI]: 0.55, 0.82). ICC according to field strength was 0.78 (95% CI: 0.67, 0.88) at 1.5 T (n = 5) and 0.64 (95% CI: 0.49, 0.78) at 3.0 T (n = 5). ICCs according to vendor were 0.83 (95% CI: 0.73, 0.91) (n = 4) and 0.65 (95% CI: 0.51, 0.79) (n = 6). Average patient level variance was 0.042 kPa, with a coefficient of variation of 10.7%. Conclusion: MR elastography is a reliable method for assessing liver stiffness, with small amounts of variability between imager manufacturers, field strengths, and pulse sequences. (C) RSNA, 2016
引用
收藏
页码:793 / 804
页数:12
相关论文
共 29 条
[1]   Magnetic resonance elastography (MRE) in assessing hepatic fibrosis: performance in a cohort of patients with histological data [J].
Batheja, Mashal ;
Vargas, Hugo ;
Silva, Annelise M. ;
Walker, Forrest ;
Chang, Yu-Hui ;
De Petris, Giovanni ;
Silva, Alvin C. .
ABDOMINAL IMAGING, 2015, 40 (04) :760-765
[2]   Meta-analysis: ARFI elastography versus transient elastography for the evaluation of liver fibrosis [J].
Bota, Simona ;
Herkner, Harald ;
Sporea, Ioan ;
Salzl, Petra ;
Sirli, Roxana ;
Neghina, Adriana M. ;
Peck-Radosavljevic, Markus .
LIVER INTERNATIONAL, 2013, 33 (08) :1138-1147
[3]   Non-invasive assessment of liver fibrosis with impulse elastography: Comparison of Supersonic Shear Imaging with ARFI and FibroScan [J].
Cassinotto, Christophe ;
Lapuyade, Bruno ;
Mouries, Amaury ;
Hiriart, Jean-Baptiste ;
Vergniol, Julien ;
Gaye, Delphine ;
Castain, Claire ;
Le Bail, Brigitte ;
Chermak, Faiza ;
Foucher, Juliette ;
Laurent, Francois ;
Montaudon, Michel ;
De Ledinghen, Victor .
JOURNAL OF HEPATOLOGY, 2014, 61 (03) :550-557
[4]   Comparative diagnostic accuracy of magnetic resonance elastography vs. eight clinical prediction rules for non-invasive diagnosis of advanced fibrosis in biopsy-proven non-alcoholic fatty liver disease: a prospective study [J].
Cui, J. ;
Ang, B. ;
Haufe, W. ;
Hernandez, C. ;
Verna, E. C. ;
Sirlin, C. B. ;
Loomba, R. .
ALIMENTARY PHARMACOLOGY & THERAPEUTICS, 2015, 41 (12) :1271-1280
[5]   Magnetic resonance elastography can discriminate normal vs. abnormal liver biopsy in candidates for live liver donation [J].
Gallegos-Orozco, Juan F. ;
Silva, Alvin C. ;
Batheja, Mashal J. ;
Chang, Yu-Hui ;
Hansen, Kathleen L. ;
Lam-Himlin, Dora ;
De Petris, Giovanni ;
Aqel, Bashar A. ;
Byrne, Thomas J. ;
Carey, Elizabeth J. ;
Douglas, David D. ;
Mulligan, David C. ;
Silva, Annelise M. ;
Rakela, Jorge ;
Vargas, Hugo E. .
ABDOMINAL IMAGING, 2015, 40 (04) :795-802
[6]   RSNA/QIBA: Shear wave speed as a biomarker for liver fibrosis staging [J].
Hall, Timothy J. ;
Milkowski, Andy ;
Garra, Brian ;
Carson, Paul ;
Palmeri, Mark ;
Nightingale, Kathy ;
Lynch, Ted ;
Alturki, Abdullah ;
Andre, Michael ;
Audiere, Stephane ;
Bamber, Jeffery ;
Barr, Richard ;
Bercoff, Jeremy ;
Bercoff, Jessica ;
Bernal, Miguel ;
Brum, Javier ;
Chan, Huan Wee ;
Chen, Shigao ;
Cohen-Bacrie, Claude ;
Couade, Mathieu ;
Daniels, Allison ;
DeWall, Ryan ;
Dillman, Jonathan ;
Ehman, Richard ;
Franchi-Abella, S. F. ;
Fromageau, Jeremie ;
Gennisson, Jean-Luc ;
Henry, Jean Pierre ;
Ivancevich, Nikolas ;
Kalin, Jan ;
Kohn, Sarah ;
Kugel, Jennifer ;
Lee, Ken ;
Liu, N. L. ;
Loupas, Thanasis ;
Mazernik, Joan ;
McAleavey, Stephen ;
Miette, Veronique ;
Metz, Stephen ;
Morel, B. M. ;
Nelson, Thomas ;
Nordberg, Eric ;
Oudry, Jennifer ;
Padwal, Monali ;
Rouze, Ned ;
Samir, Anthony ;
Sandrin, Laurent ;
Schaccitti, Janet ;
Schmitt, Cedric ;
Shamdasani, Vijay .
2013 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2013, :397-400
[7]  
Hallgren Kevin A, 2012, Tutor Quant Methods Psychol, V8, P23
[8]   Magnetic Resonance Elastography in the Liver at 3 Tesla Using a Second Harmonic Approach [J].
Herzka, D. A. ;
Kotys, M. S. ;
Sinkus, R. ;
Pettigrew, R. I. ;
Gharib, A. M. .
MAGNETIC RESONANCE IN MEDICINE, 2009, 62 (02) :284-291
[9]   Repeatability of Magnetic Resonance Elastography for Quantification of Hepatic Stiffness [J].
Hines, Catherine D. G. ;
Bley, Thorsten A. ;
Lindstrom, Mary J. ;
Reeder, Scott B. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2010, 31 (03) :725-731
[10]   Advanced Fibrosis in Nonalcoholic Fatty Liver Disease: Noninvasive Assessment with MR Elastography [J].
Kim, Donghee ;
Kim, W. Ray ;
Talwalkar, Jayant A. ;
Kim, Hwa Jung ;
Ehman, Richard L. .
RADIOLOGY, 2013, 268 (02) :411-419