The appearance of porous bone under fixation plates during fracture healing, attributed to disturbance of blood supply by the plate, has led to new plate designs with reduced plate to bone contact. The fixation stability afforded by these devices, in comparison to implants commonly used for fixation, is not well known. Therefore, the construct stiffnesses of osteotomized synthetic tibiae fixed with dynamic compression plates, external fixators, or two configurations of noncontact plates were compared in axial compression, bending, and torsion with and without cortical contact at the osteotomy site. The results of this study show that (1) the noncontact plated constructs achieve comparable fixation rigidity to constructs using dynamic compression plates or external fixators, if applied at a distance no greater than 5 mm from the surface of the tibia, and (2) the fixation rigidity of the noncontact plate decreases as the distance between plate and bone surface increases.
引用
收藏
页码:5 / 9
页数:5
相关论文
共 12 条
[1]   INTERNAL CONTACT SPLINT [J].
EGGERS, GWN .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1948, 30-A (01) :40-&
[2]  
KLAUE K, 1987, AUG P INT SOC FRAC R
[3]   COMPARISON OF THE STABILITY OF PRESS-FIT HIP-PROSTHESIS FEMORAL STEMS USING A SYNTHETIC MODEL FEMUR [J].
MCKELLOP, H ;
EBRAMZADEH, E ;
NIEDERER, PG ;
SARMIENTO, A .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1991, 9 (02) :297-305
[4]  
MULLER ME, 1991, AO MANUAL INTERNAL F, P367
[5]  
PERREN SM, 1991, INJURY, V22, P1
[6]  
PERREN SM, 1988, CLIN ORTHOP RELAT R, V232, P139
[7]  
RAMOTOWSKI W, 1991, CLIN ORTHOP RELAT R, V272, P67
[8]  
SWIONTKOWSKI MF, 1991, T ORTHOP RES SOC, V16, P66
[9]   COMPARISON OF THE DEFORMATION RESPONSE OF SYNTHETIC AND CADAVERIC FEMORA DURING SIMULATED ONE-LEGGED STANCE [J].
SZIVEK, JA ;
GEALER, RL .
JOURNAL OF APPLIED BIOMATERIALS, 1991, 2 (04) :277-280
[10]   THE USE OF SEMIRIGID CARBON-FIBER-REINFORCED PLASTIC PLATES FOR FIXATION OF HUMAN FRACTURES - RESULTS OF PRELIMINARY TRIALS [J].
TAYTON, K ;
JOHNSONNURSE, C ;
MCKIBBIN, B ;
BRADLEY, J ;
HASTINGS, G .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1982, 64 (01) :105-111