A Monte Carlo based analytic model of the in-room neutron ambient dose equivalent for a Mevion gantry-mounted passively scattered proton system

被引:4
作者
Baradaran-Ghahfarokhi, Milad [1 ]
Reynoso, Francisco [1 ]
Darafsheh, Arash [1 ]
Sun, Baozhou [1 ]
Prusator, Michael T. [1 ]
Mutic, Sasa [1 ]
Zhao, Tianyu [1 ]
机构
[1] Washington Univ, Dept Radiat Oncol, St Louis, MO 63110 USA
关键词
neutron ambient dose equivalent; Monte Carlo simulation; analytic model; passively scattered proton; shielding; RADIATION; RADIOTHERAPY; PHOTON;
D O I
10.1088/1361-6498/abaff4
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The goal of this study was to develop a Monte Carlo (MC)-based analytical model that can predict the in-room ambient dose equivalent from a Mevion gantry-mounted passively scattered proton system. The Mevion S250 and treatment vault were simulated using the MCNPX MC code. The results of the in-room neutron dose measurements, using an FHT 762 WENDI-II detector, were employed to benchmark the MC-derived values. After tuning the MCNPX MC code, for the same beam delivery parameters, the code was used to calculate the neutron spectra and ambient dose equivalent in the vault and at varying angles from the isocenter. Then, based on the calculations, an analytical model was reconstructed and data were fitted to derive the model parameters at 95% confidence intervals (CI). The MCNPX codes were tuned to within about 19% of the measured values for most of the measurements in the vault. For the maze, up to 0.08 mSv Gy(-1)discrepancies were found between the experimental measurements and MCNPX calculated results. The analytical model showed up to 18% discrepancy for distances between 100 and 600 cm from the isocenter compared to the MC calculations. The model may underestimate the neutron ambient dose equivalent up to 21% for distances less than 100 cm from the isocenter. The proposed analytical model can be used to estimate the contribution of the secondary neutron dose from the Mevion S250 for the design of local shielding inside the proton therapy treatment vault.
引用
收藏
页码:980 / 996
页数:17
相关论文
共 28 条
  • [11] Measured Neutron Spectra and Dose Equivalents From a Mevion Single-Room, Passively Scattered Proton System Used for Craniospinal Irradiation
    Howell, Rebecca M.
    Burgett, Eric A.
    Isaacs, Daniel
    Hedrick, Samantha G. Price
    Reilly, Michael P.
    Rankine, Leith J.
    Grantham, Kevin K.
    Perkins, Stephanie
    Klein, Eric E.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 95 (01): : 249 - 257
  • [12] International Commission on Radiological Protection, 1996, Ann. ICRP, V26, P1, DOI DOI 10.1016/S0116
  • [13] Proton Facility Economics: Single-Room Centers
    Kerstiens, John
    Johnstone, Gregory P.
    Johnstone, Peter A. S.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2018, 15 (12) : 1704 - 1708
  • [14] Out-of-field photon and neutron dose equivalents from step-and-shoot intensity-modulated radiation therapy
    Kry, SF
    Salehpour, M
    Followill, DS
    Stovall, M
    Kuban, DA
    White, RA
    Rosen, II
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (04): : 1204 - 1216
  • [15] M Thermo Scientific Co, 2005, FHT 762 WEND 2 WID E
  • [16] The physics of proton therapy
    Newhauser, Wayne D.
    Zhang, Rui
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (08) : R155 - R209
  • [17] WENDI: An improved neutron REM meter
    Olsher, RH
    Hsu, HH
    Beverding, A
    Kleck, JH
    Casson, WH
    Vasilik, DG
    Devine, RT
    [J]. HEALTH PHYSICS, 2000, 79 (02): : 170 - 181
  • [18] Accurate Monte Carlo simulations for nozzle design, commissioning and quality assurance for a proton radiation therapy facility
    Paganetti, H
    Jiang, H
    Lee, SY
    Kooy, HM
    [J]. MEDICAL PHYSICS, 2004, 31 (07) : 2107 - 2118
  • [19] Calculations of neutron dose equivalent exposures from range-modulated proton therapy beams
    Polf, JC
    Newhauser, WD
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (16) : 3859 - 3873
  • [20] TOPAS Simulation of the Mevion S250 compact proton therapy unit
    Prusator, Michael
    Ahmad, Salahuddin
    Chen, Yong
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (03): : 88 - 95