Experimental determination of dosimetry parameters for Sinko 125I seed source using a modified polystyrene phantom

被引:2
作者
Duan, Yuting [2 ]
Zhang, Menglong [1 ]
Wang, Gang [1 ]
Du, Liqing [3 ]
机构
[1] Shandong Med Imaging Res Inst, Jinan, Peoples R China
[2] Liaocheng Peoples Hosp, Liaocheng, Peoples R China
[3] Chinese Acad Med Sci, Inst Radiat Med, Tianjin, Peoples R China
关键词
Brachytherapy; I-125; seed; TG43; parameters; Modified phantom; BRACHYTHERAPY; RECOMMENDATIONS; LIF;
D O I
10.1007/s13246-012-0152-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Successful treatment for permanent implant brachytherapy is based on accurate measurement of dosimetry parameters for the seed sources. Literature describes the application of various types of phantom to determine the AAPM TG-43 dosimetry parameters for permanent implant seeds. Previously we created a new type of phantom used to measure the dosimetry parameters of a high dose-rate Ir-192 source. In this study, we modified the phantom to suit to a common type of I-125 seed source (Sinko BT-125-1). The dose-rate constant, radial dose function and anisotropy function of this source were measured in detail and compared with the published values of other similar in-design I-125 seed sources. The experimental results exhibit fairly small measurement uncertainties and good self-consistency. The modified phantom is demonstrated on the measurement of dosimetry parameters for the Sinko BT-125-1 I-125 seed, however, it could easily be used for similar measurements of other permanent implantation seed sources.
引用
收藏
页码:291 / 296
页数:6
相关论文
共 16 条
  • [1] Thermoluminescent dosimetry of the selectSeed 125I interstitial brachytherapy seed
    Anagnostopoulos, G
    Baltas, D
    Karaiskos, P
    Sandilos, P
    Papagiannis, P
    Sakelliou, L
    [J]. MEDICAL PHYSICS, 2002, 29 (05) : 709 - 716
  • [2] Carey E, 2008, BRACHYTHERAPY, V7, P164
  • [3] Chadwick E, 2011, BRACHYTHERAPY S1, V10, pS96
  • [4] The low energy X-ray response of the LiF:Mg:Cu:P thermoluminescent dosemeter:: a comparison with LiF:Mg:Ti
    Edwards, CR
    Mountford, PJ
    Green, S
    Palethorpe, JE
    Moloney, AJ
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2005, 78 (930) : 543 - 547
  • [5] Experimental and Monte Carlo determination of the TG-43 dosimetric parameters for the model 9011 THINSeed™ brachytherapy source
    Kennedy, R. M.
    Davis, S. D.
    Micka, J. A.
    DeWerd, L. A.
    [J]. MEDICAL PHYSICS, 2010, 37 (04) : 1681 - 1688
  • [6] Energy response of LiF and Mg2SiO4 TLDs to 10-150 KeV monoenergetic photons
    Konnai, A
    Nariyama, N
    Ohnishi, S
    Odano, N
    [J]. RADIATION PROTECTION DOSIMETRY, 2005, 115 (1-4) : 334 - 336
  • [7] PHYSICAL DOSIMETRY OF I-125 SEEDS OF A NEW DESIGN FOR INTERSTITIAL IMPLANT
    LING, CC
    YORKE, ED
    SPIRO, IJ
    KUBIATOWICZ, D
    BENNETT, D
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1983, 9 (11): : 1747 - 1752
  • [8] Loftus GP, 1984, J RES NBS, V89, P295
  • [9] Experimental determination of dosimetric characteristics of Best® 125I brachytherapy source
    Meigooni, AS
    Gearheart, DM
    Sowards, K
    [J]. MEDICAL PHYSICS, 2000, 27 (09) : 2168 - 2173
  • [10] DOSIMETRY OF INTERSTITIAL BRACHYTHERAPY SOURCES - RECOMMENDATIONS OF THE AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO 43
    NATH, R
    ANDERSON, LL
    LUXTON, G
    WEAVER, KA
    WILLIAMSON, JF
    MEIGOONI, AS
    [J]. MEDICAL PHYSICS, 1995, 22 (02) : 209 - 234