A full quantitative analysis of 18 MV photon beam from 2100 C/D-Varian clinical linear accelerator with and without flattening filter

被引:2
作者
Mahdavi, S. R. [1 ]
Ay, M. R. [2 ]
Zabihzadeh, M. [3 ,4 ,5 ]
Allahverdi, M. [2 ]
Shahriari, M. [6 ]
Hoseini-Ghahfarokhi, M. [7 ]
机构
[1] Iran Univ Med Sci, Dept Med Phys, Tehran, Iran
[2] Univ Tehran Med Sci, Dept Med Phys & Biomed Engn, Tehran, Iran
[3] Ahvaz Jundishapur Univ Med Sci, Canc Environm & Petr Pollutants Res Ctr, Ahvaz, Iran
[4] Ahvaz Jundishapur Univ Med Sci, Golestan Hosp, Fac Med, Dept Med Phys, Ahvaz, Iran
[5] Ahvaz Jundishapur Univ Med Sci, Golestan Hosp, Fac Med, Dept Clin Oncol, Ahvaz, Iran
[6] Shahid Beheshti Univ, Dept Nucl Engn, Tehran, Iran
[7] Kermanshah Univ Med Sci, Dept Radiol & Nucl Med, Kermanshah, Iran
来源
INTERNATIONAL JOURNAL OF RADIATION RESEARCH | 2019年 / 17卷 / 01期
关键词
Dosimetry; fattening filter; linear accelerator; Monte Carlo simulation; radiotherapy; MONTE-CARLO; SPECTRA;
D O I
10.18869/acadpub.ijrr.17.1.137
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: During intensity modulated radiation therapy (IMRT) technique, theoretically, presence of flattening filter (FF) across the beamline of clinical linear accelerator (linac) is not essential to obtain uniform lateral profiles due to intensity modulation of photon beams by multileaf collimators (MLCs). The aim of this study was to investigate the dosimetrical properties of 18 MV photon beam-Varian linac with and without FF. Materials and Methods: All dose measurements were performed on 18 MV, FF mode-Varian 2100C/D linac. The FF and flattening filter free (FFF) modes of linac were modeled by MCNPX 2.4. code. The photon and contaminant electrons spectra, dose rate, present depth doses (PDD), lateral dose profiles, total and collimator scatter factors and out of field doses were calculated and compared with and without FF. Results: Removing the FF increased the photon and contaminant electron fluences by factors of 5.48 and 3.94 for a 5 x 5 cm2 field size, respectively. The surface dose increased up to 155%. The flatness of dose profile was disturbed and deteriorated with increase of field size. Despite the dependence of the total scattering factor on field size, the variation of collimator scattering factors was neglected. The out-of-field dose decreased about 81.5 % for a 5 x 5 cm2 field size. Conclusion: Removing FF from the linac head increases the dose rate and decreases the out-of-field dose, but the increased skin dose and deteriorated flatness of lateral dose profile are the main disadvantages of the FFF mode.
引用
收藏
页码:137 / 146
页数:10
相关论文
共 32 条
  • [1] Allahverdi M, 2011, IRAN J RADIAT RES, V9, P15
  • [2] The characterization of unflattened photon beams from a 6 MV linear accelerator
    Cashmore, Jason
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (07) : 1933 - 1946
  • [3] Surface dose variations in 6 and 10 MV flattened and flattening filter-free (FFF) photon beams
    Cashmore, Jason
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (05): : 293 - 307
  • [4] A Monte Carlo study of a flattening filter-free linear accelerator verified with measurements
    Dalaryd, Marten
    Kragl, Gabriele
    Ceberg, Crister
    Georg, Dietmar
    McClean, Brendan
    af Wetterstedt, Sacha
    Wieslander, Elinore
    Knoos, Tommy
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (23) : 7333 - 7344
  • [5] Day M J, 1996, BJR Suppl, V25, P138
  • [6] Beam properties and stability of a flattening-filter free 7 MV beam-An overview
    Dzierma, Yvonne
    Licht, Norbert
    Nuesken, Frank
    Ruebe, Christian
    [J]. MEDICAL PHYSICS, 2012, 39 (05) : 2595 - 2602
  • [7] A multiple source model for 6 MV photon beam dose calculations using Monte Carlo
    Fix, MK
    Stampanoni, M
    Manser, P
    Born, EJ
    Mini, R
    Rüegsegger, P
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (05) : 1407 - 1427
  • [8] Delivery time comparison for intensity-modulated radiation therapy with/without flattening filter: a planning study
    Fu, WH
    Dai, JR
    Hu, YM
    Han, DS
    Song, YX
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (08) : 1535 - 1547
  • [9] Theoretical and experimental study of spectral distortions at the output of an accelerator for medical use
    Guillerminet, C
    Duverger, E
    Gschwind, R
    Makovicka, L
    Monnier, A
    Hamon, F
    David, C
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2001, 61 (3-6) : 611 - 613
  • [10] Monte Carlo comparison of superficial dose between flattening filter free and flattened beams
    Javedan, Khosrow
    Feygelman, Vladimir
    Zhang, Ray R.
    Moros, Eduardo G.
    Correa, Candace R.
    Trotti, Andy
    Li, Weiqi
    Zhang, Geoffrey G.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (04): : 503 - 508