Biomechanical Comparison of Interference Screws and Combination Screw and Sheath Devices for Soft Tissue Anterior Cruciate Ligament Reconstruction on the Tibial Side

被引:61
作者
Aga, Cathrine [1 ]
Rasmussen, Matthew T. [1 ]
Smith, Sean D. [1 ]
Jansson, Kyle S. [1 ]
LaPrade, Robert F. [1 ]
Engebretsen, Lars [1 ]
Wijdicks, Coen A. [1 ]
机构
[1] Steadman Philippon Res Inst, Dept Biomed Engn, Vail, CO 81657 USA
关键词
anterior cruciate ligament; ACL reconstruction; soft tissue graft; tibial fixation; hamstring graft; interference screw; sheath; intratunnel fixation; graft fixation; biomechanical testing; TENDON GRAFT FIXATION; PATELLAR TENDON; STRUCTURAL-PROPERTIES; ACL RECONSTRUCTION; PULLOUT STRENGTH; HAMSTRING GRAFTS; TUNNEL DILATION; PORCINE TIBIA; LENGTH; FORCE;
D O I
10.1177/0363546512474968
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The tibial fixation site has been reported to be the weakest point in anterior cruciate ligament (ACL) reconstructions. Numerous interference screws and combination screw and sheath devices are available for soft tissue fixation, and a biomechanical comparison of these devices is necessary. Hypothesis: Combination screw and sheath devices would provide superior soft tissue fixation properties compared with interference screws in a porcine model. Study Design: Controlled laboratory study. Methods: Eight different intratunnel tibial soft tissue fixation devices were biomechanically tested in a porcine model with bovine tendons, with 10 specimens per group. The soft tissue fixation devices included 3 interference screws-the Bio-Interference Screw, BIOSURE PK, and RCI Screw-and 5 combination screw and sheath devices (combination devices)-the AperFix II, BIOSURE SYNC, ExoShape, GraftBolt, and INTRAFIX. The specimens were subjected to cyclic (1000 cycles, 50-250 N, 0.5 Hz) and pull-to-failure loading (50 mm/min) with a dynamic tensile testing machine. Ultimate failure load (N), cyclic displacement (mm), pull-out stiffness (N/mm), displacement at failure (mm), load at 3 mm displacement (N), and mechanism of failure were recorded. Results: The ultimate failure loads were highest for the GraftBolt (1136 +/- 115.6 N), followed by the INTRAFIX (1127 +/- 155.0 N), AperFix II (1122 +/- 182.9 N), BIOSURE PK (990.8 +/- 182.1 N), Bio-Interference Screw (973.3 +/- 95.82 N), BIOSURE SYNC (829.5 +/- 172.4 N), RCI Screw (817.7 +/- 113.9 N), and ExoShape (814.7 +/- 178.8 N). The AperFix II, GraftBolt, and INTRAFIX devices were significantly stronger than the BIOSURE SYNC, RCI Screw, and ExoShape. Although the 3 strongest devices were combination screw and sheath devices, no significant differences were observed between the ultimate failure strengths of the screw and combination devices when compared as groups. The least amount of cyclic displacement after 1000 cycles was observed for the GraftBolt (1.38 +/- 0.27 mm), followed by the AperFix II (1.58 +/- 0.21 mm), Bio-Interference Screw (1.61 +/- 0.22 mm), INTRAFIX (1.63 +/- 0.15 mm), ExoShape (1.68 +/- 0.30 mm), BIOSURE PK (1.72 +/- 0.29 mm), BIOSURE SYNC (1.92 +/- 0.59 mm), and RCI Screw (1.97 +/- 0.39 mm). The GraftBolt allowed significantly less displacement than did the BIOSURE SYNC and RCI Screw. Similarly, no significant differences were observed between the cyclic displacements of the screws and combination devices when compared as groups. Conclusion: The combination screw and sheath devices did not provide superior soft tissue fixation properties compared with the interference screws alone in a porcine model. Although the highest ultimate failure loads and least amounts of cyclic displacement were observed for combination devices, group comparisons of screw and combination devices did not result in any significant differences for ultimate failure load and cyclic displacement. Clinical Relevance: It is important to consider that these results represent device performance in an in vitro animal model and are not directly transferrable to an in vivo clinical situation. The combination of a sheath and screw did not consistently result in improved fixation characteristics compared with interference screw fixation.
引用
收藏
页码:841 / 848
页数:8
相关论文
共 35 条
  • [1] Foam-reinforced elderly human tibia approximates young human tibia better than porcine tibia - A study of the structural properties of three soft tissue fixation devices
    Bailey, SB
    Grover, DM
    Howell, SM
    Hull, ML
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2004, 32 (03) : 755 - 764
  • [2] Graft fixation in cruciate ligament reconstruction
    Brand, J
    Weiler, A
    Caborn, DNM
    Brown, CH
    Johnson, DL
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2000, 28 (05) : 761 - 774
  • [3] A biomechanical comparison of initial soft tissue tibial fixation devices
    Caborn, DNM
    Brand, JC
    Nyland, J
    Kocabey, Y
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2004, 32 (04) : 956 - 961
  • [4] Cain EL, 2005, ORTHOPEDICS, V28, P779
  • [5] Charlick D A, 2000, Arthroscopy, V16, pE20, DOI 10.1053/jars.2000.17983
  • [6] Cohen J., 1988, Statistical power analysis for the behavioral sciences, VSecond
  • [7] Comparison of viscoelastic, structural, and material properties of double-looped anterior cruciate ligament grafts made from bovine digital extensor and human hamstring tendons
    Donahue, TLH
    Gregersen, C
    Hull, ML
    Howell, SM
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2001, 123 (02): : 162 - 169
  • [8] Soft tissue tendon graft fixation in serially dilated or extraction-drilled tibial tunnels - A porcine model study using high-resolution quantitative computerized tomography
    Dunkin, Brad S.
    Nyland, John
    Duffee, Andrew R.
    Brunelli, Jeffrey A.
    Burden, Robert
    Caborn, David
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2007, 35 (03) : 448 - 457
  • [9] Natural history of a hamstring tendon autograft used for anterior cruciate ligament reconstruction in a sheep model
    Goradia, VK
    Rochat, MC
    Kida, M
    Grana, WA
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2000, 28 (01) : 40 - 46
  • [10] The effect of bone compaction on early fixation of porous-coated implants
    Green, JR
    Nemzek, JA
    Arnoczky, SP
    Johnson, LL
    Balas, MS
    [J]. JOURNAL OF ARTHROPLASTY, 1999, 14 (01) : 91 - 97