Computer-aided characterization of mammographic masses: Accuracy of mass segmentation and its effects on characterization

被引:139
作者
Sahiner, B [1 ]
Petrick, N [1 ]
Chan, HP [1 ]
Hadjiiski, LM [1 ]
Paramagul, C [1 ]
Helvie, MA [1 ]
Gurcan, MN [1 ]
机构
[1] Univ Michigan, Dept Radiol, Ann Arbor, MI 48109 USA
关键词
breast masses; classification; computer-aided diagnosis; mammography; segmentation;
D O I
10.1109/42.974922
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Mass segmentation is used as the first step in many computer-aided diagnosis (CAD) systems for classification of breast masses as malignant or benign. The goal of this paper was to study the accuracy of an automated mass segmentation method developed in our laboratory, and to investigate the effect of the segmentation stage on the overall classification accuracy. The automated segmentation method was quantitatively compared with manual segmentation by two expert radiologists (111 and 112) using three similarity or distance measures on a data set of 100 masses. The area overlap measures between R1 and R2, the computer and R1, and the computer and R2 were 0.76 +/- 0.13, 0.74 +/- 0.11, and 0.74 +/- 0.13, respectively. The interobserver difference in these measures between the two radiologists was compared with the corresponding differences between the computer and the radiologists. Using three similarity measures and data from two radiologists, a total of six statistical tests were performed. The difference between the computer and the radiologist segmentation was significantly larger than the interobserver variability in only one test. Two sets of texture, morphological, and spiculation features, one based on the computer segmentation, and the other based on radiologist segmentation, were extracted from a data set of 249 films from 102 patients. A classifier based on stepwise feature selection and linear discriminant analysis was trained and tested using the two feature sets. The leave-one-case-out method was used for data sampling. For case-based classification, the area A(x) under the receiver operating characteristic (ROC) curve was 0.89 and 0.88 for the feature sets based on the radiologist segmentation and computer segmentation, respectively. The difference between the two ROC curves was not statistically significant.
引用
收藏
页码:1275 / 1284
页数:10
相关论文
共 40 条
  • [1] American College of Radiology, 1998, AM COLL RAD BREAST I
  • [2] [Anonymous], 1998, Applied regression analysis, DOI 10.1002/9781118625590
  • [3] Classifying mammographic mass shapes using the wavelet transform modulus-maxima method
    Bruce, LM
    Adhami, RR
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1999, 18 (12) : 1170 - 1177
  • [4] Computerized analysis of mammographic microcalcifications in morphological and texture feature spaces
    Chan, HP
    Sahiner, B
    Lam, KL
    Petrick, N
    Helvie, MA
    Goodsitt, MM
    Adler, DD
    [J]. MEDICAL PHYSICS, 1998, 25 (10) : 2007 - 2019
  • [5] COMPUTER-AIDED CLASSIFICATION OF MAMMOGRAPHIC MASSES AND NORMAL TISSUE - LINEAR DISCRIMINANT-ANALYSIS IN TEXTURE FEATURE SPACE
    CHAN, HP
    WEI, DT
    HELVIE, MA
    SAHINER, B
    ADLER, DD
    GOODSITT, MM
    PETRICK, N
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (05) : 857 - 876
  • [6] Improvement of radiologists' characterization of mammographic masses by using computer-aided diagnosis: An ROC study
    Chan, HP
    Sahiner, B
    Helvie, MA
    Petrick, N
    Roubidoux, MA
    Wilson, TE
    Adler, DD
    Paramagul, C
    Newman, JS
    Sanjay-Gopal, S
    [J]. RADIOLOGY, 1999, 212 (03) : 817 - 827
  • [7] CHAN HP, 2000, ARTIF INTELL, P179
  • [8] ON ACTIVE CONTOUR MODELS AND BALLOONS
    COHEN, LD
    [J]. CVGIP-IMAGE UNDERSTANDING, 1991, 53 (02): : 211 - 218
  • [9] MATCHED-FILTER IDENTIFICATION OF LEFT-VENTRICULAR ENDOCARDIAL BORDERS IN TRANSESOPHAGEAL ECHOCARDIOGRAMS
    DETMER, PR
    BASHEIN, G
    MARTIN, RW
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1990, 9 (04) : 396 - 404
  • [10] Galloway MM., 1975, Comput. Graphics Image Process, V4, P172, DOI [DOI 10.1016/S0146-664X(75)80008-6, 10.1016/S0146-664X(75)80008-6]