Measurement of blood access flow rate during hemodialysis from conductivity dialysance

被引:51
作者
Gotch, FA
Buyaki, R
Panlilio, F
Folden, T
机构
[1] Davies Med Ctr, Dept Dialysis Treatment & Res, San Francisco, CA 94114 USA
[2] Fresenius Med Care, Dept Clin Res, Walnut Creek, CA USA
关键词
D O I
10.1097/00002480-199905000-00007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An on line clearance monitor automated for measurements of conductivity dialysance (Dcn) was used to measure blood access flow rate (cnQac) during hemodialysis. From mathematical analysis of transport, it was shown that cnQace = [(Dcn*Decn)/(Dcn - Decn)] [1/bwf], where Dcn is measured with standard cocurrent flow of dialyzer blood (Qb) and Qac; Decn is measured with countercurrent Qb/Qac; and bwf is fractional blood water content. An identical equation was derived to measure Qac from urea dialysance (uQac, Du, Deu). In vitro studies showed excellent correlation between volumetric measurement of Qac (vQac) and cnQac, r = 0.98, n = 29, and between uQac and cnQac, r = 0.97, n = 28. In vivo studies showed comparable agreement between uQac and cnQacc, r = 0.97, n = 14. In two of the patients studied, there was unsuspected severe midgraft stenosis (no recirculation in cocurrent flow) with a Qac of 89 and 202 ml/min disclosed by both cnQac and uQac measurements. Mathematical analysis also showed that when Qb is greater than Qac and there is recirculation in cocurrent flow, the above equations always return the value Qac = Qb. An equation was derived to calculate cnQac without reversal of blood lines in this case, using Dcn calculated from the dialyzer transport coefficient and flow rates.
引用
收藏
页码:139 / 146
页数:8
相关论文
共 12 条
[1]  
Besarab A, 1996, SEMIN DIALYSIS, V9, pS21
[2]   Detection of access strictures and outlet stenoses in vascular accesses - Which test is best? [J].
Besarab, A ;
Lubkowski, T ;
Frinak, S ;
Ramanathan, S ;
Escobar, F .
ASAIO JOURNAL, 1997, 43 (05) :M543-M547
[3]  
DEPNER TA, 1995, ASAIO J, V41, pM745, DOI 10.1097/00002480-199507000-00112
[4]   THEORY AND VALIDATION OF ACCESS FLOW MEASUREMENT BY DILUTION TECHNIQUE DURING HEMODIALYSIS [J].
KRIVITSKI, NM .
KIDNEY INTERNATIONAL, 1995, 48 (01) :244-250
[5]   Hemodialysis access blood flow rates can be measured by a differential conductivity technique and are predictive of access clotting [J].
Lindsay, RM ;
Blake, PG ;
Malek, P ;
Posen, G ;
Martin, B ;
Bradfield, E .
AMERICAN JOURNAL OF KIDNEY DISEASES, 1997, 30 (04) :475-482
[6]   Accuracy and precision of access recirculation measurements by the hemodynamic recirculation monitor [J].
Lindsay, RM ;
Blake, PG ;
Malek, P ;
Posen, G ;
Martin, B ;
Bradfield, E .
AMERICAN JOURNAL OF KIDNEY DISEASES, 1998, 31 (02) :242-249
[7]  
PETITCLERC T, 1995, NEPHROL DIAL TRANSPL, V10, P212
[8]  
POLASCHEGG H, 1993, INT J ARTIF ORGANS, V16, P186
[9]   NONINVASIVE BLOOD-FLOW MEASUREMENT IN EXPANDED POLYTETRAFLUOROETHYLENE GRAFTS FOR HEMODIALYSIS ACCESS [J].
RITTGERS, SE ;
GARCIAVALDEZ, C ;
MCCORMICK, JT ;
POSNER, MP .
JOURNAL OF VASCULAR SURGERY, 1986, 3 (04) :635-642
[10]  
Sands J, 1992, ASAIO J, V38, pM524, DOI 10.1097/00002480-199207000-00090