Characterization of a clinical unit for digital radiography based on irradiation side sampling technology

被引:22
作者
Rivetti, Stefano [1 ]
Lanconelli, Nico [2 ]
Bertolini, Marco [3 ]
Nitrosi, Andrea [3 ]
Burani, Aldo [1 ]
机构
[1] Osped Sassuolo SpA, I-41049 Sassuolo, Italy
[2] Univ Bologna, Dept Phys, Alma Mater Studiorum, I-40127 Bologna, Italy
[3] Azienda Osped ASMN, Med Phys Unit, Ist Ricovero & Cura Carattere Sci, I-42123 Reggio Emilia, Italy
关键词
digital radiology; irradiation side sampling; modulation transfer function; detective quantum efficiency; CONTRAST-DETAIL ANALYSIS; DETECTIVE QUANTUM EFFICIENCY; FLAT-PANEL DETECTORS; IMAGE QUALITY; CHEST RADIOGRAPHY; PSYCHOPHYSICAL CHARACTERIZATION; PHYSICAL-CHARACTERIZATION; PERFORMANCE EVALUATION; AMORPHOUS SELENIUM; SYSTEMS;
D O I
10.1118/1.4820364
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A characterization of a clinical unit for digital radiography (FUJIFILM FDR D-EVO) is presented. This system is based on the irradiation side sampling (ISS) technology and can be equipped with two different scintillators: one traditional gadolinium-oxysulphide phosphor (GOS) and a needle structured cesium iodide (CsI) phosphor panel. Methods: The characterization was achieved in terms of response curve, modulation transfer function (MTF), noise power spectra (NPS), detective quantum efficiency (DQE), and psychophysical parameters (contrast-detail analysis with an automatic reading of CDRAD images). For both scintillation screens the authors accomplished the measurements with four standard beam conditions: RAQ3, RQA5, RQA7, and RQA9. Results: At the Nyquist frequency (3.33 lp/mm) the MTF is about 35% and 25% for CsI and GOS detectors, respectively. The CsI scintillator has better noise properties than the GOS screen in almost all the conditions. This is particularly true for low-energy beams, where the noise for the GOS system can go up to a factor 2 greater than that found for CsI. The DQE of the CsI detector reaches a peak of 60%, 60%, 58%, and 50% for the RQA3, RQA5, RQA7, and RQA9 beams, respectively, whereas for the GOS screen the maximum DQE is 40%, 44%, 44%, and 35%. The contrast-detail analysis confirms that in the majority of cases the CsI scintillator is able to provide improved outcomes to those obtained with the GOS screen. Conclusions: The limited diffusion of light produced by the ISS reading makes possible the achievement of very good spatial resolution. In fact, the MTF of the unit with the CsI panel is only slightly lower to that achieved with direct conversion detectors. The combination of very good spatial resolution, together with the good noise properties reached with the CsI screen, allows achieving DQE on average about 1.5 times greater than that obtained with GOS. In fact, the DQE of unit equipped with CsI is comparable to the best alternative methods available which are based on the same technology, and similar to others based on an a-Se direct conversion detectors. (c) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:11
相关论文
共 33 条
[1]   A comparison of digital radiography systems in terms of effective detective quantum efficiency [J].
Bertolini, Marco ;
Nitrosi, Andrea ;
Rivetti, Stefano ;
Lanconelli, Nico ;
Pattacini, Pierpaolo ;
Ginocchi, Vladimiro ;
Iori, Mauro .
MEDICAL PHYSICS, 2012, 39 (05) :2617-2627
[2]   On site evaluation of three flat panel detectors for digital radiography [J].
Borasi, G ;
Nitrosi, A ;
Ferrari, P ;
Tassoni, D .
MEDICAL PHYSICS, 2003, 30 (07) :1719-1731
[3]   Contrast-detail analysis of three flat panel detectors for digital radiography [J].
Borasi, Giovanni ;
Samei, Ehsan ;
Bertolini, Marco ;
Nitrosi, Andrea ;
Tassoni, Davide .
MEDICAL PHYSICS, 2006, 33 (06) :1707-1719
[4]   Intercomparison of methods for image quality characterization. II. Noise power spectrum [J].
Dobbins, James T., III ;
Samei, Ehsan ;
T Ranger, Nicole ;
Chen, Ying .
MEDICAL PHYSICS, 2006, 33 (05) :1466-1475
[5]   Performance evaluation of a computed radiography imaging device using a typical "front side" and novel "dual side" readout storage phosphors [J].
Fetterly, KA ;
Schueler, BA .
MEDICAL PHYSICS, 2006, 33 (02) :290-296
[6]   Performance evaluation of a "dual-side read" dedicated mammography computed radiography system [J].
Fetterly, KA ;
Schueler, BA .
MEDICAL PHYSICS, 2003, 30 (07) :1843-1854
[7]   Imaging characteristics of an amorphous silicon flat-panel detector for digital chest radiography [J].
Floyd, CE ;
Warp, RJ ;
Dobbins, JT ;
Chotas, HG ;
Baydush, AH ;
Vargas-Voracek, R ;
Ravin, CE .
RADIOLOGY, 2001, 218 (03) :683-688
[8]  
*INT EL COMM, 2003, IEC61267
[9]  
International Electrotechnical Commission, 2003, IEC622201
[10]   Characterization of noise sources for two generations of computed radiography systems using powder and crystalline photostimulable phosphors [J].
Mackenzie, Alistair ;
Honey, Ian D. .
MEDICAL PHYSICS, 2007, 34 (08) :3345-3357