A novel iterative reconstruction algorithm allows reduced dose multidetector-row CT imaging of mechanical prosthetic heart valves

被引:30
作者
Habets, Jesse [1 ]
Symersky, Petr [2 ]
de Mol, Bas A. J. M. [2 ]
Mali, Willem P. Th. M. [1 ]
Leiner, Tim [1 ]
Budde, Ricardo P. J. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiol, NL-3508 GA Utrecht, Netherlands
[2] Univ Amsterdam, Acad Med Ctr, Dept Cardiothorac Surg, NL-1105 AZ Amsterdam, Netherlands
关键词
Prosthetic heart valve; Computed tomography; Iterative reconstruction; Artifacts; Image noise; Reconstruction algorithm; COMPUTED-TOMOGRAPHY; ANGIOGRAPHY;
D O I
10.1007/s10554-011-9954-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Multidetector-row CT is promising for prosthetic heart valve (PHV) assessment but retrospectively ECG-gated scanning has a considerable radiation dose. Recently introduced iterative reconstruction (IR) algorithms may enable radiation dose reduction with retained image quality. Furthermore, PHV image quality on the CT scan mainly depends on extent of PHV artifacts. IR may decrease streak artifacts. We compared image noise and artifact volumes in scans of mechanical PHVs reconstructed with conventional filtered back projection (FBP) to lower dose scans reconstructed with IR. Four different PHVs (St. Jude, Carbomedics, ON-X and Medtronic Hall) were scanned in a pulsatile in vitro model. Ten retrospectively ECG-gated CT scans were performed of each PHV at 120 kV, 600 mAs (high-dose CTDIvol 35.3 mGy) and 120 kV, 300 mAs (low-dose CTDIvol 17.7 mGy) on a 64 detector-row scanner. Diastolic and systolic images were reconstructed with FBP (high and low-dose) and the IR algorithm (low-dose only). Hypo- and hyperdense artifact volumes were determined using two threshold filters. Image noise was measured. Mean hypo- and hyperdense artifact volumes (mm(3)) were 1,235/5,346 (high-dose FBP); 2,405/6,877 (low-dose FBP) and 1,218/5,333 (low-dose IR). Low-dose IR reconstructions had similar image noise compared to high-dose FBP (16.5 +/- A 1.7 vs. 16.3 +/- A 1.6, mean +/- A SD, respectively, P = 1.0). IR allows ECG-gated PHV imaging with similar image noise and PHV artifacts at 50% less dose compared to conventional FBP in an pulsatile in vitro model.
引用
收藏
页码:1567 / 1575
页数:9
相关论文
共 20 条
[1]   Influence of slice thickness and reconstruction kernel on the computed tomographic attenuation of coronary atherosclerotic plaque [J].
Achenbach, Stephan ;
Boehmer, Kerstin ;
Pflederer, Tobias ;
Ropers, Dieter ;
Seltmann, Martin ;
Lell, Michael ;
Anders, Katharina ;
Kuettner, Axel ;
Uder, Michael ;
Daniel, Werner G. ;
Marwan, Mohamed .
JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2010, 4 (02) :110-115
[2]   Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition [J].
Achenbach, Stephan ;
Marwan, Mohamed ;
Ropers, Dieter ;
Schepis, Tiziano ;
Pflederer, Tobias ;
Anders, Katharina ;
Kuettner, Axel ;
Daniel, Werner G. ;
Uder, Michael ;
Lell, Michael M. .
EUROPEAN HEART JOURNAL, 2010, 31 (03) :340-346
[3]  
[Anonymous], EUR RADIOL
[4]   Evaluation of Two Iterative Techniques for Reducing Metal Artifacts in Computed Tomography [J].
Boas, F. Edward ;
Fleischmann, Dominik .
RADIOLOGY, 2011, 259 (03) :894-902
[5]   Evaluation of Anatomic Valve Opening and Leaflet Morphology in Aortic Valve Bioprosthesis by Using Multidetector CT: Comparison with Transthoracic Echocardiography [J].
Chenot, Fabien ;
Montant, Patrick ;
Goffinet, Celine ;
Pasquet, Agnes ;
Vancraeynest, David ;
Coche, Emmanuel ;
Vanoverschelde, Jean-Louis ;
Gerber, Bernhard L. .
RADIOLOGY, 2010, 255 (02) :377-385
[6]   Multislice Computed Tomography in Infective Endocarditis Comparison With Transesophageal Echocardiography and Intraoperative Findings [J].
Feuchtner, Gudrun M. ;
Stolzmann, Paul ;
Dichtl, Wolfgang ;
Schertler, Thomas ;
Bonatti, Johannes ;
Scheffel, Hans ;
Mueller, Silvana ;
Plass, Andre ;
Mueller, Ludwig ;
Bartel, Thomas ;
Wolf, Florian ;
Alkadhi, Hatem .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 53 (05) :436-444
[7]   Computed tomography-old ideas and new technology [J].
Fleischmann, Dominik ;
Boas, F. Edward .
EUROPEAN RADIOLOGY, 2011, 21 (03) :510-517
[8]   Diagnostic evaluation of left-sided prosthetic heart valve dysfunction [J].
Habets, Jesse ;
Budde, Ricardo P. ;
Symersky, Petr ;
van den Brink, Renee B. ;
de Mol, Bas A. ;
Mali, Willem P. ;
van Herwerden, Lex A. ;
Chamuleau, Steven A. .
NATURE REVIEWS CARDIOLOGY, 2011, 8 (08) :466-478
[9]   Prosthetic heart valve assessment with multidetector-row CT: imaging characteristics of 91 valves in 83 patients [J].
Habets, Jesse ;
Symersky, Petr ;
van Herwerden, Lex A. ;
de Mol, Bas A. J. M. ;
Spijkerboer, Anje M. ;
Mali, Willem P. Th. M. ;
Budde, Ricardo P. J. .
EUROPEAN RADIOLOGY, 2011, 21 (07) :1390-1396
[10]   The role of ECG-gated MDCT in the evaluation of aortic and mitral mechanical valves: Initial experience [J].
Konen, Eli ;
Goitein, Orly ;
Feinberg, Micha S. ;
Eshet, Yael ;
Raanani, Ehud ;
Rimon, Uri ;
Di-Segni, Elio .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 191 (01) :26-31