Metal artifact reduction in CT using fusion based prior image

被引:48
|
作者
Wang, Jun
Wang, Shijie [1 ]
Chen, Yang
Wu, Jiasong
Coatrieux, Jean-Louis
Luo, Limin
机构
[1] Southeast Univ, Lab Image Sci & Technol LIST, Nanjing 210096, Jiangsu, Peoples R China
关键词
computed tomography; metal artifact reduction; image fusion; INTERPOLATION; SEGMENTATION; SUPPRESSION; ALGORITHM;
D O I
10.1118/1.4812424
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In computed tomography, metallic objects in the scanning field create the so-called metal artifacts in the reconstructed images. Interpolation-based methods for metal artifact reduction (MAR) replace the metal-corrupted projection data with surrogate data obtained from interpolation using the surrounding uncorrupted sinogram information. Prior-based MAR methods further improve interpolation-based methods by better estimating the surrogate data using forward projections from a prior image. However, the prior images in most existing prior-based methods are obtained from segmented images and misclassification in segmentation often leads to residual artifacts and tissue structure loss in the final corrected images. To overcome these drawbacks, the authors propose a fusion scheme, named fusion prior-based MAR (FP-MAR). Methods: The FP-MAR method consists of (i) precorrect the image by means of an interpolation-based MAR method and an edge-preserving blur filter; (ii) generate a prior image from the fusion of this precorrected image and the originally reconstructed image with metal parts removed; (iii) forward project this prior image to guide the estimation of the surrogate data using well-developed replacement techniques. Results: Both simulations and clinical image tests are carried out to show that the proposed FP-MAR method can effectively reduce metal artifacts. A comparison with other MAR methods demonstrates that the FP-MAR method performs better in artifact suppression and tissue feature preservation. Conclusions: From a wide range of clinical cases to which FP-MAR has been tested (single or multiple pieces of metal, various shapes, and sizes), it can be concluded that the proposed fusion based prior image preserves more tissue information than other segmentation-based prior approaches and can provide better estimates of the surrogate data in prior-based MAR methods. (C) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] CT Image Ring Artifact Reduction Based On Prior Image
    Yuan, C.
    Qi, H.
    Chen, Z.
    Wu, S.
    Xu, Y.
    Zhou, L.
    MEDICAL PHYSICS, 2016, 43 (06) : 3388 - 3388
  • [2] Metal Artifact Reduction with a Deep Image Prior
    Xu, Q.
    Wang, J.
    Shirato, H.
    Xing, L.
    MEDICAL PHYSICS, 2018, 45 (06) : E704 - E704
  • [3] Deep Sinogram Completion With Image Prior for Metal Artifact Reduction in CT Images
    Yu, Lequan
    Zhang, Zhicheng
    Li, Xiaomeng
    Xing, Lei
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2021, 40 (01) : 228 - 238
  • [4] CT Metal Artifact Reduction Using MR Image Patches
    Nielsen, Jonathan S.
    Edmund, Jens M.
    Van Leemput, Koen
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [5] Prior-based Metal Artifact Reduction in CT using Statistical Metal Segmentation on Projection Images
    Hegazy, M. A.
    Cho, M. H.
    Lee, S. Y.
    Kim, Kisoo
    2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,
  • [6] Image-based metal artifact reduction in CT images
    Serifovic-Trbalic, A.
    Trbalic, A.
    2016 39TH INTERNATIONAL CONVENTION ON INFORMATION AND COMMUNICATION TECHNOLOGY, ELECTRONICS AND MICROELECTRONICS (MIPRO), 2016, : 442 - 446
  • [7] A metal artifact reduction method for a dental CT based on adaptive local thresholding and prior image generation
    Hegazy, Mohamed A. A.
    Cho, Min Hyoung
    Lee, Soo Yeol
    BIOMEDICAL ENGINEERING ONLINE, 2016, 15
  • [8] A metal artifact reduction method for a dental CT based on adaptive local thresholding and prior image generation
    Mohamed A. A. Hegazy
    Min Hyoung Cho
    Soo Yeol Lee
    BioMedical Engineering OnLine, 15
  • [9] A new metal artifact reduction algorithm based on a deteriorated CT image
    Kano, Toru
    Koseki, Michihiko
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2016, 24 (06) : 901 - 912
  • [10] A prior-based metal artifact reduction algorithm for x-ray CT
    Li, Ming
    Zheng, Jian
    Zhang, Tao
    Guan, Yihui
    Xu, Pin
    Sun, Mingshan
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2015, 23 (02) : 229 - 241