Simultaneous and direct determination of urea and creatinine in human urine using a cost-effective flow injection system equipped with in-house contactless conductivity detector and LED colorimeter

被引:25
作者
Chaneam, Sumonmarn [1 ,2 ]
Kaewyai, Kamparnart [2 ,3 ,4 ]
Mantim, Thitirat [2 ,5 ]
Chaisuksant, Rasamee [1 ]
Wilairat, Prapin [6 ]
Nacapricha, Duangjai [2 ,3 ,4 ]
机构
[1] Silpakorn Univ, Fac Sci, Dept Chem, Rajamankha Nai Rd, Amphoe Muang, Nakhon Pathom 73000, Thailand
[2] FIRST Labs, Flow Innovat Res Sci & Technol Lab, Bangkok, Thailand
[3] Mahidol Univ, Fac Sci, Dept Chem, Rama 6 Rd, Bangkok 10400, Thailand
[4] Mahidol Univ, Fac Sci, Ctr Excellence Innovat Chem, Rama 6 Rd, Bangkok 10400, Thailand
[5] Srinakharinwirot Univ, Dept Chem, Fac Sci, Bangkok 10110, Thailand
[6] Mahidol Univ, Natl Doping Control Ctr, Rama 6 Rd, Bangkok 10400, Thailand
关键词
Urea; Creatinine; Urine; Flow injection analysis (FIA); Capacitively coupled contactless; conductivity detector (C4D); Paired emitter detector diodes (PEDD); CAPILLARY-ELECTROPHORESIS; AMMONIACAL NITROGEN; FUNDAMENTAL-ASPECTS; BIOSENSOR; SAMPLES; SEPARATION; DIALYSATE; CARBONATE; PHOSPHATE; GLUCOSE;
D O I
10.1016/j.aca.2019.05.003
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This work presents a cost-effective and simple flow injection analysis (FIA) system for simultaneous and direct determination of urea and creatinine in human urine. The FIA system comprises two in-house detectors, a contactless conductivity detector and a light emitting diode (LED) detector. The contactless detector was built as a flow-through detection cell with axial electrodes, commonly known as capacitively coupled contactless conductivity detector (C4D) and the diode detector was fabricated based on the concept of paired emitter detector diodes (PEDD). With appropriate dilution of urine, the sample is directly injected into a stream of glycine-NaOH buffer pH 8.8 (the gas donor stream) and is carried by the carrier through a urease minicolumn for on-line enzymatic hydrolysis. The generated NH3 diffuses from the carrier stream through a porous polytetrafluoroethylene (PTFE) membrane into a stream of deionized water (the acceptor stream) leading to an increase in the signal at the C4D due to the NH3 dissolution in the water. In this system, creatinine is determined based on the Jaffe reaction by merging a stream of alkaline picrate with the gas donor stream. The change in color is detected using the PEDD equipped with two green LEDs. Under the optimum condition, the linear range of urea and creatinine were 30-240 mg L-1 and 10-500mg L-1, with limits of detection of 9.0 mg L-1 and 0.9 mg L-1, respectively. The proposed system provides satisfactorily good precision (RSD < 3%), with sample throughput of 31 sample h(-1) for the two analytes. The FIA system tolerates potential interference commonly found in human urine. The system was successfully applied and validated with selected reference methods. (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页码:54 / 61
页数:8
相关论文
共 64 条
[1]   Citrate-capped silver nanoparticles as a probe for sensitive and selective colorimetric and spectrophotometric sensing of creatinine in human urine [J].
Alula, Melisew Tadele ;
Karamchand, Leshern ;
Hendricks, Nicolette R. ;
Blackburn, Jonathan M. .
ANALYTICA CHIMICA ACTA, 2018, 1007 :40-49
[2]  
Ballard E., 2016, AM J KIDNEY DIS, V67, pA78
[3]  
BONSNES RW, 1945, J BIOL CHEM, V158, P581
[4]   Analysis of creatine and creatinine in urine by capillary electrophoresis [J].
Burke, DG ;
MacLean, PG ;
Walker, RA ;
Dewar, PJ ;
Smith-Palmer, T .
JOURNAL OF CHROMATOGRAPHY B, 1999, 732 (02) :479-485
[5]   Green analytical method for simultaneous determination of salinity, carbonate and ammoniacal nitrogen in waters using flow injection coupled dual-channel C4D [J].
Chaneam, Sumonmarn ;
Inpota, Prawpan ;
Saisarai, Suttipong ;
Wilairat, Prapin ;
Ratanawimarnwong, Nuanlaor ;
Uraisin, Kanchana ;
Meesiri, Wanchai ;
Nacapricha, Duangjai .
TALANTA, 2018, 189 :196-204
[6]   Green analytical flow method for the determination of total sulfite in wine using membraneless gas-liquid separation with contactless conductivity detection [J].
Chantipmanee, Nattapong ;
Alahmad, Waleed ;
Sonsa-ard, Thitaporn ;
Uraisin, Kanchana ;
Ratanawimarnwong, Nuanlaor ;
Mantim, Thitirat ;
Nacapricha, Duangjai .
ANALYTICAL METHODS, 2017, 9 (43) :6107-6116
[7]   Tandem measurements of iron and creatinine by cross injection analysis with application to urine from thalassemic patients [J].
Choengchan, N. ;
Mantim, T. ;
Inpota, R. ;
Nacapricha, D. ;
Wilairat, P. ;
Jittangprasert, P. ;
Waiyawat, W. ;
Fucharoen, S. ;
Sirankpracha, P. ;
Morales, N. Phumala .
TALANTA, 2015, 133 :52-58
[8]   Determination of urea using high-performance liquid chromatography with fluorescence detection after automated derivatisation with xanthydrol [J].
Clark, Shona ;
Francis, Paul S. ;
Conlan, Xavier A. ;
Barnett, Neil W. .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1161 (1-2) :207-213
[9]  
da Silva JAF, 1998, ANAL CHEM, V70, P4339
[10]   Determination of urea, allantoin and lysine pyroglutamate in cosmetic samples by hydrophilic interaction chromatography [J].
Dallet, P ;
Labat, L ;
Kummer, E ;
Dubost, JP .
JOURNAL OF CHROMATOGRAPHY B, 2000, 742 (02) :447-452