Development and commissioning of a multileaf collimator model in Monte Carlo dose calculations for intensity-modulated radiation therapy

被引:34
作者
Jang, SY
Vassiliev, ON
Liu, HH
Mohan, R
Siebers, JV
机构
[1] Univ Texas, MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Virginia Commonwealth Univ, Dept Radiat Oncol, Richmond, VA 23298 USA
关键词
IMRT; multileaf collimators; quality assurance; Monte Carlo; commission MLC model; leaf density; interleaf air gap; tongue and groove; leaf transmission and leakage;
D O I
10.1118/1.2170598
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A multileaf collimator (MLC) model, "MATMLC," was developed to simulate MLCs for Monte Carlo (MC) dose calculations of intensity-modulated radiation therapy (IMRT). This model describes MLCs using matrices of regions, each of which can be independently defined for its material and geometry, allowing flexibility in simulating MLCs from various manufacturers. The free parameters relevant to the dose calculations with this MLC model included MLC leaf density, interleaf air gap, and leaf geometry. To commission the MLC model and its free parameters for the Varian Millennium MLC-120 (Varian Oncology Systems, Palo Alto, CA), we used the following leaf patterns: (1) MLC-blocked fields to test the effects of leaf transmission and leakage; (2) picket-fence fields to test the effects of the interleaf air gap and tongue-groove design; and (3) abutting-gap fields to test the effects of rounded leaf ends. Transmission ratios and intensity maps for these leaf patterns were calculated with various sets of modeling parameters to determine their dosimetric effects, sensitivities, and their optimal combinations to give the closest agreement with measured results. Upon commissioning the MLC model, we computed dose distributions for clinical IMRT plans using the MC system and verified the results with those from ion chamber and thermoluminescent dosimeter measurements in water phantoms and anthropomorphic phantoms. This study showed that the MLC transmission ratios were strongly dependent on both leaf density and the interleaf air gap. The effect of interleaf air gap and tongue-groove geometry can be determined most effectively through fence-type MLC patterns. Using the commissioned MLC model, we found that the calculated dose from the MC system agreed with the measured data within clinically acceptable criteria from low- to high-dose regions, showing that the model is acceptable for clinical applications. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:770 / 781
页数:12
相关论文
共 28 条
[1]  
*AAPM, 2001, 72 AAPM
[2]   A method for determining multileaf collimator transmission and scatter for dynamic intensity modulated radiotherapy [J].
Arnfield, MR ;
Siebers, JV ;
Kim, JO ;
Wu, QW ;
Keall, PJ ;
Mohan, R .
MEDICAL PHYSICS, 2000, 27 (10) :2231-2241
[3]   Intensity-modulated radiotherapy: Current status and issues of interest [J].
Boyer, AL ;
Butler, EB ;
DiPetrillo, TA ;
Engler, MJ ;
Fraass, B ;
Grant, W ;
Ling, CC ;
Low, DA ;
Mackie, TR ;
Mohan, R ;
Purdy, JA ;
Roach, M ;
Rosenman, JG ;
Verhey, LJ ;
Wong, JW ;
Cumberlin, RL ;
Stone, H ;
Palta, JR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (04) :880-914
[4]   Rapid radiographic film calibration for IMRT verification using automated MLC fields [J].
Childress, NL ;
Dong, L ;
Rosen, II .
MEDICAL PHYSICS, 2002, 29 (10) :2384-2390
[5]   Reference photon dosimetry data and reference phase space data for the 6 MV photon beam from Varian Clinac 2100 series linear accelerators [J].
Cho, SH ;
Vassiliev, ON ;
Lee, S ;
Liu, HH .
MEDICAL PHYSICS, 2005, 32 (01) :137-148
[6]   Dose measurements compared with Monte Carlo simulations of narrow 6 MV multileaf collimator shaped photon beams [J].
De Vlamynck, K ;
Palmans, H ;
Verhaegen, F ;
De Wagter, C ;
De Neve, W ;
Thierens, H .
MEDICAL PHYSICS, 1999, 26 (09) :1874-1882
[7]   Efficient particle transport simulation through beam modulating devices for Monte Carlo treatment planning [J].
Fippel, M .
MEDICAL PHYSICS, 2004, 31 (05) :1235-1242
[8]   Development and validation of a BEAMnrc component module for accurate Monte Carlo modelling of the Varian dynamic Millennium multileaf collimator [J].
Heath, E ;
Seuntjens, J .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (24) :4045-4063
[9]   Investigation of variance reduction techniques for Monte Carlo photon dose calculation using XVMC [J].
Kawrakow, I ;
Fippel, M .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (08) :2163-2183
[10]   A Monte Carlo study of radiation transport through multileaf collimators [J].
Kim, JO ;
Siebers, JV ;
Keall, PJ ;
Arnfield, MR ;
Mohan, R .
MEDICAL PHYSICS, 2001, 28 (12) :2497-2506