Design and implementation of a closed-loop control system for infusion of propofol guided by bispectral index (BIS)

被引:33
作者
Reboso, J. A. [1 ]
Mendez, J. A. [2 ]
Reboso, H. J. [2 ]
Leon, A. M. [1 ]
机构
[1] Hosp Univ Canarias, Dept Anesthesiol & Reanimat, San Cristobal la Laguna, Spain
[2] Univ La Laguna, Dept Syst Engn & Automat & Comp Architecture & Te, E-38207 San Cristobal la Laguna, Spain
关键词
TARGET-CONTROLLED INFUSION; ANESTHESIA; REMIFENTANIL; PERFORMANCE; TITRATION; INDUCTION; HYPNOSIS;
D O I
10.1111/j.1399-6576.2012.02738.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background This study describes the design of a hypnosis closed-loop control system with propofol. The controller used a proportional-integral (PI) algorithm with the bispectral index (BIS) as the feedback signal. Our hypothesis was that a PI closed-loop control could be applied in clinical practice safely keeping the BIS within a pre-determined target range. Methods The adjustment of the PI parameters was based on simulation. The procedure had three steps: obtaining a patient model using data from 12 patients, designing and adjusting the controller in simulation, and fine tuning the PI parameters in a pilot study (10 patients). The resulting controller was tested in 24 American Society of Anesthesiology (ASA) III patients. The controller directly decides the infusion rate of propofol, and no model is necessary in its online operation. The BIS target was set to 50. Remifentanil was used for analgesia. Results We evaluated the efficiency and safety of the automatic feedback system. It worked properly in all the patients. The median performance error was -1.62, and the median absolute performance error was 11.03. Average propofol-normalized consumption was 5.3 +/- 1.8?mg/kg/h. Mean percentage of BIS in the range 4060 was 83%. Mean time to open eyes was 8 +/- 4?min. Time to extubation was 9 +/- 5?min. Hemodynamic adverse event or intraoperative awareness were not recorded. Conclusions The closed-loop system was able to maintain the BIS within an acceptable range of levels. The control of a propofol infusion guided by the BIS is feasible without hemodynamic instability in ASA I/II patients.
引用
收藏
页码:1032 / 1041
页数:10
相关论文
共 50 条
  • [21] Thoracic Epidural Analgesia With Levobupivacaine Reduces Remifentanil and Propofol Consumption Evaluated by Closed-Loop Titration Guided by the Bispectral Index: A Double-Blind Placebo-Controlled Study
    Dumans-Nizard, Virginie
    Le Guen, Morgan
    Sage, Edouard
    Chazot, Thierry
    Fischler, Marc
    Liu, Ngai
    ANESTHESIA AND ANALGESIA, 2017, 125 (02) : 635 - 642
  • [22] Robust closed-loop control of hypnosis with propofol using WAVCNS index as the controlled variable
    Hahn, Jin-Oh
    Dumont, Guy A.
    Ansermino, J. Mark
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2012, 7 (05) : 517 - 524
  • [23] Robust closed-loop control of induction and maintenance of propofol anesthesia in children
    West, Nicholas
    Dumont, Guy A.
    van Heusden, Klaske
    Petersen, Christian L.
    Khosravi, Sara
    Soltesz, Kristian
    Umedaly, Aryannah
    Reimer, Eleanor
    Ansermino, J. Mark
    PEDIATRIC ANESTHESIA, 2013, 23 (08) : 712 - 719
  • [24] BIS feedback closed-loop target-controlled infusion of propofol or etomidate in elderly patients with spinal surgery
    Zheng, Zhonglei
    Su, Yuqiang
    Fan, Xiaoying
    Zhang, Wanping
    Li, Jing
    Xue, Sha
    AMERICAN JOURNAL OF TRANSLATIONAL RESEARCH, 2023, 15 (02): : 1231 - 1238
  • [25] Optimal BIS reference functions for closed-loop induction of anesthesia with propofol
    Jarrett, Ryan T.
    Blair, James L.
    Shotwell, Matthew S.
    COMPUTERS IN BIOLOGY AND MEDICINE, 2022, 144
  • [26] A semi-adaptive control approach to closed-loop medication infusion
    Jin, Xin
    Kim, Chang-Sei
    Dumont, Guy A.
    Ansermino, J. Mark
    Hahn, Jin-Oh
    INTERNATIONAL JOURNAL OF ADAPTIVE CONTROL AND SIGNAL PROCESSING, 2017, 31 (02) : 240 - 254
  • [27] Design and Evaluation of a Closed-Loop Anesthesia System With Robust Control and Safety System
    West, Nicholas
    van Heusden, Klaske
    Gorges, Matthias
    Brodie, Sonia
    Rollinson, Aryannah
    Petersen, Christian L.
    Dumont, Guy A.
    Ansermino, J. Mark
    Merchant, Richard N.
    ANESTHESIA AND ANALGESIA, 2018, 127 (04) : 883 - 894
  • [28] Nitrous oxide does not produce a clinically important sparing effect during closed-loop delivered propofol-remifentanil anaesthesia guided by the bispectral index: a randomized multicentre study
    Liu, N.
    Le Guen, M.
    Boichut, N.
    Genty, A.
    Herail, T.
    Schmartz, D.
    Khefif, G.
    Landais, A.
    Bussac, J. J.
    Charmeau, A.
    Baars, J.
    Rehberg, B.
    Tricoche, S.
    Chazot, T.
    Sessler, D. I.
    Fischler, M.
    BRITISH JOURNAL OF ANAESTHESIA, 2014, 112 (05) : 842 - 851
  • [29] Comparison of closed loop vs. manual administration of propofol using the Bispectral index in cardiac surgery
    Agarwal, J.
    Puri, G. D.
    Mathew, P. J.
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2009, 53 (03) : 390 - 397
  • [30] Comparing Closed-loop Control of Drug Infusion using MPC and PID
    Setati, Tiro
    Perold, Willem Jacobus
    Fourie, Pieter Rousseau
    Withey, Daniel
    PROCEEDINGS OF THE 15TH INTERNATIONAL JOINT CONFERENCE ON BIOMEDICAL ENGINEERING SYSTEMS AND TECHNOLOGIES (BIODEVICES), VOL 1, 2021, : 126 - 133