Evaluation of localization errors for craniospinal axis irradiation delivery using volume modulated arc therapy and proposal of a technique to minimize such errors

被引:41
作者
Myers, Pamela [1 ]
Stathakis, Sotirios [1 ]
Mavroidis, Panayiotis [1 ]
Esquivel, Carlos [1 ]
Papanikolaou, Niko [1 ]
机构
[1] Univ Texas Hlth Sci Ctr San Antonio, Dept Radiat Oncol, Canc Therapy & Res Ctr, Sch Med, San Antonio, TX 78229 USA
关键词
Craniospinal axis irradiation; VMAT; Setup errors; RADIOTHERAPY;
D O I
10.1016/j.radonc.2013.05.026
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To dosimetrically evaluate the effects of improper patient positioning in the junction area of a VMAT cranio-spinal axis irradiation technique consisting of one superior and one inferior arc and propose a solution to minimize these patient setup errors. Methods: Five (n = 5) cranio-spinal axis irradiation patients were planned with 2 arcs: one superior and one inferior. In order to mimic patient setup errors, the plans were recalculated with the inferior isocenter shifted by: 1, 2, 5, and 10 mm superiorly, and 1, 2, 5, and 10 mm inferiorly. The plans were then compared with the corresponding original, non-shifted arc plans on the grounds of target metrics such as conformity number and homogeneity index, as well as several normal tissue dose descriptors. "Gradient-optimized" plans were then created for each patient in an effort to reduce dose discrepancies due to setup errors. Results: Percent differences were calculated in order to compare each of the eight shifted plans with the original non-shifted arc plan, which corresponds to the ideal patient setup. The conformity number was on average lower by 0.9%, 2.7%, 5.8%, and 9.1% for the 1, 2, 5, and 10 mm inferiorly-shifted plans and 0.4%, 0.8%, 2.8%, and 6.0% for the respective superiorly-shifted plans. The homogeneity indices were, averaged among the five patients and they indicated less homogeneous dose distributions by 0.03%, 0.3%, 1.0%, and 2.8% for the inferior shifts and 0.2%, 1.2%, 6.3%, and 15.3% for the superior shifts. Overall, the mean doses to the organs at risk deviate by less than 2% for the 1, 2, and 5 mm shifted plans. The 10 mm shifted plans, however, showed average percent differences, over all studied organs, from the original plan of up to 5.6%. Using "gradient-optimized" plans, the average dose differences were reduced by 0.2%, 0.5%, 1.2%, and 2.1% for 1, 2, 5, and 10 mm shifts, respectively compared to the originally optimized plans, and the maximum dose differences were reduced by 11.7%, 8.5%, 12.4%, and 13.9% on average for the 1, 2, 5, and 10 mm shifted plans. Conclusions: Setup errors related to isocenter shifting should be minimized in order to provide the patient with the most dosimetrically accurate treatment possible. Errors of 1-2 mm can negatively affect the quality of the delivered treatment, most notably in the arc junction area, but the deterioration of the treatment plan accuracy is not as problematic as in the cases of larger errors such as 5-10 mm. By employing a new planning technique, the dose differences due to setup errors can be greatly reduced. (C) 2013 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:107 / 113
页数:7
相关论文
共 17 条
[1]   Dose calculations for external photon beams in radiotherapy [J].
Ahnesjö, A ;
Aspradakis, MM .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (11) :R99-R155
[2]   Helical Volumetric Modulated Arc Therapy for Treatment of Craniospinal Axis [J].
Bedford, James L. ;
Lee, Young K. ;
Saran, Frank H. ;
Warrington, Alan P. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 83 (03) :1047-1054
[3]   QUANTITATIVE ANALYSES OF NORMAL TISSUE EFFECTS IN THE CLINIC (QUANTEC): AN INTRODUCTION TO THE SCIENTIFIC ISSUES [J].
Bentzen, Soren M. ;
Constine, Louis S. ;
Deasy, Joseph O. ;
Eisbruch, Avi ;
Jackson, Andrew ;
Marks, Lawrence B. ;
Ten Haken, Randall K. ;
Yorke, Ellen D. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (03) :S3-S9
[4]   Conformity index:: A review [J].
Feuvret, L ;
Noël, G ;
Mazeron, JJ ;
Bey, P .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 64 (02) :333-342
[5]   Cranio-spinal irradiation with volumetric modulated arc therapy: A multi-institutional treatment experience [J].
Fogliata, Antonella ;
Bergstrom, Stefan ;
Cafaro, Ines ;
Clivio, Alessandro ;
Cozzi, Luca ;
Dipasquale, Giovanna ;
Hallstrom, Per ;
Mancosu, Pietro ;
Navarria, Piera ;
Nicolini, Giorgia ;
Parietti, Emanuela ;
Pesce, Gianfranco A. ;
Richetti, Antonella ;
Scorsetti, Marta ;
Vanetti, Eugenio ;
Weber, Damien C. .
RADIOTHERAPY AND ONCOLOGY, 2011, 99 (01) :79-85
[6]  
Halperin EC, 2008, PEREZ BRADYS PRINCIP
[7]  
Khan FM., 2007, TREATMENT PLANNING R, VSecond
[8]   DEVELOPMENT AND EVALUATION OF MULTIPLE ISOCENTRIC VOLUMETRIC MODULATED ARC THERAPY TECHNIQUE FOR CRANIOSPINAL AXIS RADIOTHERAPY PLANNING [J].
Lee, Young K. ;
Brooks, Corrinne J. ;
Bedford, James L. ;
Warrington, Alan P. ;
Saran, Frank H. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 82 (02) :1006-1012
[9]  
Michalski Jeff M, 2002, J Appl Clin Med Phys, V3, P310, DOI 10.1120/1.1508013
[10]  
Myers P, 2013, INT J MED PHYS CLIN, P2