Optimization of on-line solid phase extraction and HPLC conditions using response surface methodology for determination of WM-5 in mouse plasma and its application to pharmacokinetic study

被引:15
作者
Liu, Lei [1 ,2 ]
Wen, Ya-Bin [3 ]
Liu, Kang-Ning [1 ,2 ]
Sun, Liang [1 ,2 ]
Wu, Meng [2 ,4 ]
Han, Gui-Fang [2 ,4 ]
Lu, Ya-Xin [1 ,2 ]
Wang, Qing-Ming [2 ,4 ]
Yin, Zheng [1 ,2 ]
机构
[1] Nankai Univ, Coll Pharm, Tianjin 300071, Peoples R China
[2] Nankai Univ, State Key Lab Elementoorgan Chem, Tianjin 300071, Peoples R China
[3] Nankai Univ, Coll Life Sci, Tianjin 300071, Peoples R China
[4] Nankai Univ, Coll Chem, Tianjin 300071, Peoples R China
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2013年 / 923卷
基金
中国国家自然科学基金;
关键词
WM-5; On-line solid-phase extraction; HPLC-DAD; Pharmacokinetic study; Response surface methodology; PERFORMANCE LIQUID-CHROMATOGRAPHY; TANDEM MASS-SPECTROMETRY; CYTOTOXIC ALKALOIDS; EXPERIMENTAL-DESIGN; SAMPLE PREPARATION; PHENANTHROINDOLIZIDINE; ACID; DRUGS;
D O I
10.1016/j.jchromb.2013.01.024
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Response surface methodology (RSM) was utilized for rapid and systematic optimization of on-line solid-phase extraction (SPE) parameters to maximize the response and separation of WM-5. The optimization was performed with Box-Behnken designs. Four major parameters were investigated for their contributions to the response and separation of WM-5, with a total of 29 experiments being performed for each instrument, respectively. Quantitative determination of WM-5 in mouse plasma was performed to evaluate the statistical significance of the parameters on chromatographic response. A fully automated on-line SPE and high-performance liquid chromatography (HPLC) with diode array detection (DAD) method was developed for the determination of WM-5 in mouse plasma. Calibration curve with good linearity (r = 0.9989) was obtained in the range of 20-4000 ng/mL in mouse plasma. The limit of detection (LOD) and lower limit of quantification (LLOQ) of the assay were 6 ng/mL and 20 ng/mL, respectively. The overall intra-day and the inter-day variations were less than 1.90%. The recovery of the method was in the range of 93.74-96.33% with RSD less than 3.06%. The optimized method demonstrated good performance in terms of specificity, LLOQ linearity, recovery, precision and accuracy, and was successfully applied to quantify WM-5 in mouse plasma to support the pharmacokinetic study. (c) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:8 / 15
页数:8
相关论文
共 36 条
  • [1] Ali M, 2001, PHARMAZIE, V56, P188
  • [2] [Anonymous], 2001, Guidance for industry, bioanalytical method validation
  • [3] Ashri NY, 2011, BIOANALYSIS, V3, P2003, DOI [10.4155/bio.11.201, 10.4155/BIO.11.201]
  • [4] Optimal experimental design in chromatography
    Atkinson, Anthony C.
    Tobias, Randall D.
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2008, 1177 (01) : 1 - 11
  • [5] Response surface methodology (RSM) as a tool for optimization in analytical chemistry
    Bezerra, Marcos Almeida
    Santelli, Ricardo Erthal
    Oliveira, Eliane Padua
    Villar, Leonardo Silveira
    Escaleira, Luciane Amlia
    [J]. TALANTA, 2008, 76 (05) : 965 - 977
  • [6] Chemler Sherry R, 2009, Curr Bioact Compd, V5, P2
  • [7] Automated Online Solid Phase Extraction Ultra High Performance Liquid Chromatography Method Coupled with Tandem Mass Spectrometry for Determination of Forty-Two Therapeutic Drugs and Drugs of Abuse in Human Urine
    Chiuminatto, Ugo
    Gosetti, Fabio
    Dossetto, Paolo
    Mazzucco, Eleonora
    Zampieri, Davide
    Robotti, Elisa
    Gennaro, Maria Carla
    Marengo, Emilio
    [J]. ANALYTICAL CHEMISTRY, 2010, 82 (13) : 5636 - 5645
  • [8] PHYSIOLOGICAL-PARAMETERS IN LABORATORY-ANIMALS AND HUMANS
    DAVIES, B
    MORRIS, T
    [J]. PHARMACEUTICAL RESEARCH, 1993, 10 (07) : 1093 - 1095
  • [9] Eeckhaut A.V., 2009, J. Chromatogr B, V877, P2198, DOI DOI 10.1016/J.JCHR0MB.2009.01.003
  • [10] Box-Behnken design: An alternative for the optimization of analytical methods
    Ferreira, S. L. C.
    Bruns, R. E.
    Ferreira, H. S.
    Matos, G. D.
    David, J. M.
    Brandao, G. C.
    da Silva, E. G. P.
    Portugal, L. A.
    dos Reis, P. S.
    Souza, A. S.
    dos Santos, W. N. L.
    [J]. ANALYTICA CHIMICA ACTA, 2007, 597 (02) : 179 - 186