Comparison of Keratometric Data Obtained by Automated Keratometer, Dicon CT 200, Allegro Topolyzer, and Pentacam

被引:8
作者
Gonen, Tansu [1 ]
Cosar, Banu [2 ]
Sener, Bozkurt [3 ]
Keskinbora, Kadircan H. [1 ]
机构
[1] Namik Kemal Univ, Sch Med, Dept Ophthalmol, TR-59100 Tekirdag, Turkey
[2] Acibadem Univ, Sch Med, Dept Ophthalmol, Istanbul, Turkey
[3] Acibadem Maslak Hosp Eye Clin, Istanbul, Turkey
关键词
CORNEAL; VIDEOKERATOGRAPHY; REPRODUCIBILITY; REPEATABILITY; TOPOGRAPHY; AGREEMENT; ACCURACY; SYSTEMS;
D O I
10.3928/1081597X-20120723-04
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To compare the keratometric values measured by the automated keratometer, two Placido-based computerized topography systems (Dicon CT 200 [Vismed Inc] and Allegro Topolyzer [WaveLight Inc]), and Scheimpflug analysis (Pentacam [Oculus Optikgerate GmbH]). METHODS: The keratometric data of 200 eyes from 200 patients evaluated for refractive surgery were reviewed retrospectively. Mean simulated keratometry (Sim K) and mean corneal astigmatism measured by the four devices were compared using repeated measures analysis of variance with Bonferroni correction. The analysis of agreement between two measurements was assessed using the method of Bland and Altman. RESULTS: Mean Sim K as measured by the automated keratometer, Dicon CT 200, Allegro Topolyzer, and Pentacam was 43.39 +/- 1.50 diopters (D), 43.55 +/- 1.50 D, 43.45 +/- 1.50 D, and 43.43 +/- 1.45 D, respectively. The Dicon CT 200 measured the mean Sim K to be steeper and the automated keratometer measured the mean Sim K to be flatter than the other devices. Significant differences in corneal astigmatism were noted among the four devices except Dicon CT 200 versus Allegro Topolyzer and Allegro Topolyzer versus Pentacam comparisons (P <.013). For mean Sim K, the 95% limits of agreement between the Pentacam and other three devices were significantly wider than the other comparisons. In Bland-Altman plots comparing the Pentacam to the other devices, extreme outliers were present in 11 (5.5%) eyes. CONCLUSIONS: Because of the wide distribution range and presence of extreme outliers, Pentacam data should be used cautiously in IOL power calculation and astigmatic keratotomy procedures. [J Refract Surg. 2012;28(8):557-561.] doi:10.3928/1081597X-20120723-04
引用
收藏
页码:557 / +
页数:6
相关论文
共 13 条
[1]  
Binder Perry S., 1995, CLAO Journal, V21, P133
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Reliability and repeatability of the Pentacam on corneal curvatures [J].
Chen, Davie ;
Lam, Andrew K. C. .
CLINICAL AND EXPERIMENTAL OPTOMETRY, 2009, 92 (02) :110-118
[4]   Comparison of different techniques of anterior chamber depth and keratometric measurements [J].
Elbaz, Uri ;
Barkana, Yaniv ;
Gerber, Yariv ;
Avni, Isaac ;
Zadok, David .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 143 (01) :48-53
[5]   CALCULATING THE SURGICALLY INDUCED REFRACTIVE CHANGE FOLLOWING OCULAR SURGERY [J].
HOLLADAY, JT ;
CRAVY, TV ;
KOCH, DD .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1992, 18 (05) :429-443
[6]   Repeatability and Reproducibility of Corneal Curvature Measurements Using the Pentacam and Keratron Topography Systems [J].
Kawamorita, Takushi ;
Nakayama, Nanami ;
Uozato, Hiroshi .
JOURNAL OF REFRACTIVE SURGERY, 2009, 25 (06) :539-544
[7]   Comparison and reproducibility of corneal thickness and curvature readings obtained by the Galilei and the Orbscan II analysis systems [J].
Menassa, Nardine ;
Kaufmann, Claude ;
Goggin, Michael ;
Job, Oliver M. ;
Bachmann, Lucas M. ;
Thiel, Michael A. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (10) :1742-1747
[8]   Corneal topography with Scheimpflug imaging and videokeratography: Comparative study of normal eyes [J].
Read, Scott A. ;
Collins, Michael J. ;
Iskander, Robert ;
Davis, Brett A. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (06) :1072-1081
[9]   COMPARISON OF THE SRK-II FORMULA AND OTHER 2ND GENERATION FORMULAS [J].
SANDERS, DR ;
RETZLAFF, J ;
KRAFF, MC .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1988, 14 (02) :136-141
[10]   Accuracy of Scheimpflug corneal power measurements for intraocular lens power calculation [J].
Savini, Giacomo ;
Barboni, Piero ;
Carbonelli, Michele ;
Hoffer, Kenneth J. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (07) :1193-1197