Significant association of coronary stent fracture with in-stent restenosis in sirolimus-eluting stents

被引:18
作者
Kim, Jung-Sun [2 ]
Lee, Seung-Yul [2 ]
Lee, Jung Myung [2 ]
Yoon, Young Won [1 ]
Ahn, Chul-Min [2 ]
Kim, Myung-Hyun [1 ]
Min, Pil-Ki [1 ]
Ko, Young-Guk [2 ]
Hong, Bum-Kee [1 ]
Choi, Donghoon [2 ]
Kwon, Hyuck Moon [1 ]
Jang, Yangsoo [2 ]
Shim, Won-Heum [2 ]
机构
[1] Yonsei Univ, Coll Med, Yong Dong Severance Hosp, Div Cardiol, Seoul 135720, South Korea
[2] Yonsei Univ, Coll Med, Yong Dong Severance Hosp, Div Cardiol,Yonsei Cardiovasc Ctr, Seoul 135720, South Korea
关键词
restenosis; sirolimus-eluting stent; stent fracture; CLINICAL IMPACT; IMPLANTATION; PREDICTORS; LESIONS;
D O I
10.1097/MCA.0b013e32830fd101
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Several reports have suggested that stent fractures in sirolimus-eluting stents (SESs) might be related to in-stent restenosis (ISR). However, the role of stent strut fracture in ISR has not been clearly elucidated. Therefore, we investigated the association of the SES fracture and ISR. Methods From 2003 to 2006, SES implantations with follow-up coronary angiography (CAG) for 628 lesions in 557 patients were analyzed. We reviewed clinical and procedural factors that might affect SES fracture and ISR. The median time interval from stent implantation to follow-up CAG was 9 months (range: 2-30 months). Results ISR occurred in 38 patients (5.7%), and 21 stent fractures (3.3%) were identified by follow-up CAG. Fourteen cases occurred in the left anterior descending artery, and seven occurred in the right coronary artery. The binary ISR rate in the stent fracture group was higher compared with that of the nonfracture group (38.1% vs. 4.6%, P < 0.001). Predictors of ISR as estimated by multivariate analysis were a stent diameter less than 2.75 mm [odds ratio (OR)=2.76, P=0.012], a stent length over 28 mm (OR=3.30, P=0.024), and stent fracture (OR=11.03, P < 0.001) after controlling for the angiographic and clinical risk factors of ISR. Conclusion Stent fracture was an independent predictor of ISR and may be one of the crucial mechanisms of ISR after implantation of an SES. Coron Artery Dis 20:59-63 (c) 2009 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:59 / 63
页数:5
相关论文
共 50 条
  • [41] Incidence and clinical impact of coronary stent fracture after sirolimus-eluting stent implantation
    Aoki, Jiro
    Nakazawa, Gaku
    Tanabe, Kengo
    Hoye, Angela
    Yamamoto, Hirosada
    Nakayama, Tomohiro
    Onuma, Yoshinobu
    Higashikuni, Yasutomi
    Otsuki, Syuji
    Yagishita, Atsuhiko
    Yachi, Sen
    Nakajima, Hiroyoshi
    Hara, Kazuhiro
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2007, 69 (03) : 380 - 386
  • [42] Comparison of neointimal morphology of in-stent restenosis with sirolimus-eluting stents versus bare metal stents: Virtual histology-intravascular ultrasound analysis
    Yamamoto Y.
    Otani H.
    Iwasaka J.
    Park H.
    Sakuma T.
    Kamihata H.
    Iwasaka T.
    [J]. Cardiovascular Intervention and Therapeutics, 2011, 26 (3) : 186 - 192
  • [43] Fracture of a sirolimus-eluting stent in renal artery stenosis
    Sala, Carla
    Rescaldani, Marta
    Burdick, Larry
    Danzi, Gian Battista
    [J]. JOURNAL OF CARDIOVASCULAR MEDICINE, 2015, 16 : S42 - S44
  • [44] Treatment of in-stent restenosis with sirolimus-eluting magnesium bioresorbable scaffolds: optical coherence tomography insights
    Cuesta, Javier
    Rivero, Fernando
    Bastante, Teresa
    Antuna, Paula
    Jimenez-Mendez, Cesar
    Garcia-Guimaraes, Marcos
    Alfonso, Fernando
    [J]. CORONARY ARTERY DISEASE, 2022, 33 (05) : 362 - 367
  • [45] Differences in tissue characterization of restenotic neointima between sirolimus-eluting stent and bare-metal stent: integrated backscatter intravascular ultrasound analysis for in-stent restenosis
    Ando, Hirohiko
    Amano, Tetsuya
    Takashima, Hiroaki
    Harada, Kazuhiro
    Kitagawa, Katsuhide
    Suzuki, Akihiro
    Kunimura, Ayako
    Shimbo, Yusaku
    Harada, Ken
    Yoshida, Tomohiro
    Kato, Bunichi
    Uetani, Tadayuki
    Kato, Masataka
    Matsubara, Tatsuaki
    Kumagai, Soichiro
    Yoshikawa, Daiji
    Isobe, Satoshi
    Ishii, Hideki
    Murohara, Toyoaki
    [J]. EUROPEAN HEART JOURNAL-CARDIOVASCULAR IMAGING, 2013, 14 (10) : 996 - 1001
  • [46] The Supralimus® sirolimus-eluting stent
    Lemos, Pedro A.
    Bienert, Igor
    [J]. EXPERT REVIEW OF MEDICAL DEVICES, 2013, 10 (03) : 295 - 300
  • [47] Intravascular ultrasound assessment of optimal stent area to prevent in-stent restenosis after zotarolimus-, everolimus-, and sirolimus-eluting stent implantation
    Song, Hae-Geun
    Kang, Soo-Jin
    Ahn, Jung-Min
    Kim, Won-Jang
    Lee, Jong-Young
    Park, Duk-Woo
    Lee, Seung-Whan
    Kim, Young-Hak
    Lee, Cheol Whan
    Park, Seong-Wook
    Park, Seung-Jung
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2014, 83 (06) : 873 - 878
  • [48] Restenosis and stent fracture following sirolimus-eluting stent (SES) implantation - A serial quantitative coronary angiography (QCA) and intravascular ultrasound (IVUS) study
    Okumura, Masanori
    Ozaki, Yukio
    Ishii, Junichi
    Kan, Shino
    Naruse, Hiroyuki
    Matsui, Shigeru
    Ishikawa, Makoto
    Hattori, Kousuke
    Gochi, Tomoko
    Nakano, Tadashi
    Yamada, Akira
    Kato, Shigeru
    Motoyama, Sadako
    Sarai, Masayoshi
    Takagi, Yasushi
    Ismail, Tevfik F.
    Nomura, Masanori
    Hishida, Hitoshi
    [J]. CIRCULATION JOURNAL, 2007, 71 (11) : 1669 - 1677
  • [49] Treatment of in-stent restenosis with sirolimus-eluting-stents:: results from the prospective German Cypher stent registry
    Moellmann, Helge
    Elsaesser, Albrecht
    Nef, Holger
    Schneider, Steffen
    Nienaber, Christoph A.
    Richardt, Gert
    Weber, Michael
    Kelm, Malte
    Levenson, Benny
    Bonzel, Tassilo
    Tebbe, Ulrich
    Sabin, Georg
    Pfannebecker, Thomas
    Senges, Jochen
    Hamm, Christian W.
    [J]. CLINICAL RESEARCH IN CARDIOLOGY, 2008, 97 (07) : 432 - 440
  • [50] Effects of sirolimus-eluting stent on calcified coronary lesions
    Jian-Jun, Li
    Bo, Xu
    Yue-Jin, Yang
    Ji-Lin, Chen
    Shu-Bing, Qiao
    Wei-Hua, Ma
    Xue-Wen, Qin
    Min, Yao
    Hai-Bo, Liu
    Yong-Jian, Wu
    Jin-Qing, Yuan
    Jue, Chen
    Shi-Jie, You
    Jun, Dai
    Ran, Xia
    Run-Lin, Gao
    [J]. CHINESE MEDICAL JOURNAL, 2008, 121 (01) : 6 - 11