Investigating the relationship between virtual cystoscopy image quality and CT slice thickness

被引:9
作者
Lalondrelle, S. [1 ,3 ]
Sohaib, S. A. [2 ]
Castellano, I. A. [4 ]
Mears, D. [2 ]
Huddart, R. [1 ,3 ]
Khoo, V. [1 ,3 ]
机构
[1] Royal Marsden NHS Fdn Trust, Acad Urol Unit, London, England
[2] Royal Marsden NHS Fdn Trust, Dept Radiol, London, England
[3] Inst Canc Res, London SW3 6JB, England
[4] Royal Marsden NHS Fdn Trust, Dept Phys, London, England
关键词
MATERIAL-FILLED BLADDER; MULTIDETECTOR ROW CT; COMPUTED-TOMOGRAPHY; COLONOGRAPHY; COMPLICATIONS; OPTIMIZATION; RESECTION; PHANTOM; POLYPS; TUMORS;
D O I
10.1259/bjr/99567374
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To investigate the effect of reconstruction slice thickness on image quality at CT virtual cystoscopy (VC). Methods: Pelvic CT examinations in bladder cancer patients were reconstructed at different slice thicknesses (0.6-5 mm) and intervals, and resulting VC images assessed. Quality indicators were ridging, holes, floaters and dimpling artefacts, tumour definition, and an overall score, ranked 1 (best) to 7 (worst). CT number and standard deviation (SD) for bladder contents and bladder wall were recorded. The mean SD was used as a measure of noise, and the contrast-to-noise ratio (CNR) was calculated as the CT number difference between them divided by the average image noise. The mean CNR across the three levels was used for analysis. Each qualitative image quality measure was compared with CT number, noise and CNR measurements. Results: Dimpling artefacts increased with thinner slice reconstruction and correlated with increased noise, often resulting in poor tumour definition. The best overall image quality score was seen for VC images reconstructed at 1.2 mm slice thickness, probably because of the competing effects of spatial resolution and CNR. Conclusion: A slice thickness reconstruction <1.2 mm does not provide for better image quality at VC owing to the presence of increased noise.
引用
收藏
页码:1112 / 1117
页数:6
相关论文
共 26 条
[21]   Bladder tumor detection at virtual cytoscopy [J].
Song, JH ;
Francis, IR ;
Platt, JF ;
Cohan, RH ;
Mohsin, J ;
Kielb, SJ ;
Korobkin, M ;
Montie, JE .
RADIOLOGY, 2001, 218 (01) :95-100
[22]   16-MDCT cystoscopy in the evaluation of neoplasms of the urinary bladder [J].
Tsampoulas, Constantine ;
Tsili, Athina C. ;
Giannakis, Dimitrios ;
Alamanos, Yiannis ;
Sofikitis, Nikolaos ;
Efremidis, Stavros C. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (03) :729-735
[23]   Computed tomographic virtual cystoscopy for the detection of urinary bladder neoplasms [J].
Tsili, AC ;
Tsampoulas, C ;
Chatziparaskevas, N ;
Silakos, A ;
Kalef-Ezra, J ;
Sofikitis, N ;
Efremidis, SC .
EUROPEAN UROLOGY, 2004, 46 (05) :579-585
[24]   CT colonography: Protocol optimization with multi-detector row CT - Study in an anthropomorphic colon phantom [J].
Wessling, J ;
Fischbach, R ;
Meier, N ;
Allkemper, T ;
Klusmeier, J ;
Ludwig, K ;
Heindel, W .
RADIOLOGY, 2003, 228 (03) :753-759
[25]   Detection of colorectal polyps: Comparison of multi-detector row CT and MR colonography in a colon phantom [J].
Wessling, Johannes ;
Fischbach, Roman ;
Borchert, Alexandra ;
Kugel, Harald ;
Allkemper, Thomas ;
Osada, Nani ;
Heindel, Walter .
RADIOLOGY, 2006, 241 (01) :125-131
[26]   Virtual cystoscopy in the evaluation of bladder tumors [J].
Yazgan, C ;
Fitoz, S ;
Atasoy, C ;
Turkolmez, K ;
Yagci, C ;
Akyar, S .
CLINICAL IMAGING, 2004, 28 (02) :138-142