FRoG: An independent dose and LETd prediction tool for proton therapy at ProBeam® facilities

被引:19
作者
Kopp, Benedikt [1 ,2 ,3 ,4 ,5 ,6 ,7 ]
Jensen, Maria Fuglsang [8 ]
Mein, Stewart [1 ,2 ,3 ,4 ,5 ,6 ]
Hoffmann, Lone [9 ]
Nystrom, Hakan [8 ,10 ]
Falk, Marianne [8 ]
Haberer, Thomas [11 ]
Abdollahi, Amir [1 ,2 ,3 ,4 ,5 ,6 ]
Debus, Jurgen [5 ,6 ,7 ,11 ,12 ,13 ,14 ]
Mairani, Andrea [11 ,15 ]
机构
[1] Heidelberg Univ Hosp UKHD, Natl Ctr Tumor Dis NCT, Clin Cooperat Unit Translat Radiat Oncol, D-69120 Heidelberg, Germany
[2] German Canc Res Ctr, D-69120 Heidelberg, Germany
[3] Heidelberg Fac Med MFHD, Div Mol & Translat Radiat Oncol, Dept Radiat Oncol, D-69120 Heidelberg, Germany
[4] Heidelberg Univ Hosp UKHD, Heidelberg Ion Beam Therapy Ctr HIT, D-69120 Heidelberg, Germany
[5] German Canc Res Ctr, German Canc Consortium DKTK Core Ctr Heidelberg, D-69120 Heidelberg, Germany
[6] Heidelberg Univ, Natl Ctr Radiat Oncol NCRO, Heidelberg Inst Radiat Oncol HIRO, Clin Cooperat Unit Radiat Oncol, D-69120 Heidelberg, Germany
[7] Heidelberg Univ, Dept Phys & Astron, D-69120 Heidelberg, Germany
[8] Aarhus Univ Hosp, Danish Ctr Particle Therapy, DK-8200 Aarhus, Denmark
[9] Aarhus Univ Hosp, Dept Med Phys, DK-8200 Aarhus, Denmark
[10] Skandion Clin, S-75237 Uppsala, Sweden
[11] Heidelberg Univ Hosp, Heidelberg Ion Beam Therapy Ctr HIT, Dept Radiat Oncol, D-69120 Heidelberg, Germany
[12] Heidelberg Univ Hosp, Dept Radiat Oncol, D-69120 Heidelberg, Germany
[13] Heidelberg Inst Radiat Oncol HIRO, D-69120 Heidelberg, Germany
[14] Natl Ctr Tumor Dis NCT, D-69120 Heidelberg, Germany
[15] Natl Ctr Oncol Hadrontherapy CNAO, Med Phys, I-27100 Pavia, Italy
关键词
FRoG; independent dose engine; ion-beam therapy; LETd; robustness analysis; RELATIVE BIOLOGICAL EFFECTIVENESS; MONTE-CARLO SIMULATIONS; SCANNED PROTON; ION THERAPY; GPU; CODE; OPTIMIZATION; MODEL;
D O I
10.1002/mp.14417
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Particle therapy is becoming increasingly available world-wide for precise tumor targeting, its favorable depth dose deposition, and increased biological damage to tumor tissue compared to conventional photon therapy. As demand increases for improved robustness and conformality, next-generation secondary dose calculation engines are needed to verify treatment plans independently and provide estimates for clinical decision-making factors, such as dose-averaged linear energy transfer (LETd) and relative biological effectiveness (RBE). Method: FRoG (Fast dose Recalculation on GPU) has been installed and commissioned at the Danish Centre for Particle Therapy (DCPT). FRoG was developed for synchrotron-based facilities and has previously demonstrated good agreement with gold-standard Monte Carlo simulations and measurements. In this work, additions and modifications to FRoG's pencil beam algorithm to support the ion beam delivery with cyclotron-based technology as used at the DCPT, range shifter (RS) implementation, and robustness analysis methods are presented. FRoG dose predictions are compared to measurements and predictions of the clinical treatment planning system (TPS) Eclipse (Varian Medical Systems, Palo Alto, United States of America, CA, v.13.7.16) in both homogenous and heterogeneous scenarios using a solid-water/water and a half-head anthropomorphic phantom, respectively. Additional capabilities of FRoG are explored by performing a plan robustness analysis, analyzing dose and LETd for ten patients. Results: Mid-target measurements in spread-out Bragg Peaks (SOBP) were on average within -0.19% +/- 0.30% and <= 0.5% of FRoG predictions for irradiations without and with RS, respectively. Average 3%/2mm 3D gamma-analysis passing rates were 99.1% for similar to 200 patient plan QA comparisons. Measurement with an anthropomorphic head-phantom yielded a gamma-passing rate >98%. Overall, maximum target differences in D-02% of <2% between the TPS and FRoG were observed for patient plans. The robustness analysis study accounting for range, delivery, and positioning uncertainties revealed small differences in target dose and a maximum LETdVH02%(LET (d)received by 2% of the volume having dose larger than 1% of maximum dose) values below 10.1 keV/mu m to the brain stem. Conclusion: We demonstrate that auxiliary dose calculation systems like FRoG can yield excellent agreement to measurements comparable to clinical beam models. Through this work, application of FRoG as a secondary engine at third party cyclotron-based particle treatment facilities is now established for dose verification as well as providing further insight on LETd and variable RBE distributions for protons, currently absent from the standard clinical TPS. (c) 2020 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine. [https://doi.org/10.1002/mp.14417]
引用
收藏
页码:5274 / 5286
页数:13
相关论文
共 32 条
[1]   Integration and evaluation of automated Monte Carlo simulations in the clinical practice of scanned proton and carbon ion beam therapy [J].
Bauer, J. ;
Sommerer, F. ;
Mairani, A. ;
Unholtz, D. ;
Farook, R. ;
Handrack, J. ;
Frey, K. ;
Marcelos, T. ;
Tessonnier, T. ;
Ecker, S. ;
Ackermann, B. ;
Ellerbrock, M. ;
Debus, J. ;
Parodi, K. .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (16) :4635-4659
[2]   The FLUKA Code: Developments and Challenges for High Energy and Medical Applications [J].
Boehlen, T. T. ;
Cerutti, F. ;
Chin, M. P. W. ;
Fosso, A. ;
Ferrari, A. ;
Ortega, P. G. ;
Mairani, A. ;
Sala, P. R. ;
Smirnov, G. ;
Vlachoudisl, V. .
NUCLEAR DATA SHEETS, 2014, 120 :211-214
[3]   Including robustness in multi-criteria optimization for intensity-modulated proton therapy [J].
Chen, Wei ;
Unkelbach, Jan ;
Trofimov, Alexei ;
Madden, Thomas ;
Kooy, Hanne ;
Bortfeld, Thomas ;
Craft, David .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (03) :591-608
[4]   FRoG-A New Calculation Engine for Clinical Investigations with Proton and Carbon Ion Beams at CNAO [J].
Choi, KyungDon ;
Mein, Stewart B. ;
Kopp, Benedikt ;
Magro, Giuseppe ;
Molinelli, Silvia ;
Ciocca, Mario ;
Mairani, Andrea .
CANCERS, 2018, 10 (11)
[5]   Sub-second pencil beam dose calculation on GPU for adaptive proton therapy [J].
da Silva, Joakim ;
Ansorge, Richard ;
Jena, Rajesh .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (12) :4777-4795
[6]  
Ferrari A., 2005, Fluka, CERN-Libr., V55, P100, DOI DOI 10.2172/877507
[7]   THE LINEAR-QUADRATIC FORMULA AND PROGRESS IN FRACTIONATED RADIOTHERAPY [J].
FOWLER, JF .
BRITISH JOURNAL OF RADIOLOGY, 1989, 62 (740) :679-694
[8]   Validation of a GPU-based Monte Carlo code (gPMC) for proton radiation therapy: clinical cases study [J].
Giantsoudi, Drosoula ;
Schuemann, Jan ;
Jia, Xun ;
Dowdell, Stephen ;
Jiang, Steve ;
Paganetti, Harald .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (06) :2257-2269
[9]   The heidelberg ion therapy center [J].
Haberer, T ;
Debus, J ;
Eickhoff, H ;
Jäkel, O ;
Schulz-Ertner, D ;
Weber, U .
RADIOTHERAPY AND ONCOLOGY, 2004, 73 :S186-S190
[10]   Modeling late effects in hypofractionated stereotactic radiotherapy [J].
Hoban, PW ;
Jones, LC ;
Clark, BG .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (01) :199-210