Antipsychotics; Protein precipitation; UHPLC-MS/MS; Method validation; Therapeutic drug monitoring; SOLID-PHASE EXTRACTION; ELECTROSPRAY-IONIZATION; GAS-CHROMATOGRAPHY; HUMAN SERUM; QUANTITATIVE-DETERMINATION; CHEMICAL-IONIZATION; PSYCHOTROPIC-DRUGS; HALOPERIDOL; METABOLITES; ZUCLOPENTHIXOL;
D O I
10.1016/j.jpba.2013.07.041
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A selective and sensitive method was developed for the simultaneous quantification of seven typical antipsychotic drugs (cis-chlorprothixene, flupentixol, haloperidol, levomepromazine, pipamperone, promazine and zuclopenthixol) in human plasma. Ultra-high performance liquid chromatography (UHPLC) was used for complete separation of the compounds in less than 4.5 min on an Acquity UPLC BEH C18 column (2.1 mm x 50 mm; 1.7 mu m) with a gradient elution of ammonium formate buffer pH 4.0 and acetonitrile at a flow rate of 400 mu l/min. Detection was performed on a tandem quadrupole mass spectrometer (MS/MS) equipped with an electrospray ionization interface. A simple protein precipitation procedure with acetonitrile was used for sample preparation. Thanks to the use of stable isotope-labeled internal standards for all analytes, internal standard-normalized matrix effects were in the range of 92-108%. The method was fully validated to cover large concentration ranges of 0.2-90 ng/ml for haloperidol, 0.5-90 ng/ml for flupentixol, 1-450 ng/ml for levomepromazine, promazine and zuclopenthixol and 2-900 ng/ml for cis-chlorprothixene and pipamperone. Trueness (89.1-114.8%), repeatability (1.8-9.9%), intermediate precision (1.9-16.3%) and accuracy profiles (<30%) were in accordance with the latest international recommendations. The method was successfully used in our laboratory for routine quantification of more than 500 patient plasma samples for therapeutic drug monitoring. To the best of our knowledge, this is the first UHPLC-MS/MS method for the quantification of the studied drugs with a sample preparation based on protein precipitation. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:36 / 44
页数:9
相关论文
共 50 条
  • [31] Quantification of clenbuterol at trace level in human urine by ultra-high pressure liquid chromatography-tandem mass spectrometry
    Nicoli, Raul
    Petrou, Michael
    Badoud, Flavia
    Dvorak, Jiri
    Saugy, Martial
    Baume, Norbert
    JOURNAL OF CHROMATOGRAPHY A, 2013, 1292 : 142 - 150
  • [32] Determination of trihexyphenidyl in human plasma by ultra high performance liquid chromatography-tandem mass spectrometry
    Yang Yufei
    Xia Yunyan
    Wu Sha
    Zou Qiaogen
    CHINESE JOURNAL OF CHROMATOGRAPHY, 2019, 37 (12) : 1291 - 1296
  • [33] Rapid liquid chromatography tandem mass spectrometry determination of rivaroxaban levels in human plasma for therapeutic drug monitoring
    Varga, Andreea
    Serban, Razvan Constantin
    Muntean, Daniela Lucia
    Tatar, Cristina Maria
    Farczadi, Lenard
    Tilea, Ioan
    REVISTA ROMANA DE MEDICINA DE LABORATOR, 2017, 25 (02): : 145 - 155
  • [34] A multiresidue approach for the simultaneous quantification of antibiotics in macroalgae by ultra-high performance liquid chromatography-tandem mass spectrometry
    Leston, Sara
    Freitas, Andreia
    Rosa, Joao
    Barbosa, Jorge
    Lemos, Marco F. L.
    Pardal, Miguel Angelo
    Ramos, Fernando
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2016, 1033 : 361 - 367
  • [35] Development and validation of an ultra-high performance liquid chromatography-tandem mass spectrometry method to measure creatinine in human urine
    Fraselle, S.
    De Cremer, K.
    Coucke, W.
    Glorieux, G.
    Vanmassenhove, J.
    Schepers, E.
    Neirynck, N.
    Van Overmeire, I.
    Van Loco, J.
    Van Biesen, W.
    Vanholder, R.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2015, 988 : 88 - 97
  • [36] Ultra-Performance Liquid Chromatography-Tandem Mass Spectrometry for Simultaneous Determination of Antipsychotic Drugs in Human Plasma and Its Application in Therapeutic Drug Monitoring
    Qi, Yingjie
    Liu, Guangxuan
    DRUG DESIGN DEVELOPMENT AND THERAPY, 2021, 15 : 463 - 479
  • [37] Simultaneous quantification of atenolol and chlorthalidone in human plasma by ultra-performance liquid chromatography-tandem mass spectrometry
    Shah, Jaivik V.
    Patel, Daxesh P.
    Shah, Priyanka A.
    Sanyal, Mallika
    Shrivastav, Pranav S.
    BIOMEDICAL CHROMATOGRAPHY, 2016, 30 (02) : 208 - 216
  • [38] Separation and quantification of two diastereomers of a Drug Candidate in rat plasma by ultra-high pressure liquid chromatography mass spectrometry
    Wang, Haiping
    Edom, Richard W.
    Kumar, Sanjeev
    Vincent, Stella
    Shen, Zhongzhou
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 854 (1-2): : 26 - 34
  • [39] A NEW STRATEGY FOR PENTACHLOROPHENOL MONITORING IN WATER SAMPLES USING ULTRA-HIGH PERFORMANCE LIQUID CHROMATOGRAPHY - MASS SPECTROMETRY TANDEM
    Kadmi, Yassine
    Favier, Lidia
    Harja, Maria
    Simion, Andrei Ionut
    Rusu, Lacramioara
    Wolbert, Dominique
    ENVIRONMENTAL ENGINEERING AND MANAGEMENT JOURNAL, 2015, 14 (03): : 567 - 574
  • [40] Therapeutic drug monitoring of amikacin: quantification in plasma by liquid chromatography-tandem mass spectrometry and work experience of clinical pharmacists
    Xu, Lijie
    Cheng, Xuefang
    Zhu, Guanhua
    Hu, Juanni
    Li, Qin
    Fan, Guorong
    EUROPEAN JOURNAL OF HOSPITAL PHARMACY, 2022, 29 (E1) : e77 - e82