A clinical comparison of the Flexiblade™ and macintosh laryngoscopes for laryngeal exposure in anesthetized adults

被引:7
作者
Cheung, RWW
Irwin, MG
Law, BCW
Chan, CK
机构
[1] Queen Elizabeth Hosp, Dept Anaesthesia, Hong Kong, Hong Kong, Peoples R China
[2] Univ Hong Kong, Queen Elizabeth Hosp, Dept Anaesthesiol, Hong Kong, Hong Kong, Peoples R China
关键词
D O I
10.1213/01.ane.0000185035.89265.3c
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The Flexiblade(TM) is a laryngoscope with a flexible blade. To evaluate the efficacy of the Flexiblade(TM) compared with the classic Macintosh laryngoscope, we performed a clinical study in 200 paralyzed patients undergoing elective surgery requiring general anesthesia and endotracheal intubation. Direct laryngoscopy was performed with a size 3 Macintosh laryngoscope and the Flexiblade(TM), with and without activation of the lever. The laryngeal views were recorded, without manipulation, according to the Cormack and Lehane classification. No laryngoscopic view obtained by the Macintosh blade was worse than that obtained by the Flexiblade(TM) without the lever activated. The Macintosh blade improved 58.5% of non-Grade I views obtained by the Flexiblade(TM) with its lever not activated. However, when the Flexiblade(TM) lever was activated, 39.6% of non-Grade I views obtained by the Macintosh blade were improved, whereas 84.5% of non-Grade I views obtained by the inactivated Flexiblade(TM) were improved. Activating the Flexiblade(TM) lever never caused a deterioration of view. In only one case was the view better with the Macintosh blade than that with the activated Flexiblade(TM). We conclude that the Flexiblade(TM), after lever activation, is significantly better than the Macintosh laryngoscope for laryngeal visualization in paralyzed adults (P < 0.0001).
引用
收藏
页码:626 / 630
页数:5
相关论文
共 14 条
[1]  
Arino JJ, 2003, CAN J ANAESTH, V50, P501, DOI 10.1007/BF03021064
[2]   CRITERIA FOR ESTIMATING LIKELIHOOD OF DIFFICULTY OF ENDOTRACHEAL INTUBATION WITH THE MACINTOSH LARYNGOSCOPE [J].
BELLHOUSE, CP ;
DORE, C .
ANAESTHESIA AND INTENSIVE CARE, 1988, 16 (03) :329-337
[3]   DIFFICULT TRACHEAL INTUBATION IN OBSTETRICS [J].
CORMACK, RS ;
LEHANE, J .
ANAESTHESIA, 1984, 39 (11) :1105-1111
[4]  
Groom P, 1997, ANAESTHESIA, V52, P802
[5]   DIFFICULT LARYNGOSCOPY MADE EASY WITH A BURP [J].
KNILL, RL .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1993, 40 (03) :279-282
[6]   The use of the McCoy laryngoscope in patients with simulated cervical spine injuries [J].
Laurent, SC ;
deMelo, AE ;
AlexanderWilliams, JM .
ANAESTHESIA, 1996, 51 (01) :74-75
[7]  
Law BCW, 2001, ANAESTHESIA, V56, P914
[8]   A CLINICAL SIGN TO PREDICT DIFFICULT TRACHEAL INTUBATION - A PROSPECTIVE-STUDY [J].
MALLAMPATI, SR ;
GATT, SP ;
GUGINO, LD ;
DESAI, SP ;
WARAKSA, B ;
FREIBERGER, D ;
LIU, PL .
CANADIAN ANAESTHETISTS SOCIETY JOURNAL, 1985, 32 (04) :429-434
[9]   AN ANALYSIS OF LARYNGOSCOPE BLADE SHAPE AND DESIGN - NEW CRITERIA FOR LARYNGOSCOPE EVALUATION [J].
MARKS, RRD ;
HANCOCK, R ;
CHARTERS, P .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1993, 40 (03) :262-270
[10]   Flexiblade laryngoscope [J].
Moynham, D .