An iterative block-shifting approach to retention time alignment that preserves the shape and area of gas chromatography-mass spectrometry peaks

被引:26
作者
Chae, Minho [1 ,2 ]
Reis, Robert J. Shmookler [2 ,3 ,4 ]
Thaden, John J. [2 ,4 ]
机构
[1] Univ Arkansas, UALR UAMS Joint Grad Program Bioinformat, Little Rock, AR 72204 USA
[2] Univ Arkansas Med Sci, Dept Geriatr, Little Rock, AR 72205 USA
[3] Univ Arkansas Med Sci, Dept Biochem & Mol Biol, Little Rock, AR 72205 USA
[4] Cent Arkansas Vet Healthcare Syst LRVA 151, Little Rock, AR 72205 USA
关键词
D O I
10.1186/1471-2105-9-S9-S15
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Background: Metabolomics, petroleum and biodiesel chemistry, biomarker discovery, and other fields which rely on high-resolution profiling of complex chemical mixtures generate datasets which contain millions of detector intensity readings, each uniquely addressed along dimensions of time (e.g., retention time of chemicals on a chromatographic column), a spectral value (e.g., mass-to-charge ratio of ions derived from chemicals), and the analytical run number. They also must rely on data preprocessing techniques. In particular, inter-run variance in the retention time of chemical species poses a significant hurdle that must be cleared before feature extraction, data reduction, and knowledge discovery can ensue. Alignment methods, for calibrating retention reportedly (and in our experience) can misalign matching chemicals, falsely align distinct ones, be unduly sensitive to chosen values of input parameters, and result in distortions of peak shape and area. Results: We present an iterative block-shifting approach for retention-time calibration that detects chromatographic features and qualifies them by retention time, spectrum, and the effect of their inclusion on the quality of alignment itself. Mass chromatograms are aligned pairwise to one selected as a reference. In tests using a 45-run GC-MS experiment, block-shifting reduced the absolute deviation of retention by greater than 30-fold. It compared favourably to COW and XCMS with respect to alignment, and was markedly superior in preservation of peak area. Conclusion: Iterative block-shifting is an attractive method to align GC-MS mass chromatograms that is also generalizable to other two-dimensional techniques such as HPLC-MS.
引用
收藏
页数:11
相关论文
共 25 条
  • [1] MET-IDEA: Data extraction tool for mass spectrometry-based metabolomics
    Broeckling, Corey D.
    Reddy, Indira R.
    Duran, Anthony L.
    Zhao, Xuechun
    Sumner, Lloyd W.
    [J]. ANALYTICAL CHEMISTRY, 2006, 78 (13) : 4334 - 4341
  • [2] Chromatographic alignment by warping and dynamic programming as a pre-processing tool for PARAFAC modelling of liquid chromatography-mass spectrometry data
    Bylund, D
    Danielsson, R
    Malmquist, G
    Markides, KE
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2002, 961 (02) : 237 - 244
  • [3] Mass spectrometry-based metabolomics
    Dettmer, Katja
    Aronov, Pavel A.
    Hammock, Bruce D.
    [J]. MASS SPECTROMETRY REVIEWS, 2007, 26 (01) : 51 - 78
  • [4] DRAPER J, 2006, 2 INT SCI M MET SOC
  • [5] Improved peak detection in mass spectrum by incorporating continuous wavelet transform-based pattern matching
    Du, Pan
    Kibbe, Warren A.
    Lin, Simon M.
    [J]. BIOINFORMATICS, 2006, 22 (17) : 2059 - 2065
  • [6] Metabolomics spectral formatting, alignment and conversion tools (MSFACTs)
    Duran, AL
    Yang, J
    Wang, LJ
    Sumner, LW
    [J]. BIOINFORMATICS, 2003, 19 (17) : 2283 - 2293
  • [7] Classification of weathered petroleum oils by multi-way analysis of gas chromatography-mass spectrometry data using PARAFAC2 parallel factor analysis
    Ebrahimi, Diako
    Li, Jianfeng
    Hibbert, David Brynn
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2007, 1166 (1-2) : 163 - 170
  • [8] Parametric time warping
    Eilers, PHC
    [J]. ANALYTICAL CHEMISTRY, 2004, 76 (02) : 404 - 411
  • [9] High-speed peak matching algorithm for retention time alignment of gas chromatographic data for chemometric analysis
    Johnson, KJ
    Wright, BW
    Jarman, KH
    Synovec, RE
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2003, 996 (1-2) : 141 - 155
  • [10] MZmine: toolbox for processing and visualization of mass spectrometry based molecular profile data
    Katajamaa, M
    Miettinen, J
    Oresic, M
    [J]. BIOINFORMATICS, 2006, 22 (05) : 634 - 636