Cardiac Perforation and Lead Dislodgement after Implantation of a MR-Conditional Pacing Lead: A Single-Center Experience

被引:19
作者
Elmouchi, Darryl A. [1 ,2 ]
Rosema, Shelly [1 ]
Vanoosterhout, Stacie M. [1 ]
Khan, Mohsin [3 ]
Davis, Alan T. [2 ]
Gauri, Andre J. [1 ,2 ]
Finta, Bohuslav [1 ,2 ]
Woelfel, Alan K. [1 ,2 ]
Chalfoun, Nagib T. [1 ,2 ]
机构
[1] Spectrum Hlth, Frederik Meijer Heart & Vasc Inst, Grand Rapids, MI USA
[2] Michigan State Univ, Coll Human Med, Dept Internal Med, Grand Rapids, MI 49525 USA
[3] Grand Rapids Med Educ Partners, Grand Rapids, MI USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2014年 / 37卷 / 01期
关键词
pacemakers; magnetic resonance imaging; cardiac tamponade; complications; PACEMAKER; COMPLICATIONS; SYSTEM; PREDICTORS; SAFETY;
D O I
10.1111/pace.12293
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
ObjectiveThe purpose of our study was to determine if the rate of lead-related complications was increased with the Medtronic CapSureFix MRI SureScan 5086 MRI pacing lead (5086; Medtronic Inc., Minneapolis, MN, USA) compared to the previous generation of Medtronic CapSureFix Novus 5076 pacing lead (5076). BackgroundThe 5086 lead is a newly introduced active-fixation pacemaker lead designed to be used conditionally in a magnetic resonance (MR) scanner. This lead has specific design changes compared to the previous generation of 5076 pacing leads. MethodsThis study was a retrospective case control study of 65 consecutive patients implanted with two 5086 leads compared to 92 consecutive control patients implanted with two 5076 leads over a 14-month period at a high-volume tertiary care hospital. ResultsPericarditis, pericardial effusion, cardiac tamponade, or death within 30 days of implant were seen in eight patients from the 5086 cohort and two from the 5076 cohort (odds ratio 6.3, 95% confidence interval 1.3-30.8, P = 0.02). Lead dislodgement occurred in four of the 5086 patients and in none of the 5076 patients (P < 0.03). ConclusionsIn a high-volume center, the incidence of pericarditis, cardiac perforation, tamponade, death, and lead dislodgement was significantly higher with the MR-conditional Medtronic 5086 lead when compared to the previous generation Medtronic 5076 lead.
引用
收藏
页码:4 / 10
页数:7
相关论文
共 18 条
  • [1] AGGARWAL RK, 1995, BRIT HEART J, V73, P571
  • [2] 50th Anniversary of the First Successful Permanent Pacemaker Implantation in the United States: Historical Review and Future Directions
    Beck, Hiroko
    Boden, William E.
    Patibandla, Sushmitha
    Kireyev, Dmitriy
    Gupta, Vipul
    Campagna, Franklin
    Cain, Michael E.
    Marine, Joseph E.
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2010, 106 (06) : 810 - 818
  • [3] Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator
    Danik, Stephan B.
    Mansour, Moussa
    Singh, Jagmeet
    Reddy, Vivek Y.
    Ellinor, Patrick T.
    Milan, David
    Heist, E. Kevin
    d'Avila, Andre
    Ruskin, Jeremy N.
    Mela, Theofanie
    [J]. HEART RHYTHM, 2007, 4 (04) : 439 - 442
  • [4] Clinical and electrical performance of currently available MRI-safe pacing systems. Do all devices perform in the same way?
    Forleo, Giovanni B.
    Della Rocca, Domenico G.
    Panattoni, Germana
    Schirripa, Valentina
    Sergi, Domenico
    Magliano, Giulia
    Tesauro, Manfredi
    Santini, Luca
    Romeo, Francesco
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 2013, 167 (05) : 2340 - 2341
  • [5] Acute Left Hemothorax as a Late Complication of an Active-Fixation Pacemaker Lead
    Forleo, Giovanni B.
    Zeitani, Jacob
    Perretta, Tommaso
    Della Rocca, Domenico G.
    Santini, Luca
    Simonetti, Giovanni
    Romeo, Francesco
    [J]. ANNALS OF THORACIC SURGERY, 2013, 95 (03) : 1081 - 1084
  • [6] Safety and efficacy of a new magnetic resonance imaging-compatible pacing system: Early results of a prospective comparison with conventional dual-chamber implant outcomes
    Forleo, Giovanni B.
    Santini, Luca
    Della Rocca, Domenico G.
    Romano, Valentina
    Papavasileiou, Lida P.
    Magliano, Giulia
    Sgueglia, Marianna
    Romeo, Francesco
    [J]. HEART RHYTHM, 2010, 7 (06) : 750 - 754
  • [7] Randomized trial of pacemaker and lead system for safe scanning at 1.5 Tesla
    Gimbel, J. Rod
    Bello, David
    Schmitt, Matthias
    Merkely, Bela
    Schwitter, Juerg
    Hayes, David L.
    Sommer, Torsten
    Schloss, Edward J.
    Chang, Yanping
    Willey, Sarah
    Kanal, Emanuel
    [J]. HEART RHYTHM, 2013, 10 (05) : 685 - 691
  • [8] Delayed lead perforation: A disturbing trend
    Khan, MN
    Joseph, G
    Khaykin, Y
    Ziada, KM
    Wilkoff, BL
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (03): : 251 - 253
  • [9] First MRI-Safe Pacemaker Receives Conditional Approval From FDA
    Mitka, Mike
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2011, 305 (10): : 985 - 986
  • [10] How to perform magnetic resonance imaging on patients with implantabte cardiac arrhythmia devices
    Nazarian, Saman