Optimization of the Coil to Focalize the Electrical Field Induced by Magnetic Stimulation in the Human Brain

被引:0
|
作者
Pu, Lina [1 ]
Yin, Tao [1 ]
An, Hao [1 ]
Li, Song [1 ]
Liu, Zhipeng [1 ]
机构
[1] Chinese Acad Med Sci, Peking Union Med Coll, Inst Biomed Engn, Tianjin, Peoples R China
关键词
magnetic stimulating coil; focus; induced electrical field; comparing coefficient;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The design of the magnetic stimulating coil is one of the key technique in the magnetic stimulating system, it is the hotspot in the research of magnetic stimulation. A coefficient to evaluate the focus of the electrical field intensity induced by magnetic stimulation is firstly presented in this study, which was used as an index of focus, Matlab was used as the tool to simulate the distribution of induced electrical field, result is obtained that the four-leaf rose shape coil is the best compared with other five different shapes of coil. As a result of optimizing other parameters of the coil, the smaller radius of the coil is the better choice and the angle between four-leaf rose coil of 102 degree is better than other value. Furthermore, when the four-leaf coil with angle of 112 degree assembled with a coil of 89, the more focused electrical field is obtained. The coefficient proposed in this study may be a guideline to the optimization of the coil's parameters.
引用
收藏
页码:366 / 372
页数:7
相关论文
共 50 条
  • [41] RESPONSES OF A HUMAN HAND MUSCLE TO CUTANEOUS STIMULATION AND MAGNETIC BRAIN-STIMULATION, BUT NOT ELECTRICAL BRAIN-STIMULATION, ARE TASK-DEPENDENT
    DATTA, AK
    HARRISON, LM
    STEPHENS, JA
    JOURNAL OF PHYSIOLOGY-LONDON, 1988, 401 : P47 - P47
  • [42] COIL PLACEMENT IN MAGNETIC BRAIN-STIMULATION RELATED TO SKULL AND BRAIN ANATOMY
    MEYER, BU
    BRITTON, TC
    KLOTEN, H
    STEINMETZ, H
    BENECKE, R
    ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1991, 81 (01): : 38 - 46
  • [43] Effect of stimulation frequency on hippocampal electric field induced by deep-brain magnetic stimulation
    Xu, Xiang-yang
    Deng, Bin
    Wang, Jiang
    Yi, Guo-sheng
    AIP ADVANCES, 2023, 13 (01)
  • [44] Design and field calculation of coil array for functional magnetic stimulation
    Liu, Ji-Cheng
    Huang, Ka-Ma
    Hu, Ya-Yi
    Hua, Wei
    Hangtian Yixue Yu Yixue Gongcheng/Space Medicine and Medical Engineering, 2004, 17 (05):
  • [45] MAGNETIC COIL VERSUS ELECTRICAL-STIMULATION OF MONKEY MOTOR CORTEX
    AMASSIAN, VE
    QUIRK, GJ
    STEWART, M
    JOURNAL OF PHYSIOLOGY-LONDON, 1987, 394 : P119 - P119
  • [46] Modelling of eddy currents induced in human brain by transcranial magnetic stimulation
    Institute of Electrical Engineering, Warsaw, Poland
    不详
    不详
    COMPEL Int J Comput Math Electr Electron Eng, 2 (257-260):
  • [47] Modelling of eddy currents induced in human brain by transcranial magnetic stimulation
    Krawczyk, A
    Wiak, S
    Zyss, T
    COMPEL-THE INTERNATIONAL JOURNAL FOR COMPUTATION AND MATHEMATICS IN ELECTRICAL AND ELECTRONIC ENGINEERING, 1998, 17 (1-3) : 257 - 260
  • [48] Histological alterations induced by electrode implantation and electrical stimulation in the human brain: A review
    Van Kuyck, Kris
    Welkenhuysen, Marleen
    Arckens, Lutgarde
    Sciot, Raf
    Nuttin, Bart
    NEUROMODULATION, 2007, 10 (03): : 244 - 261
  • [49] Effects of coil orientation and magnetic field shield on transcranial magnetic stimulation in cats
    Nakatoh, S
    Kitagawa, H
    Kawaguchi, Y
    Nakamura, H
    Takano, H
    Tsuji, H
    MUSCLE & NERVE, 1998, 21 (09) : 1172 - 1180
  • [50] Distortion of the electric field distribution induced in the brain during transcranial magnetic stimulation
    Kim, D. -H.
    Choi, N. -S.
    Won, C.
    Georghiou, G. E.
    IET SCIENCE MEASUREMENT & TECHNOLOGY, 2010, 4 (01) : 12 - 20