Optimization of the Coil to Focalize the Electrical Field Induced by Magnetic Stimulation in the Human Brain

被引:0
|
作者
Pu, Lina [1 ]
Yin, Tao [1 ]
An, Hao [1 ]
Li, Song [1 ]
Liu, Zhipeng [1 ]
机构
[1] Chinese Acad Med Sci, Peking Union Med Coll, Inst Biomed Engn, Tianjin, Peoples R China
关键词
magnetic stimulating coil; focus; induced electrical field; comparing coefficient;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The design of the magnetic stimulating coil is one of the key technique in the magnetic stimulating system, it is the hotspot in the research of magnetic stimulation. A coefficient to evaluate the focus of the electrical field intensity induced by magnetic stimulation is firstly presented in this study, which was used as an index of focus, Matlab was used as the tool to simulate the distribution of induced electrical field, result is obtained that the four-leaf rose shape coil is the best compared with other five different shapes of coil. As a result of optimizing other parameters of the coil, the smaller radius of the coil is the better choice and the angle between four-leaf rose coil of 102 degree is better than other value. Furthermore, when the four-leaf coil with angle of 112 degree assembled with a coil of 89, the more focused electrical field is obtained. The coefficient proposed in this study may be a guideline to the optimization of the coil's parameters.
引用
收藏
页码:366 / 372
页数:7
相关论文
共 50 条
  • [21] Numerical simulation of electromagnetic field distribution induced in brain by electrical stimulation
    Kim, H. J.
    Meng, Z. J.
    Sajib, S. Z. K.
    Chauhan, M.
    Jeong, W. C.
    Wi, H.
    Kwon, O. I.
    Woo, E. J.
    Oh, T. I.
    ELECTRONICS LETTERS, 2014, 50 (15) : 1045 - 1046
  • [22] PERCUTANEOUS STIMULATION OF THE HUMAN-BRAIN - A COMPARISON OF ELECTRICAL AND MAGNETIC STIMULI
    HESS, CW
    MILLS, KR
    MURRAY, NMF
    JOURNAL OF PHYSIOLOGY-LONDON, 1986, 378 : P35 - P35
  • [23] DIFFERENT SITES OF ACTION OF ELECTRICAL AND MAGNETIC STIMULATION OF THE HUMAN-BRAIN
    DAY, BL
    THOMPSON, PD
    DICK, JP
    NAKASHIMA, K
    MARSDEN, CD
    NEUROSCIENCE LETTERS, 1987, 75 (01) : 101 - 106
  • [24] OPTIMIZATION OF COIL DESIGN FOR NEURONAL EXCITATION BY MAGNETIC STIMULATION
    FERGUSON, AS
    DURAND, D
    DALBASTI, T
    IMAGES OF THE TWENTY-FIRST CENTURY, PTS 1-6, 1989, 11 : 1254 - 1255
  • [25] Coil Arrays Modeling and optimization for Transcranial Magnetic Stimulation
    Liu Jicheng
    Lu Jiqing
    Liu Cui
    Hu Yayi
    PROCEEDINGS OF THE 2009 2ND INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS, VOLS 1-4, 2009, : 922 - +
  • [26] Dependence of Electromagnetic Fields in Human Brain on Coil Bending Angle in Transcranial Magnetic Stimulation
    Lu, Mai
    Ueno, Shoogo
    2014 XXXITH URSI GENERAL ASSEMBLY AND SCIENTIFIC SYMPOSIUM (URSI GASS), 2014,
  • [27] Modelling of the Electromagnetic Field Distributions Induced by Different Transcranial Magnetic Stimulation Coil Configurations
    Xie, Bin
    Deng, Bin
    Wang, Jiang
    Yi, Guosheng
    Song, Zhenxi
    Lin, Lin
    2018 13TH WORLD CONGRESS ON INTELLIGENT CONTROL AND AUTOMATION (WCICA), 2018, : 1531 - 1536
  • [28] Coil optimization for homogeneous magnetic field with small leakage field
    Sato, S
    Sakaguchi, S
    Futamata, K
    Katou, K
    IEEE TRANSACTIONS ON MAGNETICS, 2000, 36 (04) : 649 - 653
  • [29] MAGNETIC STIMULATION OF THE HUMAN BRAIN
    UENO, S
    MATSUDA, T
    ANNALS OF THE NEW YORK ACADEMY OF SCIENCES-SERIES, 1992, 649 : 366 - 368
  • [30] Focusing-analysis of induced electric field in magnetic brain stimulation
    Shenshan
    Zheng, XL
    IEEE-EMBS ASIA PACIFIC CONFERENCE ON BIOMEDICAL ENGINEERING - PROCEEDINGS, PTS 1 & 2, 2000, : 730 - 731