Prototype of electrochemical sensor for measurements of volatile organic compounds in gases

被引:22
作者
Gebicki, J. [1 ]
Kloskowski, A. [2 ]
Chrzanowski, W. [2 ]
机构
[1] Gdansk Univ Technol, Dept Chem & Proc Engn, Fac Chem, PL-80233 Gdansk, Poland
[2] Gdansk Univ Technol, Dept Phys Chem, Fac Chem, PL-80233 Gdansk, Poland
关键词
Sensor; VOC; Ionic liquid; Limit of detection; Sensitivity; TEMPERATURE IONIC LIQUIDS; WALLED CARBON NANOTUBES; 1-ETHYL-3-METHYLIMIDAZOLIUM BIS(TRIFLUOROMETHYLSULFONYL)IMIDE; MEMBRANE; AMMONIA; DIMETHYLFORMAMIDE; ACETALDEHYDE; ELECTRODES; COMPOSITE; OXIDATION;
D O I
10.1016/j.snb.2012.12.025
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The paper presents a laboratory prototype of an electrochemical sensor for measurements of volatile organic compounds (VOCs) in gases, utilizing ionic liquids and commercial screen printed electrodes by DropSens. The following ionic liquids have been tested as the electrolyte and redox reaction environment: 1-butyl, 3-methylimidazolium dicyanoamide ([BMIM][N(CN)(2)]), 1-octyl, 3-methylimidazolium tetrafluoroborate ([OMIM][BF4]), 1-butyl, 3-methylimidazolium bis (trifluoromethanesulfonyl)imide ([BMIM][Tf2N]). VOCs investigated in the studies were: benzaldehyde, methyl benzoate, acetophenone. Gaseous reference mixtures have been prepared using permeation technique. Two metrological parameters of the sensor have been subjected to validation: limit of detection and sensitivity. The most advantageous parameters have been exhibited by the sensor for measurement of benzaldehyde in argon environment, where [OMIM][BF4] has been the ionic liquid employed. The limit of detection was at a level of 1 ppm by volume, whereas sensitivity amounted to 200 nA/ppm. The limit of detection has been evaluated in three ways: (a) based on the residual standard deviation of the calibration curve, (b) using standard deviation of the intercept of the calibration curve, and (c) utilizing standard deviation of the series of the results obtained for blank sample. (c) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:1173 / 1179
页数:7
相关论文
共 38 条
[1]   Use of room temperature ionic liquids in gas sensor design [J].
Buzzeo, MC ;
Hardacre, C ;
Compton, RG .
ANALYTICAL CHEMISTRY, 2004, 76 (15) :4583-4588
[2]   Elucidation of the electrochemical oxidation pathway of ammonia in dimethylformamide and the room temperature ionic liquid, 1-ethyl-3-methylimidazolium bis(trifluoromethylsulfonyl)imide [J].
Buzzeo, MC ;
Giovanelli, D ;
Lawrence, NS ;
Hardacre, C ;
Seddon, KR ;
Compton, RG .
ELECTROANALYSIS, 2004, 16 (11) :888-896
[3]   Ultrathin Films of Single-Walled Carbon Nanotubes for Electronics and Sensors: A Review of Fundamental and Applied Aspects [J].
Cao, Qing ;
Rogers, John A. .
ADVANCED MATERIALS, 2009, 21 (01) :29-53
[4]   Amperometric SO2 gas sensors based on solid polymer electrolytes [J].
Chiou, CY ;
Chou, TC .
SENSORS AND ACTUATORS B-CHEMICAL, 2002, 87 (01) :1-7
[5]   Temporal variation of carbonyl compound concentrations at a semi-rural site in Denmark [J].
Christensen, CS ;
Skov, H ;
Nielsen, T ;
Lohse, C .
ATMOSPHERIC ENVIRONMENT, 2000, 34 (02) :287-296
[6]   Electrosynthesis in room-temperature ionic liquids: benzaldehyde reduction [J].
Doherty, AP ;
Brooks, CA .
ELECTROCHIMICA ACTA, 2004, 49 (22-23) :3821-3826
[7]   Ionic liquids. Green solvents for the future [J].
Earle, MJ ;
Seddon, KR .
PURE AND APPLIED CHEMISTRY, 2000, 72 (07) :1391-1398
[8]   Metrological parameters of sulphur dioxide amperometric sensor containing addition of aprotic solvents [J].
Gebicki, J. ;
Chachulski, B. .
SENSORS AND ACTUATORS B-CHEMICAL, 2009, 141 (01) :249-255
[9]   Influence of Analyte Flow Rate on Signal and Response Time of the Amperometric Gas Sensor with Nafion Membrane [J].
Gebicki, J. ;
Chachulski, B. .
ELECTROANALYSIS, 2009, 21 (14) :1568-1576
[10]   Effect of oxygenation time on signal of a sensor based on ionic liquids [J].
Gebicki, Jacek ;
Kloskowski, Adam ;
Chrzanowski, Wojciech .
ELECTROCHIMICA ACTA, 2011, 56 (27) :9910-9915