Delivery of titrated oxygen via a self-inflating resuscitation bag

被引:18
作者
Young, Paul [1 ,2 ]
Pilcher, Janine [2 ]
Patel, Mitesh [2 ]
Cameron, Laird [2 ]
Braithwaite, Irene [2 ]
Weatherall, Mark [1 ,2 ,3 ]
Beasley, Richard [1 ,2 ]
机构
[1] Capital & Coast Dist Hlth Board, Wellington, New Zealand
[2] Med Res Inst New Zealand, Wellington, New Zealand
[3] Univ Otago, Sch Med, Wellington, New Zealand
关键词
Oxygen delivery; Resuscitation; Bag-valve mask; CARDIAC-ARREST; ARTERIAL HYPEROXIA;
D O I
10.1016/j.resuscitation.2012.08.330
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Aim: To investigate whether titration of inspired oxygen can be achieved through adjustment of oxygen flow into a self-inflating resuscitation bag with a reservoir of a type used in standard ambulance practice. Methods: In a series of bench experiments, oxygen was delivered via a flow metre to a 1500 ml self-inflating resuscitation bag with a 2500 ml reservoir bag and connected to a test lung. The oxygen concentration delivered to the test lung by manual inflation of the resuscitation bag was measured using an anaesthetic machine while the delivered tidal volume was measured using a respirometer. The delivered oxygen concentration was measured at flows of 0.5, 2, 6, 12 and 15 l min(-1) for tidal volumes of 300, 600, and 900 ml with bag inflation rates of 10, 20 and 30 min(-1). Results: A wide range of delivered oxygen concentrations ranging between 24% and 99.5% were achieved with different oxygen flows, tidal volumes, and inflation rates. Overall, the mean delivered oxygen concentration increased significantly with each of the increments of oxygen flow tested (p < 0.001 for all comparisons). Conclusions: Effective titration of oxygen delivery can be achieved using adjustment of oxygen flow with a standard self-inflating resuscitation bag and reservoir. (C) 2012 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:391 / 394
页数:4
相关论文
共 8 条
[1]   Oxygen concentrations at the rear and front of the bag in bag valve mask devices during oxygenation of neonates [J].
Bakr, AF ;
El-Attar, A .
RESUSCITATION, 2005, 66 (02) :217-223
[2]   Arterial hyperoxia and in-hospital mortality after resuscitation from cardiac arrest [J].
Bellomo, Rinaldo ;
Bailey, Michael ;
Eastwood, Glenn M. ;
Nichol, Alistair ;
Pilcher, David ;
Hart, Graeme K. ;
Reade, Michael C. ;
Egi, Moritoki ;
Cooper, D. James .
CRITICAL CARE, 2011, 15 (02)
[3]  
Godoy Armando Carlos Franco de, 2009, Rev. Bras. Anestesiol., V59, P21, DOI 10.1590/S0034-70942009000100004
[4]   Oxygen therapy post-cardiac arrest: The 'Goldilocks' principle? [J].
Hommers, Clare .
RESUSCITATION, 2010, 81 (12) :1605-1606
[5]   Relationship Between Supranormal Oxygen Tension and Outcome After Resuscitation From Cardiac Arrest [J].
Kilgannon, J. Hope ;
Jones, Alan E. ;
Parrillo, Joseph E. ;
Dellinger, R. Phillip ;
Milcarek, Barry ;
Hunter, Krystal ;
Shapiro, Nathan I. ;
Trzeciak, Stephen .
CIRCULATION, 2011, 123 (23) :2717-2722
[6]   Association Between Arterial Hyperoxia Following Resuscitation From Cardiac Arrest and In-Hospital Mortality [J].
Kilgannon, J. Hope ;
Jones, Alan E. ;
Shapiro, Nathan I. ;
Angelos, Mark G. ;
Milcarek, Barry ;
Hunter, Krystal ;
Parrillo, Joseph E. ;
Trzeciak, Stephen .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2010, 303 (21) :2165-2171
[7]   Comparison of 30 and the 100% inspired oxygen concentrations during early post-resuscitation period:: a randomised controlled pilot study [J].
Kuisma, M ;
Boyd, J ;
Voipio, V ;
Alaspää, A ;
Roine, RO ;
Rosenberg, P .
RESUSCITATION, 2006, 69 (02) :199-206
[8]   The effect of hyperoxia following cardiac arrest - A systematic review and meta-analysis of animal trials [J].
Pilcher, Janine ;
Weatherall, Mark ;
Shirtcliffe, Philippa ;
Bellomo, Rinaldo ;
Young, Paul ;
Beasley, Richard .