Non-Invasive Hemodynamic Assessment of Aortic Coarctation: Validation with In Vivo Measurements

被引:53
作者
Itu, Lucian [1 ,2 ]
Sharma, Puneet [1 ]
Ralovich, Kristof [1 ,3 ]
Mihalef, Viorel [1 ]
Ionasec, Razvan [1 ]
Everett, Allen [4 ]
Ringel, Richard [4 ]
Kamen, Ali [1 ]
Comaniciu, Dorin [1 ]
机构
[1] Siemens Corp Res & Technol, Princeton, NJ 08540 USA
[2] Transilvania Univ Brasov, Brasov 500024, Romania
[3] Tech Univ Munich, D-85748 Garching, Germany
[4] Johns Hopkins Univ, Sch Med, Baltimore, MD 21287 USA
关键词
Pressure gradient; CFD; Reduced-order models; Coarctation of aorta; Non-invasive; PC-MRI; PEAK-TO-PEAK; BLOOD-FLOW; ARTERIAL; PRESSURE; SIMULATION; GRADIENT; MODELS;
D O I
10.1007/s10439-012-0715-0
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We propose a CFD-based approach for the non-invasive hemodynamic assessment of pre- and post-operative coarctation of aorta (CoA) patients. Under our approach, the pressure gradient across the coarctation is determined from computational modeling based on physiological principles, medical imaging data, and routine non-invasive clinical measurements. The main constituents of our approach are a reduced-order model for computing blood flow in patient-specific aortic geometries, a parameter estimation procedure for determining patient-specific boundary conditions and vessel wall parameters from non-invasive measurements, and a comprehensive pressure-drop formulation coupled with the overall reduced-order model. The proposed CFD-based algorithm is fully automatic, requiring no iterative tuning procedures for matching the computed results to observed patient data, and requires approximately 6-8 min of computation time on a standard personal computer (Intel Core2 Duo CPU, 3.06 GHz), thus making it feasible for use in a clinical setting. The initial validation studies for the pressure-drop computations have been performed on four patient datasets with native or recurrent coarctation, by comparing the results with the invasively measured peak pressure gradients recorded during routine cardiac catheterization procedure. The preliminary results are promising, with a mean absolute error of less than 2 mmHg in all the patients.
引用
收藏
页码:669 / 681
页数:13
相关论文
共 25 条
[1]  
[Anonymous], J CARDIOVASC MAGN RE
[2]   In Vivo Validation of Numerical Prediction for Turbulence Intensity in an Aortic Coarctation [J].
Arzani, Amirhossein ;
Dyverfeldt, Petter ;
Ebbers, Tino ;
Shadden, Shawn C. .
ANNALS OF BIOMEDICAL ENGINEERING, 2012, 40 (04) :860-870
[3]  
Bessems D., 2007, THESIS TU EINDHOVEN
[4]   Computational Fluid Dynamic Simulations of Aortic Coarctation Comparing the Effects of Surgical- and Stent-Based Treatments on Aortic Compliance and Ventricular Workload [J].
Coogan, Jessica Shih ;
Chan, Frandics Pak ;
Taylor, Charles Anthony ;
Feinstein, Jeffrey Allan .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2011, 77 (05) :680-691
[5]  
Formaggia Luca, 2006, Computer Methods in Biomechanics and Biomedical Engineering, V9, P273, DOI 10.1080/10255840600857767
[6]   Analytical modeling of the instantaneous pressure gradient across the aortic valve [J].
Garcia, D ;
Pibarot, P ;
Durand, LG .
JOURNAL OF BIOMECHANICS, 2005, 38 (06) :1303-1311
[7]   Velocity-encoded cine MR imaging in aortic coarctation: Functional assessment of hemodynamic events [J].
Hom, Jeffrey J. ;
Ordovas, Karen ;
Reddy, Gautham P. .
RADIOGRAPHICS, 2008, 28 (02) :407-416
[8]   Measuring aortic pulse wave velocity using high-field cardiovascular magnetic resonance: comparison of techniques [J].
Ibrahim, El-Sayed H. ;
Johnson, Kevin R. ;
Miller, Alan B. ;
Shaffer, Jean M. ;
White, Richard D. .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2010, 12
[9]  
Kadem L, 2006, J HEART VALVE DIS, V15, P609
[10]   A new approach for the evaluation of the severity of coarctation of the aorta using Doppler velocity index and effective orifice area: In vitro validation and clinical implications [J].
Keshavarz-Motamed, Z. ;
Garcia, J. ;
Maftoon, N. ;
Bedard, E. ;
Chetaille, P. ;
Kadem, L. .
JOURNAL OF BIOMECHANICS, 2012, 45 (07) :1239-1245