HERMETIC IMPLANTABLE ANTENNA INSIDE VITREOUS HUMOR SIMULATING FLUID

被引:13
作者
Permana, H. [1 ]
Fang, Q. [1 ]
Rowe, W. S. T. [1 ]
机构
[1] RMIT Univ, Sch Elect & Comp Engn, Melbourne, Vic 3000, Australia
来源
PROGRESS IN ELECTROMAGNETICS RESEARCH-PIER | 2012年 / 133卷
关键词
RETINAL PROSTHESIS; BIOTELEMETRY COMMUNICATION; BANDWIDTH ENHANCEMENT; MACULAR DEGENERATION; DESIGN; DEVICES; SKIN;
D O I
10.2528/PIER12090806
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Retinal prosthesis system is currently being developed in various places around the world. this system involved data transfer between an implanted antenna inside an eyeball and an external camera that is located just in front of the eyeball. While there are plenty of publications about the stimulating electrodes or the processing unit of the system itself, very limited amount has been published regarding the wireless communication link between the two antennas despite the fact that the electromagnetic wave will propagate through a complex medium in the form of Vitreous Humor. this paper will discuss about the constraints associated with implanting an antenna into an eyeball. An antenna design and simulation was performed with the aid of High Frequency Structure Simulator (HFSS) and its Finite Element Method (FEM) mathematical solver in the operating frequency of 402-405 MHz. The antenna, which was a 4 layer microstrip antenna, was positioned at the centre of a spherical model filled with homogeneous Vitreous Humor material. Antenna performances that include return loss, bandwidth, gain, radiation pattern, and SAR value are analysed and compared against those of other implantable antennas operating in Medical Implant Communication Service (MICS) band. Free space and simulating fluid measurements were also conducted on the fabricated antenna to validate the simulation results. It was concluded that the fabricated antenna was able to produce the similar performance to the simulation results and hence at the same level as the other antennas operating in material with lower dielectric constants and conductivities.
引用
收藏
页码:571 / 590
页数:20
相关论文
共 26 条
[1]  
Balanis ConstantineA., 2010, Antenna Theory, VThird
[2]  
Blake L. V, 2009, ANTENNAS FUNDAMENTAL, P371
[3]  
Friedman DS, 2004, ARCH OPHTHALMOL-CHIC, V122, P564
[4]   The dielectric properties of biological tissues .2. Measurements in the frequency range 10 Hz to 20 GHz [J].
Gabriel, S ;
Lau, RW ;
Gabriel, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2251-2269
[5]   Rectenna Application of Miniaturized Implantable Antenna Design for Triple-Band Biotelemetry Communication [J].
Huang, Fu-Jhuan ;
Lee, Chien-Ming ;
Chang, Chia-Lin ;
Chen, Liang-Kai ;
Yo, Tzong-Chee ;
Luo, Ching-Hsing .
IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, 2011, 59 (07) :2646-2653
[6]  
IEEE, 1979, IEEE STNAD TEST PROC
[7]  
Johnson R.C., 1993, ANTENNA ENG HDB, V3rd
[8]   Design of a dual-band implantable antenna and development of skin mimicking gels for continuous glucose monitoring [J].
Karacolak, Tutku ;
Hood, Aaron Z. ;
Topsakal, Erdem .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2008, 56 (04) :1001-1008
[9]   In Vivo Verification of Implantable Antennas Using Rats as Model Animals [J].
Karacolak, Tutku ;
Cooper, Robert ;
Butler, James ;
Fisher, Stephen ;
Topsakal, Erdem .
IEEE ANTENNAS AND WIRELESS PROPAGATION LETTERS, 2010, 9 :334-337
[10]   Electrical Properties of Rat Skin and Design of Implantable Antennas for Medical Wireless Telemetry [J].
Karacolak, Tutku ;
Cooper, Robert ;
Topsakal, Erdem .
IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, 2009, 57 (09) :2806-2812