End-column amperometric detection of 6-mercaptopurine by capillary zone electrophoresis

被引:0
作者
Wang, XP [1 ]
You, TY [1 ]
Yang, XR [1 ]
Wang, EK [1 ]
机构
[1] Chinese Acad Sci, Changchun Inst Appl Chem, Electroanalyt Chem Lab, Natl Analyt Res Ctr Electrochem & Spect, Changchun 130022, Peoples R China
关键词
capillary electrophoresis; amperometric detection; 6-mercaptopurine;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
End-column amperometric detection of 6-mercaptopurine by capillary zone electrophoresis was described with high resolution and speed. The detection conditions were optimized and the electrochemical behavior was observed. Under the optimal conditions: detection potential of 1.2 V ( vs. Ag/AgCl), operating voltage of 15 kV, sample injection of 3 s at 15 kV and 10 mmol/l Na2HPO4 buffer, the detection limit for 6-MP was as low as 1 x 10(-7) mol/L and the linear range was from 5 x 10(-4) to 5 x 10(-6) mol/L with the relative coefficient of 0.995. The RSD of reproducibility for peak current and migration time was 2.5% and 1.2%, respectively. This method was utilized in assay real sample of human mine and bovine serum albumin containing 6-mercaptopurine.
引用
收藏
页码:1141 / 1145
页数:5
相关论文